Corneal and epithelial thickness in keratoconus: A comparison of ultrasonic pachymetry Orbscan II, and optical coherence tomography

被引:97
作者
Haque, S [1 ]
Simpson, T [1 ]
Jones, L [1 ]
机构
[1] Univ Waterloo, Sch Optometry, Ctr Contact Lens Res, Waterloo, ON N2L 3G1, Canada
关键词
D O I
10.3928/1081-597X-20060501-11
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To compare corneal thickness measurements in individuals with keratoconus using optical coherence tomography (OCT), Orbscan II, and ultrasonic pachymetry and to measure epithelial and stromal thickness in these individuals using OCT. METHODS: Twenty individuals with keratoconus and 20 controls (without keratoconus) were enrolled. The Orbscan If was used to locate the steepest area of the cornea, which was taken to represent the cone apex. Each instrument was used to obtain four total corneal thickness measurements-from the cone apex, corneal center, mid-nasal, and mid-temporal cornea. Optical coherence tomography scans were analyzed to provide epithelial and stromal thickness readings. RESULTS: In individuals with keratoconus, mean central corneal thickness (CCT) measured by ultrasonic pachymetry Orbscan, and OCT was 494.2 +/- 50.0 mu m, 438.6 +/- 47.7 mu m, and 433.5 +/- 39.7 mu m, respectively. The central keratoconic cornea was 57.7 mu m thinner than the normal cornea (post-hoc P<.001). The cone apex was thinner than the central cornea (P<.001). Keratoconic epithelium was 48.2 +/- 5.5 mu m centrally and 42.1 +/- 4.5 mu m at the apex. Central keratoconic epithelium was 4.7 mu m thinner and central keratoconic stroma was 57.8 mu m thinner than the normal cornea (P<.001, respectively). Comparing instruments, Orbscan and OCT correlated in CCT measurement (r=0.890) and apical thickness (r=0.846). All instruments produced similar readings for mid-nasal and mid-temporal corneal thickness in participants with keratoconus (P>.05). CONCLUSIONS: Ultrasonic pachymetry produced the highest corneal thickness readings in the center and apex, compared to Orbscan 11 and OCT Centrally, the total cornea, epithelium, and stroma were thinner in individuals with keratoconus than in normal individuals.
引用
收藏
页码:486 / 493
页数:8
相关论文
共 36 条
  • [1] Keratoconus evaluation using the Orbscan Topography System
    Auffarth, GU
    Wang, L
    Völcker, HE
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (02) : 222 - 228
  • [2] Central corneal thickness measurement with a retinal optical coherence tomography device versus standard ultrasonic pachymetry
    Bechmann, M
    Thiel, MJ
    Neubauer, AS
    Ullrich, S
    Ludwig, K
    Kenyon, KR
    Ulbig, MW
    [J]. CORNEA, 2001, 20 (01) : 50 - 54
  • [3] Accuracy of Orbscan optical pachymetry in corneas with haze
    Boscia, F
    La Tegola, MG
    Alessio, G
    Sborgia, C
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (02) : 253 - 258
  • [4] Human corneal thickness and its impact on intraocular pressure measures: A review and meta-analysis approach
    Doughty, MJ
    Zaman, ML
    [J]. SURVEY OF OPHTHALMOLOGY, 2000, 44 (05) : 367 - 408
  • [5] Comparison of corneal pachymetry using ultrasound and Orbscan II
    Fakhry, MA
    Artola, A
    Belda, JL
    Ayala, MJ
    Alió, JL
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (02) : 248 - 252
  • [6] FERNANDEZ MJG, 2002, CLAO J, V28, P221
  • [7] Optical coherence tomography
    Fujimoto, JG
    [J]. COMPTES RENDUS DE L ACADEMIE DES SCIENCES SERIE IV PHYSIQUE ASTROPHYSIQUE, 2001, 2 (08): : 1099 - 1111
  • [8] Corneal pachymetry in normal and keratoconic eyes: Orbscan II versus ultrasound
    Gherghel, D
    Hosking, SL
    Mantry, S
    Banerjee, S
    Naroo, SA
    Shah, S
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (06) : 1272 - 1277
  • [9] Central and peripheral corneal thickness measurement with Orbscan II and topographical ultrasound pachymetry
    González-Méijome, JM
    Cerviño, A
    Yebra-Pimentel, E
    Parafita, MA
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (01) : 125 - 132
  • [10] CENTRAL AND PERIPHERAL CORNEAL THICKNESS IN KERATOCONUS AND NORMAL-PATIENT GROUPS
    GROMACKI, SJ
    BARR, JT
    [J]. OPTOMETRY AND VISION SCIENCE, 1994, 71 (07) : 437 - 441