Comparison of methods to measure corneal power for intraocular lens power calculation using a rotating Scheimpflug camera

被引:70
作者
Savini, Giacomo [1 ]
Barboni, Piero [2 ]
Carbonelli, Michele [1 ]
Hoffer, Kenneth J. [3 ]
机构
[1] GB Bietti Fdn IRCCS, Rome, Italy
[2] Studio Oculist Azeglio, Bologna, Italy
[3] Univ Calif Los Angeles, Jules Stein Eye Inst, Los Angeles, CA 90024 USA
关键词
CATARACT-SURGERY; AUTOMATED KERATOMETRY; REFRACTIVE OUTCOMES; ACCURACY; TOPOGRAPHY; REPRODUCIBILITY; REPEATABILITY; BIOMETRY; PENTACAM; SYSTEM;
D O I
10.1016/j.jcrs.2012.11.022
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To assess the accuracy of corneal power measurements provided by a Scheimpflug camera (Pentacam HR) for intraocular lens (IOL) power calculation in unoperated eyes and compare the results with those of simulated keratometry (SimK) performed with a Placido-disk corneal topographer (Keratron). SETTING: Private practice. DESIGN: Evaluation of diagnostic test. METHODS: Eight Scheimpflug camera corneal power measurements were analyzed: (1) average K, (2) true net power calculated using the Gaussian optics formula, (3) total corneal refractive power at 2.0 mm calculated by ray tracing on a ring and (4) as the average of the zone inside the ring, (5) total corneal refractive power at 3.0 mm on a ring and (6) as the average of the zone inside the ring, (7) the equivalent K reading at 3.0 mm and (8) at 4.5 mm. The IOL power was calculated using the Hoffer Q, Holladay 1, and SRK/T formulas. RESULTS: No statistically significant differences were observed between any corneal power measurements, including simulated K, in 41 consecutive patients. The latter showed slightly lower mean absolute errors with all 3 formulas (range 0.26 to 0.27 diopter [D]). The Scheimpflug camera gave the lowest median absolute errors with all formulas; that is, the 3.0 mm equivalent K reading with the Hoffer Q formula (0.18 D) and Holladay 1 formula (0.17 D) and the 2.0 mm total corneal refractive power ring with the SRK/T formula (0.18 D). CONCLUSION: Corneal power measurements provided by the Scheimpflug camera and Placido disk corneal topographer displayed comparable accuracy in IOL power calculation.
引用
收藏
页码:598 / 604
页数:7
相关论文
共 32 条
[1]  
American Academy of Ophthalmology, 1996, BAS CLIN SCI COURS O, P176
[2]   Intraocular lens power calculation after corneal refractive surgery: Double-K method [J].
Aramberri, J .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (11) :2063-2068
[3]   Formula choice: Hoffer Q, Holladay 1, or SRK/T and refractive outcomes in 8108 eyes after cataract surgery with biometry by partial coherence interferometry [J].
Aristodemou, Petros ;
Cartwright, Nathaniel E. Knox ;
Sparrow, John M. ;
Johnston, Robert L. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (01) :63-71
[4]  
Dupont WD, 2012, PS POWER SAMPLE SIZE
[5]   Intraocular lens power calculation after laser in situ keratomileusis for myopia and hyperopia - A standardized approach [J].
Feiz, V ;
Mannis, MJ ;
Garcia-Ferrer, F ;
Kandavel, G ;
Darlington, JK ;
Kim, E ;
Caspar, J ;
Wang, JL ;
Wang, W .
CORNEA, 2001, 20 (08) :792-797
[6]   Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis [J].
Haigis, W ;
Lege, B ;
Miller, N ;
Schneider, B .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2000, 238 (09) :765-773
[7]  
Hoffer KJ, 2007, J CATARACT REFR SURG, V33, P2, DOI 10.1016/j.jcrs.2006.08.054
[8]   Intraocular lens power calculation after previous laser refractive surgery [J].
Hoffer, Kenneth J. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (04) :759-765
[9]   Clinical results using the Holladay 2 intraocular lens power formula [J].
Hoffer, KJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (08) :1233-1237
[10]  
Holladay JT, 2010, J REFRACT SURG, V26, P387