Temporal stability of ambulatory stroke volume and cardiac output measured by impedance cardiography

被引:30
作者
Goedhart, AD [1 ]
Kupper, N [1 ]
Willemsen, G [1 ]
Boomsma, DI [1 ]
de Geus, EJC [1 ]
机构
[1] Free Univ Amsterdam, Dept Biol Psychol, NL-1081 BT Amsterdam, Netherlands
关键词
sympathetic nervous system; temporal stability; hemodynamic regulation; impedance cardiography;
D O I
10.1016/j.biopsycho.2005.08.004
中图分类号
B84 [心理学];
学科分类号
04 ; 0402 ;
摘要
Recently, devices have become available that allow non-invasive measurement of stroke volume and cardiac output through ambulatory thorax impedance recording. If such recordings have adequate temporal stability, they offer great potential to further our understanding of how repeated or chronic cardiovascular activation in response to naturalistic events may contribute to cardiovascular disease. In this study, 24 It ambulatory impedance-derived systolic time intervals, stroke volume and cardiac output were measured in 65 healthy subjects across an average time span of 3 years and 4 months. Stability was computed separately for sleep and daytime recordings. To avoid confounding by differences in posture and physical activity across measurement days, temporal stability was computed using sitting activities only. During the day intraclass correlations were moderate for stroke volume (.29-.46) and cardiac output (.33-.46) and good for systolic time intervals (.55-.81). When test-retest comparison was limited to two comparable days (two work days or two leisure days), correlations for both SV (.42-.46) and CO (.43-.50) improved. Conclusion: Moderate long-term temporal stability is found for individual differences in ambulatory stroke volume and cardiac output measured by impedance cardiography. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:110 / 117
页数:8
相关论文
共 50 条
[21]   Reproducibility of cardiac output derived by impedance cardiography during postural changes and exercise [J].
Martin G. Schultz ;
Rachel E. D. Climie ;
Sonja B. Nikolic ;
Kiran D. Ahuja ;
James E. Sharman .
Artery Research, 2012, 6 (2) :78-84
[22]   Reproducibility of cardiac output derived by impedance cardiography during postural changes and exercise [J].
Schultz, Martin G. ;
Climie, Rachel E. D. ;
Nikolic, Sonja B. ;
Ahuja, Kiran D. ;
Sharman, James E. .
ARTERY RESEARCH, 2012, 6 (02) :78-84
[23]   Monitoring of Stroke Volume through Impedance Cardiography Using an Artificial Neural Network [J].
Naidu, S. M. M. ;
Bagal, Uttam R. ;
Pandey, Prem C. ;
Hardas, Suhas ;
Khambete, Niranjan D. .
2015 TWENTY FIRST NATIONAL CONFERENCE ON COMMUNICATIONS (NCC), 2015,
[24]   IMPEDANCE CARDIOGRAPHY FOR CARDIAC-OUTPUT ESTIMATION IN PACEMAKER PATIENTS - REVIEW OF THE LITERATURE [J].
OVSYSHCHER, I ;
FURMAN, S .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (07) :1412-1422
[25]   Left ventricular stroke volume measurement by impedance cardiography correlates with echocardiography in neonates [J].
ME Blohm ;
J Hartwich ;
D Obrecht ;
G Müller ;
J Weil ;
D Singer .
Critical Care, 16 (Suppl 1)
[26]   Bernstein-Osypka stroke volume equation for impedance cardiography: citation correction [J].
Donald P. Bernstein .
Intensive Care Medicine, 2007, 33 :923-923
[27]   STROKE VOLUME MEASUREMENT DURING SUPINE AND UPRIGHT CYCLE EXERCISE BY IMPEDANCE CARDIOGRAPHY [J].
MOON, JK ;
COGGAN, AR ;
HOPPER, MK ;
BAKER, LE ;
COYLE, EF .
ANNALS OF BIOMEDICAL ENGINEERING, 1994, 22 (05) :514-523
[28]   Stroke volume and systolic time intervals: beat-to-beat comparison between echocardiography and ambulatory impedance cardiography in supine and tilted positions [J].
Cybulski, G ;
Michalak, E ;
Kozluk, E ;
Piatkowska, A ;
Niewiadomski, W .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2004, 42 (05) :707-711
[29]   Stroke volume and systolic time intervals: Beat-to-beat comparison between echocardiography and ambulatory impedance cardiography in supine and tilted positions [J].
G. Cybulski ;
E. Michalak ;
E. Koźluk ;
A. Piątkowska ;
W. Niewiadomski .
Medical and Biological Engineering and Computing, 2004, 42 :707-711
[30]   Impedance cardiography for cardiac output estimation - Reliability of wrist-to-ankle electrode configuration [J].
Paredes, Oscar Luis ;
Shite, Junya ;
Shinke, Toshiro ;
Watanabe, Satoshi ;
Otake, Hiromasa ;
Matsumoto, Daisuke ;
Imuro, Yusuke ;
Ogasawara, Daisuke ;
Sawada, Takahiro ;
Yokoyama, Mitsuhiro .
CIRCULATION JOURNAL, 2006, 70 (09) :1164-1168