Detection of lead(II) using an glassy carbon electrode modified with Nafion, carbon nanotubes and benzo-18-crown-6

被引:38
作者
Anandhakumar, Sukeri [1 ]
Mathiyarasu, Jayaraman [1 ]
机构
[1] CSIR Cent Electrochem Res Inst, Electrod & Electrocatalysis Div, Karaikkudi 630006, Tamil Nadu, India
关键词
Carbon nanotubes; Crown ether; Anodic stripping voltammetry; Selectivity; Lead; STRIPPING VOLTAMMETRIC DETERMINATION; HEAVY-METALS; WATER SAMPLES; PRECONCENTRATION; SPECTROMETRY; EXPOSURE; MERCURY;
D O I
10.1007/s00604-013-1022-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A glassy carbon electrode was modified with Nafion, carbon nanotubes and benzo-18-crown-6 to give an electrode for the selective determination of lead(II) via square wave anodic stripping voltammetry. The use of carbon nanotubes with their extraordinary electrical conductivity and strong adsorption ability warrants high sensitivity. Benzo-18-crown-6 is employed as a "molecular scavenger" because of its excellent selectivity for lead(II). The modified electrode shows enhanced sensitivity, reproducibility and selectivity for lead(II) even without applying an electrical potential during the accumulation time. It responds linearly to lead(II) in the 1 to 30 nM concentration range (with a correlation coefficient of 0.9992) after a 10-min accumulation time. The detection limit is 1 nM. The sensor exhibits excellent selectivity over other heavy metal ions such as Cd(II), Cu(II), Zn(II), and Hg(II).
引用
收藏
页码:1065 / 1071
页数:7
相关论文
共 32 条
[1]  
Bard AJ., 1980, ELECTROCHEMICAL METH
[2]   Very low lead exposures and children's neurodevelopment [J].
Bellinger, David C. .
CURRENT OPINION IN PEDIATRICS, 2008, 20 (02) :172-177
[3]   Adsorptive stripping voltammetric detection of daunomycin at a bismuth bulk electrode [J].
Bucková, M ;
Gründler, P ;
Flechsig, GU .
ELECTROANALYSIS, 2005, 17 (5-6) :440-444
[4]   Lead exposure, IQ, and behavior in urban 5- to 7-year-olds: Does lead affect Behavior only by lowering IQ? [J].
Chen, Aimin ;
Cai, Bo ;
Dietrich, Kim N. ;
Radcliffe, Jerilynn ;
Rogan, Walter J. .
PEDIATRICS, 2007, 119 (03) :E650-E658
[5]  
Crowley K, 2002, ELECTROANAL, V14, P1077, DOI 10.1002/1521-4109(200208)14:15/16<1077::AID-ELAN1077>3.0.CO
[6]  
2-3
[7]   Mercury film electrodes: developments, trends and potentialities for electroanalysis [J].
Economou, A ;
Fielden, PR .
ANALYST, 2003, 128 (03) :205-212
[8]   Anodic stripping voltammetry of heavy metals at nanocrystalline boron-doped diamond electrode [J].
El Tall, Omar ;
Jaffrezic-Renault, Nicole ;
Sigaud, Monique ;
Vittori, Olivier .
ELECTROANALYSIS, 2007, 19 (11) :1152-1159
[9]   Highly selective optical-sensing film for lead(II) determination in water samples [J].
Ensafi, Ali A. ;
Far, A. Katiraei ;
Meghdadi, S. .
JOURNAL OF HAZARDOUS MATERIALS, 2009, 172 (2-3) :1069-1075
[10]   DETERMINATION OF CADMIUM, LEAD AND COPPER IN WATER SAMPLES BY FLAME ATOMIC-ABSORPTION SPECTROMETRY WITH PRECONCENTRATION BY FLOW-INJECTION ONLINE SORBENT EXTRACTION [J].
FANG, ZL ;
GUO, TZ ;
WELZ, B .
TALANTA, 1991, 38 (06) :613-619