How Inaccurate is Weight as a Metric for Patient Size? Comparing Patient Weight to Effective Diameter for Size-Specific Dose Estimation

被引:2
|
作者
Cook, Tessa S. [1 ]
Chadalavada, Seetharam C. [1 ]
Boonn, William W. [1 ]
机构
[1] Hosp Univ Penn, Dept Radiol, Philadelphia, PA 19104 USA
来源
MEDICAL IMAGING 2013: ADVANCED PACS-BASED IMAGING INFORMATICS AND THERAPEUTIC APPLICATIONS | 2013年 / 8674卷
关键词
patient-specific dose monitoring; computed tomography; size-specific dose estimates; patient size metrics; CT;
D O I
10.1117/12.2006132
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
One of the biggest challenges in dose monitoring is customization of CT dose estimates to the patient. Patient size remains a highly significant variable. One metric that has previously been used for patient size is patient weight, though this is often criticized as inaccurate. In this work, we compare patients' weight to their effective diameters obtained from a CT scan of the chest or the abdomen. CT exams of the chest (N= 163) and abdomen/pelvis (N= 168) performed on adult patients in July 2012 were randomly selected for analysis. The effective diameter of the patient for each exam was determined using the central slice of the scan region for each exam using eXposure (TM) (Radimetrics, Inc., Toronto, Canada). In some cases, the same patient had both a chest and abdominopelvic CT, so effective diameters from both regions were analyzed. In this small sample size, there appears to be a linear relationship between patient weight and effective diameter when measured in the mid-chest and mid-abdomen of adult patients. However, for each weight, patient effective diameter can vary by 5 cm from the regression line in both the chest and the abdomen. A 5-cm difference corresponds to a difference of approximately 0.2 in the chest and 0.3 in the abdomen/pelvis for the correction factors recommended for size-specific dose estimation by the AAPM. This preliminary data suggests that weight-based CT protocoling may in fact be appropriate for some adults. However, more work is needed to identify those patients in whom weight-based protocoling is not appropriate.
引用
收藏
页数:6
相关论文
共 48 条
  • [1] SIMPLE METHOD OF SIZE-SPECIFIC DOSE ESTIMATES CALCULATION FROM PATIENT WEIGHT ON COMPUTED TOMOGRAPHY
    Iriuchijima, Akiko
    Fukushima, Yasuhiro
    Nakajima, Takahito
    Tsushima, Yoshito
    Ogura, Akio
    RADIATION PROTECTION DOSIMETRY, 2018, 178 (02) : 208 - 212
  • [2] Can Patient's Body Weight Represent Body Diameter for Pediatric Size-Specific Dose Estimate in Thoracic and Abdominal Computed Tomography?
    Kritsaneepaiboon, Supika
    Eng-chuan, Suwadee
    Yoykaew, Saowapark
    JOURNAL OF CLINICAL IMAGING SCIENCE, 2019, 9 : 1 - 8
  • [3] Size-Specific Dose Estimates for Evaluation of Individual Patient Dose in CT Protocol for Renal Colic
    Waszczuk, Lukasz A.
    Guzinski, Maciej
    Czarnecka, Anna
    Sasiadek, Marek J.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 205 (01) : 100 - 105
  • [4] Relationship between size-specific dose estimates and image quality in computed tomography depending on patient size
    Kawashima, Hiroki
    Ichikawa, Katsuhiro
    Hanaoka, Shinsuke
    Matsubara, Kosuke
    Takata, Tadanori
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (04): : 246 - 251
  • [5] Size-specific dose estimates in pediatric brain CT in relation to age and weight
    Fujii, Kaoru
    Inoue, Yusuke
    Itoh, Hiroyasu
    Sasa, Ryosuke
    Hata, Hirofumi
    Mitsui, Kohei
    RADIATION PROTECTION DOSIMETRY, 2023, 199 (10) : 1075 - 1081
  • [6] Accuracy of patient dose estimation in cone beam computed tomography in breast irradiation by size-specific dose estimates with position correction
    Ueno, Hiroyuki
    Matsubara, Kosuke
    Bou, Sayuri
    Hizume, Masato
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (12):
  • [7] Optimizing CT radiation dose based on patient size and image quality: the size-specific dose estimate method
    Larson, David B.
    PEDIATRIC RADIOLOGY, 2014, 44 : 501 - 505
  • [8] Optimizing CT radiation dose based on patient size and image quality: the size-specific dose estimate method
    David B. Larson
    Pediatric Radiology, 2014, 44 : 501 - 505
  • [9] CNN application for automated determination of the patient's size to obtain the size-specific dose estimated in CT
    Hernandez-Davila, Erik R.
    Torres-Garcia, Eugenio
    Aranda-Lara, Liliana
    Roldan-Valadez, Ernesto
    Isaac-Olive, Keila
    Flores-Reyes, Mario
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2024, 10 (03)
  • [10] Evaluation of direct point dose estimation based on the distribution of the size-specific dose estimate
    Anam, Choirul
    Sutanto, Heri
    Amilia, Riska
    Marini, Rini
    Barokah, Sinta Nur
    Osman, Noor Diyana
    Dougherty, Geoff