Development of a portable therapeutic and high intensity ultrasound system for military, medical, and research use

被引:34
作者
Lewis, George K., Jr. [1 ]
Olbricht, William L. [2 ]
机构
[1] Cornell Univ, Dept Biomed Engn, Ithaca, NY 14853 USA
[2] Cornell Univ, Sch Chem & Biomol Engn, Ithaca, NY 14853 USA
基金
美国国家卫生研究院; 美国国家科学基金会;
关键词
biomedical transducers; ultrasonic transducers;
D O I
10.1063/1.3020704
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have developed a portable high power ultrasound system with a very low output impedance amplifier circuit (less than 0.3 Omega) that can transfer more than 90% of the energy from a battery supply to the ultrasound transducer. The system can deliver therapeutic acoustical energy waves at lower voltages than those in conventional ultrasound systems because energy losses owing to a mismatched impedance are eliminated. The system can produce acoustic power outputs over the therapeutic range (greater then 50 W) from a PZT-4, 1.54 MHz, and 0.75 in diameter piezoelectric ceramic. It is lightweight, portable, and powered by a rechargeable battery. The portable therapeutic ultrasound unit has the potential to replace "plug-in" medical systems and rf amplifiers used in research. The system is capable of field service on its internal battery, making it especially useful for military, ambulatory, and remote medical applications.
引用
收藏
页数:7
相关论文
共 22 条
[1]   Current status of HIFU and cryotherapy in prostate cancer - A review [J].
Aus, Gunnar .
EUROPEAN UROLOGY, 2006, 50 (05) :927-934
[2]   Physical mechanisms of the therapeutic effect of ultrasound - (A review) [J].
Bailey, MR ;
Khokhlova, VA ;
Sapozhnikov, OA ;
Kargl, SG ;
Crum, LA .
ACOUSTICAL PHYSICS, 2003, 49 (04) :369-388
[3]  
Cesario D, 2008, CONTEMP CARDIOL, P209, DOI 10.1007/978-1-59745-163-5_15
[4]  
CRUM LA, 2007, PROJECT REPORT NASA
[5]  
FRY FJ, 1986, P ULTR S, P1001
[6]  
HARR G, 2007, INT J HYPERTHER, V23, P89, DOI DOI 10.1080/02656730601186138
[7]   Arrays of multielement ultrasound applicators for interstitial hyperthermia [J].
Lee, RJ ;
Buchanan, M ;
Kleine, LJ ;
Hynynen, K .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1999, 46 (07) :880-890
[8]   Cost-effective broad-band electrical impedance spectroscopy measurement circuit and signal analysis for piezo-materials and ultrasound transducers [J].
Lewis, George K., Jr. ;
Lewis, George K., Sr. ;
Olbricht, William .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2008, 19 (10)
[9]  
LEWIS GK, 2008, P M AC, V2, P1
[10]  
LEWIS GK, 2008, P INT S THER ULTR