Analysis of elastohydrodynamic lubrication in McKee-Farrar metal-on-metal hip joint replacement

被引:12
作者
Yew, A [1 ]
Udofia, I [1 ]
Jagatia, M [1 ]
Jin, ZM [1 ]
机构
[1] Univ Bradford, Sch Engn Design & Technol, Bradford BD7 1DP, W Yorkshire, England
关键词
elastohydrodynamic lubrication; metal-on-metal hip prosthesis; finite element;
D O I
10.1243/095441104322807721
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An elastohydrodynamic lubrication (EHL) analysis was carried out in this study for a typical McKee-Farrar metal-on-metal hip prosthesis under a simple steady state rotation. The finite element method was used initially to investigate the effect of the cement and bone on the predicted contact pressure distribution between the two articulating surfaces under dry conditions, and subsequently to determine the elastic deformation of both the femoral and the acetabular components required for the lubrication analysis. Both Reynolds equation and the elasticity equation were coupled and solved numerically using the finite difference method. Important features in reducing contact stresses and promoting fluid-film lubrication associated with the McKee-Farrar metal-on-metal hip implant were identified as the large femoral head and the thin acetabular cup. For the typical McKee-Farrar metal-on-metal hip prosthesis considered under typical walking conditions, an increase in the femoral head radius from 14 to 17.4 mm (for a fixed radial clearance of 79 mum) was shown to result in a 25 per cent decrease in the maximum dry contact pressure and a 60 per cent increase in the predicted minimum film thickness. Furthermore, the predicted maximum contact pressure considering both the cement and the bone was found to be decreased by about 80 per cent, while the minimum film thickness was predicted to be increased by 50 per cent. Despite a significant increase in the predicted minimum lubricating film thickness due to the large femoral head and the thin acetabular cup, a mixed lubrication regime was predicted for the McKee-Farrar metal-on-metal hip implant under estimated in vivo steady state walking conditions, depending on the surface roughness of the bearing surfaces. This clearly demonstrated the important influences of the material, design and manufacturing parameters on the tribological performance of these hard-on-hard hip prostheses. Furthermore, in the present contact mechanics analysis, the significant increase in the elasticity due to the relatively thin acetabular cup was not found to cause equatorial contact and gripping of the ball.
引用
收藏
页码:27 / 34
页数:8
相关论文
共 50 条
[21]   Contact mechanics of a novel metal-on-metal total hip replacement [J].
Besong, AA ;
Lee, R ;
Farrar, R ;
Jin, ZM .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2001, 215 (H6) :543-548
[22]   Different Loading and Motion Applied on Hip Simulators Affects the Lubrication of Metal-on-Metal Hip Implants [J].
Gao, Leiming ;
Yang, Peiran ;
Wang, Fengcai ;
Jin, Zhongmin .
ADVANCED TRIBOLOGY, 2009, :44-+
[23]   Long-Term Contact-Coupled Wear Prediction for Total Metal-on-Metal Hip Joint Replacement [J].
Harun, Muhamad Noor ;
Jin, Zhongmin ;
Wang, Fengchai ;
Fisher, John .
ADVANCED TRIBOLOGY, 2009, :835-+
[24]   Transient elas tohydro dynamic lubrication analysis of metal-on-metal hip implant under simulated walking conditions [J].
Liu, F ;
Jin, ZM ;
Hirt, F ;
Rieker, C ;
Roberts, P ;
Grigoris, P .
JOURNAL OF BIOMECHANICS, 2006, 39 (05) :905-914
[25]   Lubrication of metal-on-metal hip joints: The effect of protein content and load on film formation and wear [J].
Myant, C. ;
Underwood, R. ;
Fan, J. ;
Cann, P. M. .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2012, 6 :30-40
[26]   Numerical lubrication simulation of metal-on-metal artificial hip joint replacements: ball-in-socket model and ball-on-plane model [J].
Wang, W-Z ;
Wang, F-C ;
Jin, Z-M ;
Dowson, D. ;
Hu, Y-Z .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART J-JOURNAL OF ENGINEERING TRIBOLOGY, 2009, 223 (J7) :1073-1082
[27]   A hip joint simulator study of the performance of metal-on-metal joints - Part II: Design [J].
Dowson, D ;
Hardaker, C ;
Flett, M ;
Isaac, GH .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :124-130
[28]   Prediction of contact mechanics in metal-on-metal Total Hip Replacement for parametrically comprehensive designs and loads [J].
Donaldson, Finn E. ;
Nyman, Edward, Jr. ;
Coburn, James C. .
JOURNAL OF BIOMECHANICS, 2015, 48 (10) :1828-1835
[29]   Numerical solutions for the elastic deformation of spherical bearing surfaces of metal-on-metal hip joint implants [J].
Gao, L. M. ;
Meng, Q. E. ;
Wang, F. C. ;
Yang, P. R. ;
Jin, Z. M. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART J-JOURNAL OF ENGINEERING TRIBOLOGY, 2010, 224 (J8) :797-805
[30]   Analysis of Run-in-Stage Wear Behavior and Contact Mechanics of Metal-on-Metal Hip Joint Bearings with Different Radial Clearances [J].
Chen, Yan ;
Li, Yunping ;
Kurosu, Shingo ;
Meng, Qingen ;
Tang, Ning ;
Koizumi, Yuichiro ;
Chiba, Akihiko .
MATERIALS TRANSACTIONS, 2015, 56 (06) :826-834