Relationship Between Speech Recognition in Quiet and Noise and Fitting Parameters, Impedances and ECAP Thresholds in Adult Cochlear Implant Users

被引:19
作者
de Graaff, Feike [1 ]
Lissenberg-Witte, Birgit I. [2 ]
Kaandorp, Marre W. [1 ]
Merkus, Paul [1 ]
Goverts, S. Theo [1 ]
Kramer, Sophia E. [1 ]
Smits, Cas [1 ]
机构
[1] Vrije Univ Amsterdam, Amsterdam Publ Hlth Res, Amsterdam UMC, Otolaryngol Head & Neck Surg,Ear & Hearing, POB 7057, NL-1007 MB Amsterdam, Netherlands
[2] Vrije Univ Amsterdam, Amsterdam UMC, Epidemiol & Biostat, Amsterdam, Netherlands
关键词
Cochlear implant; ECAP; Fitting parameters; Impedance; Multivariable linear regression; NRT thresholds; Speech recognition; INPUT DYNAMIC-RANGE; PERCEPTION; LEVEL; HEARING; PERFORMANCE; INTERFACE; PROCESSOR; ABILITIES; POSITION; PROGRAM;
D O I
10.1097/AUD.0000000000000814
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objectives: The objective of this study was to identify parameters which are related to speech recognition in quiet and in noise of cochlear implant (CI) users. These parameters may be important to improve current fitting practices. Design: Adult CI users who visited the Amsterdam UMC, location VUmc, for their annual follow-up between January 2015 and December 2017 were retrospectively identified. After applying inclusion criteria, the final study population consisted of 138 postlingually deaf adult Cochlear CI users. Prediction models were built with speech recognition in quiet and in noise as the outcome measures, and aided sound field thresholds, and parameters related to fitting (i.e., T and C levels, dynamic range [DR]), evoked compound action potential thresholds and impedances as the independent variables. A total of 33 parameters were considered. Separate analyses were performed for postlingually deafened CI users with late onset (LO) and CI users with early onset (EO) of severe hearing impairment. Results: Speech recognition in quiet was not significantly different between the LO and EO groups. Speech recognition in noise was better for the LO group compared with the EO group. For CI users in the LO group, mean aided thresholds, mean electrical DR, and measures to express the impedance profile across the electrode array were identified as predictors of speech recognition in quiet and in noise. For CI users in the EO group, the mean T level appeared to be a significant predictor in the models for speech recognition in quiet and in noise, such that CI users with elevated T levels had worse speech recognition in quiet and in noise. Conclusions: Significant parameters related to speech recognition in quiet and in noise were identified: aided thresholds, electrical DR, T levels, and impedance profiles. The results of this study are consistent with previous study findings and may guide audiologists in their fitting practices to improve the performance of CI users. The best performance was found for CI users with aided thresholds around the target level of 25 dB HL, and an electrical DR between 40 and 60 CL. However, adjustments of T and/or C levels to obtain aided thresholds around the target level and the preferred DR may not always be acceptable for individual CI users. Finally, clinicians should pay attention to profiles of impedances other than a flat profile with mild variations.
引用
收藏
页码:935 / 947
页数:13
相关论文
共 59 条
  • [1] Blamey P, 1996, Audiol Neurootol, V1, P293
  • [2] Factors Affecting Auditory Performance of Postlinguistically Deaf Adults Using Cochlear Implants: An Update with 2251 Patients
    Blamey, Peter
    Artieres, Franoise
    Baskent, Deniz
    Bergeron, Francois
    Beynon, Andy
    Burke, Elaine
    Dillier, Norbert
    Dowell, Richard
    Fraysse, Bernard
    Gallego, Stephane
    Govaerts, Paul J.
    Green, Kevin
    Huber, Alexander M.
    Kleine-Punte, Andrea
    Maat, Bert
    Marx, Mathieu
    Mawman, Deborah
    Mosnier, Isabelle
    O'Connor, Alec Fitzgerald
    O'Leary, Stephen
    Rousset, Alexandra
    Schauwers, Karen
    Skarzynski, Henryk
    Skarzynski, Piotr H.
    Sterkers, Olivier
    Terranti, Assia
    Truy, Eric
    Van de Heyning, Paul
    Venail, Frederic
    Vincent, Christophe
    Lazard, Diane S.
    [J]. AUDIOLOGY AND NEURO-OTOLOGY, 2013, 18 (01) : 36 - 47
  • [3] FACTORS PREDICTING POSTOPERATIVE SENTENCE SCORES IN POSTLINGUISTICALLY DEAF ADULT COCHLEAR IMPLANT PATIENTS
    BLAMEY, PJ
    PYMAN, BC
    CLARK, GM
    DOWELL, RC
    GORDON, M
    BROWN, AM
    HOLLOW, RD
    [J]. ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1992, 101 (04) : 342 - 348
  • [4] Bosman AJ, 1995, AUDIOLOGY, V34, P260
  • [5] Neural Response Telemetry Reconsidered: I. The Relevance of ECAP Threshold Profiles and Scaled Profiles to Cochlear Implant Fitting
    Botros, Andrew
    Psarros, Colleen
    [J]. EAR AND HEARING, 2010, 31 (03) : 367 - 379
  • [6] Neural Response Telemetry Reconsidered: II. The Influence of Neural Population on the ECAP Recovery Function and Refractoriness
    Botros, Andrew
    Psarros, Colleen
    [J]. EAR AND HEARING, 2010, 31 (03) : 380 - 391
  • [7] The relationship between EAP and EABR thresholds and levels used to program the nucleus 24 speech processor: Data from adults
    Brown, CJ
    Hughes, ML
    Luk, B
    Abbas, PJ
    Wolaver, A
    Gervais, J
    [J]. EAR AND HEARING, 2000, 21 (02) : 151 - 163
  • [8] Busby P A, 2002, Cochlear Implants Int, V3, P87, DOI 10.1179/cim.2002.3.2.87
  • [9] Effects of Threshold Adjustment on Speech Perception in Nucleus Cochlear Implant Recipients
    Busby, Peter A.
    Arora, Komal
    [J]. EAR AND HEARING, 2016, 37 (03) : 303 - 311
  • [10] Cochlear L, 2012, CLIN GUIDANCE DOCUME