Characterisation and optimisation of ion discrimination in a mini ion funnel for a miniature mass spectrometer

被引:14
作者
Zhang, Xiaohua [1 ]
Huo, Xinming [2 ]
Tang, Fei [2 ]
Zha, Zhuoyue [3 ]
Yang, Fengbo [3 ]
Ni, Yafan [3 ]
Wang, Yanjun [3 ]
Zhou, Mingfei [1 ]
Wang, Xiaohao [2 ,4 ]
机构
[1] Fudan Univ, Dept Chem, Shanghai 200433, Peoples R China
[2] Tsinghua Univ, Dept Precis Instrument, State Key Lab Precis Measurement Technol & Instrm, Beijing 100084, Peoples R China
[3] Anyeep Instrumentat Co, Suzhou 215129, Peoples R China
[4] Tsinghua Univ, Grad Sch Shenzhen, Div Adv Mfg, Shenzhen 518055, Peoples R China
基金
中国国家自然科学基金;
关键词
ATMOSPHERIC-PRESSURE; MULTICAPILLARY INLET; INTERFACE; SENSITIVITY; TRANSMISSION; DESIGN; OPTICS;
D O I
10.1039/c9ay00677j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Ion funnels have been miniaturised and integrated into miniature ion trap mass spectrometers to improve instrument sensitivity. However, the ion discrimination effect of the ion funnel constrains its application. Previous research has generally reported that the ion discrimination is related mainly to the mass of the ions. In this paper, the Runge-Kutta algorithm was used to characterise the trajectories of ions in the funnel under collisions with gas molecules. It is theoretically and empirically proved that, in a continuous atmospheric-pressure-interfaced miniature mass spectrometer, ion discrimination in the funnel shows two aspects: mass discrimination and collision cross-section discrimination. It was found that DC gradient, RF amplitude and operating pressure of the ion funnel will affect ion transmission. Among these parameters, RF amplitude is the key parameter controlling ion discrimination. To weaken the ion discrimination, RF amplitude scanning method was introduced in the injection section of the ion trap to improve ion transmission efficiency over the entire mass range. After optimisation of the ion funnel, the dynamic range of the instrument concentration reached four orders of magnitude. The detection limit of roxithromycin reached a low level of 1 ng mL(-1). The results of this study increase the theoretical understanding of the ion funnel, providing guidance for applying ion funnels in improving the sensitivity of miniature mass spectrometers and expanding their actual applicability.
引用
收藏
页码:2551 / 2558
页数:8
相关论文
共 32 条
[1]   A novel planar ion funnel design for miniature ion optics [J].
Chaudhary, A. ;
van Amerom, Friso H. W. ;
Short, R. T. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2014, 85 (10)
[2]   Rectangular Ion Funnel: A New Ion Funnel Interface for Structures for Loss less Ion Manipulations [J].
Chen, Tsung-Chi ;
Webb, Ian K. ;
Prost, Spencer A. ;
Harrer, Marques B. ;
Norheim, Randolph V. ;
Tang, Keqi ;
Ibrahim, Yehia M. ;
Smith, Richard D. .
ANALYTICAL CHEMISTRY, 2015, 87 (01) :716-722
[3]   Improved ion optics for introduction of ions into a 9.4-T Fourier transform ion cyclotron resonance mass spectrometer [J].
Chen, Yu ;
Leach, Franklin E., III ;
Kaiser, Nathan K. ;
Dang, Xibei ;
Ibrahim, Yehia M. ;
Norheim, Randolph V. ;
Anderson, Gordon A. ;
Smith, Richard D. ;
Marshall, Alan G. .
JOURNAL OF MASS SPECTROMETRY, 2015, 50 (01) :280-284
[4]  
GERLICH D, 1992, ADV CHEM PHYS, V82, P1
[5]   THEORY OF HIGH-RESOLUTION MASS-SPECTROMETRY ACHIEVED VIA RESONANCE EJECTION IN THE QUADRUPOLE ION TRAP [J].
GOERINGER, DE ;
WHITTEN, WB ;
RAMSEY, JM ;
MCLUCKEY, SA ;
GLISH, GL .
ANALYTICAL CHEMISTRY, 1992, 64 (13) :1434-1439
[6]  
Guo Q., 2018, CHINESE CHEM LETT, V29, P46
[7]   Characteristics of a rf-only hexapole ion-guide interface for plasma-source time-of-flight mass spectrometry [J].
Guzowski, JP ;
Hieftje, GM .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (08) :781-792
[8]   Ion funnel trap interface for orthogonal time-of-flight mass spectrometry [J].
Ibrahim, Yehia ;
Belov, Mikhail E. ;
Tolmachev, Aleksey V. ;
Prior, David C. ;
Smith, Richard D. .
ANALYTICAL CHEMISTRY, 2007, 79 (20) :7845-7852
[9]   Improving mass spectrometer sensitivity using a high-pressure electrodynamic ion funnel interface [J].
Ibrahim, Yehia ;
Tang, Keqi ;
Tolmachev, Aleksey V. ;
Shvartsburg, Alexandre A. ;
Smith, Richard D. .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2006, 17 (09) :1299-1305
[10]   Improving Ion Mobility Measurement Sensitivity by Utilizing Helium in an Ion Funnel Trap [J].
Ibrahim, Yehia M. ;
Garimella, Sandilya V. B. ;
Tolmachev, Aleksey V. ;
Baker, Erin S. ;
Smith, Richard D. .
ANALYTICAL CHEMISTRY, 2014, 86 (11) :5295-5299