Percutaneous Microwave Tumor Ablation Is Safe in Patients with Cardiovascular Implantable Electronic Devices: A Single-Institutional Retrospective Review

被引:3
作者
Swietlik, John F. [1 ]
Longo, Katherine C. [1 ]
Knott, Emily A. [1 ]
Hinshaw, J. Louis [1 ]
Brace, Christopher L. [2 ]
Alexander, Marci [1 ]
Lubner, Meghan G. [1 ]
Wells, Shane A. [1 ]
Lee, Fred T., Jr. [1 ]
Ziemlewicz, Timothy J. [1 ]
机构
[1] Univ Wisconsin, Dept Radiol, 600 Highland Ave, Madison, WI 53792 USA
[2] Univ Wisconsin, Dept Med Phys, 600 Highland Ave, Madison, WI 53792 USA
关键词
RADIOFREQUENCY ABLATION; CARDIAC DEVICES; INTERFERENCE; EFFICACY;
D O I
10.1016/j.jvir.2018.12.001
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The risk of electromagnetic interference between microwave (MW) ablation and cardiac implantable electronic devices (CIEDs), ie, pacemakers and defibrillators, has not been fully evaluated. Fourteen MW ablations (kidney, n = 8; liver, n = 5; lung, n = 1) were performed in 13 patients with CIEDs in normal operating mode. Electrocardiography tracings, cardiovascular complications, tumor size, tumor-to-CIED distance, and tumor-to-device lead distance were recorded. Mean tumor size was 2.9 cm, mean tumor-to-CIED distance was 26.4 cm (range, 9-30 cm), and mean tumor-to-lead distance was 12.1 cm (range, 3.5-20 cm). No device-based cardiovascular complications or class C or higher complications per Society of Interventional Radiology criteria were identified. MW ablation appears to be safe in select patients with CIEDs.
引用
收藏
页码:396 / 400
页数:5
相关论文
共 18 条
[1]  
Boston Scientific, 2008, RAD ABL IMPL DEV SYS
[2]   Microwave tissue ablation: Biophysics, technology, and applications [J].
Brace C.L. .
Critical Reviews in Biomedical Engineering, 2010, 38 (01) :65-78
[3]   Renal Mass and Localized Renal Cancer: AUA Guideline [J].
Campbell, Steven ;
Uzzo, Robert G. ;
Allaf, Mohamad E. ;
Bass, Eric B. ;
Cadeddu, Jeffrey A. ;
Chang, Anthony ;
Clark, Peter E. ;
Davis, Brian J. ;
Derweesh, Ithaar H. ;
Giambarresi, Leo ;
Gervais, Debra A. ;
Hu, Susie L. ;
Lane, Brian R. ;
Leibovich, Bradley C. ;
Pierorazio, Philip M. .
JOURNAL OF UROLOGY, 2017, 198 (03) :520-529
[4]   Safety and Efficacy of Percutaneous Microwave Hepatic Ablation Near the Heart [J].
Carberry, George A. ;
Smolock, Amanda R. ;
Cristescu, Mircea ;
Wells, Shane A. ;
Ziemlewicz, Timothy J. ;
Lubner, Meghan G. ;
Hinshaw, J. Louis ;
Brace, Christopher L. ;
Lee, Fred T., Jr. .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2017, 28 (04) :490-497
[5]  
Cohan L., 2008, Cardiac Pacing for the Clinician, P595
[6]  
Crossley GH, 2011, HEART RHYTHM, V8, P1114, DOI 10.1016/j.hrthm.2010.12.023
[7]   Pacemaker reprogramming after radiofrequency ablation of a lung neoplasm [J].
Donohoo, Jay H. ;
Anderson, Maria T. ;
Mayo-Smith, William W. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 189 (04) :890-892
[8]   Hepatocellular carcinoma [J].
Forner, Alejandro ;
Llovet, Josep M. ;
Bruix, Jordi .
LANCET, 2012, 379 (9822) :1245-1255
[9]   The National ICD Registry Report: Version 2.1 including leads and pediatrics for years 2010 and 2011 [J].
Kremers, Mark S. ;
Hammill, Stephen C. ;
Berul, Charles I. ;
Koutras, Christina ;
Curtis, Jeptha S. ;
Wang, Yongfei ;
Beachy, Jim ;
Meisnere, Laura Blum ;
Conyers, Del M. ;
Reynolds, Matthew R. ;
Heidenreich, Paul A. ;
Al-Khatib, Sana M. ;
Pina, Ileana L. ;
Blake, Kathleen ;
Walsh, Mary Norine ;
Wilkoff, Bruce L. ;
Shalaby, Alaa ;
Masoudi, Frederick A. ;
Rumsfeld, John .
HEART RHYTHM, 2013, 10 (04) :E59-E65
[10]   Long-term survival after permanent pacemaker implantation in young adults: 30 year experience [J].
Mayosi, BM ;
Little, F ;
Millar, RNS .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (03) :407-412