Mechanical Comparison of Headless Screw Fixation and Locking Plate Fixation for Talar Neck Fractures

被引:15
作者
Karakasli, Ahmet [1 ]
Hapa, Onur [1 ]
Erduran, Mehmet [1 ]
Dincer, Cemal [1 ]
Cecen, Berivan [2 ]
Havitcioglu, Hasan [1 ]
机构
[1] Dokuz Eylul Univ, Fac Med, Dept Orthopaed & Traumatol, TR-35340 Izmir, Turkey
[2] Dokuz Eylul Univ, Dept Biomech, Inst Hlth Sci, TR-35340 Izmir, Turkey
关键词
headless screw; locking plate; talar neck fractures; DYNAMIC COMPRESSION PLATE; BODY FRACTURES; SURGICAL-TREATMENT; CANCELLOUS BONE; STABILITY; LCP;
D O I
10.1053/j.jfas.2015.04.002
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
For talar neck fractures, open reduction and internal fixation have been thought to facilitate revascularization and prevent osteonecrosis. Newer screw systems allow for placement of cannulated headless screws, which provide compression by virtue of a variable pitch thread. The present study compared the biomechanical fixation strength of cannulated headless variable-pitch screw fixation and locking plate fixation. A reproducible talar neck fracture was created in 14 fresh cadaver talar necks. Talar head fixation was then performed using 2 cannulated headless variable-pitch 4-mm/5-mm diameter (4/5) screws (Acutrak; Acumed, Hillsboro, OR) and locking plate fixation. Headless variable-pitch screw fixation had lower failure displacement than did locking plate fixation. No statistically significant differences were found in failure stiffness, yield stiffness (p = .655), yield load (p = .142), or ultimate load between the 2 fixation techniques. Cannulated headless variable-pitch screw fixation resulted in better failure displacement than locking plate fixation in a cadaveric talus model and could be considered a viable option for talus fracture fixation. Headless, fully threaded, variable-pitch screw fixation has inherent advantages compared with locking plate fixation, because it might cause less damage to the articular surface and can compress the fracture for improved reduction. Additionally, plate fixation can increase the risk of avascular necrosis owing to the wider incision and dissection of soft tissues. (C) 2015 by the American College of Foot and Ankle Surgeons. All rights reserved.
引用
收藏
页码:905 / 909
页数:5
相关论文
共 19 条
[1]   Biomechanical testing of the locking compression plate: When does the distance between bone and implant significantly reduce construct stability? [J].
Ahmad, M. ;
Nanda, R. ;
Bajwa, A. S. ;
Candal-Couto, J. ;
Green, S. ;
Hui, A. C. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2007, 38 (03) :358-364
[2]   Comminuted talar neck fractures: A mechanical comparison of fixation techniques [J].
Attiah, Mohammed ;
Sanders, David W. ;
Valdivia, Gonzalo ;
Cooper, Ian ;
Ferreira, Louis ;
MacLeod, Mark D. ;
Johnson, James A. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2007, 21 (01) :47-51
[3]   Fixation Strength of Anteriorly Inserted Headless Screws for Talar Neck Fractures [J].
Capelle, Jonathan H. ;
Couch, Cory G. ;
Wells, Kevin M. ;
Morris, Randal P. ;
Buford, William L., Jr. ;
Merriman, David J. ;
Panchbhavi, Vinod K. .
FOOT & ANKLE INTERNATIONAL, 2013, 34 (07) :1012-1016
[4]  
Chateau PBF, 2002, J ORTHOP TRAUMA, V16, P213
[5]   Varus malalignment of the talar neck - Its effect on the position of the foot and on subtalar motion [J].
Daniels, TR ;
Smith, JW ;
Ross, TI .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1996, 78A (10) :1559-1567
[6]   Mechanical properties of cancellous bone of the distal humerus [J].
Dunham, CE ;
Takaki, SE ;
Johnson, JA ;
Dunning, CE .
CLINICAL BIOMECHANICS, 2005, 20 (08) :834-838
[7]   Total talar fracture - Long-term results of internal fixation of talar fractures. A multicentric study of 114 cases [J].
Fournier, A. ;
Barba, N. ;
Steiger, V. ;
Lourdais, A. ;
Frin, J. -M. ;
Williams, T. ;
Falaise, V. ;
Pineau, V. ;
de Chou, E. Salle ;
Noailles, T. ;
Carvalhana, G. ;
Ruhlmann, F. ;
Huten, D. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2012, 98 (04) :S48-S55
[8]   Mechanical properties of subchondral cancellous bone of the radial head [J].
Gordon, KD ;
Duck, TR ;
King, GJW ;
Johnson, JA .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2003, 17 (04) :285-289
[9]   Talar Neck Fractures: A Systematic Review of the Literature [J].
Halvorson, Jason J. ;
Winter, S. Bradley ;
Teasdall, Robert D. ;
Scott, Aaron T. .
JOURNAL OF FOOT & ANKLE SURGERY, 2013, 52 (01) :56-61
[10]   Open reduction and stable fixation of isolated, displaced talar neck and body fractures [J].
Lindvall, E ;
Haidukewych, G ;
DiPasquale, T ;
Herscovici, D ;
Sanders, R .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (10) :2229-2234