What is the risk of stress risers for interprosthetic fractures of the femur? A biomechanical analysis

被引:55
作者
Lehmann, Wolfgang [1 ]
Rupprecht, Martin [1 ]
Nuechtern, Jacob [1 ]
Melzner, Daniel [1 ]
Sellenschloh, Kai [2 ]
Kolb, Jan [1 ]
Fensky, Florian [1 ]
Hoffmann, Michael [1 ]
Pueschel, Klaus [3 ]
Morlock, Michael [2 ]
Rueger, Johannes M. [1 ]
机构
[1] Univ Med Ctr Hamburg Eppendorf, Dept Trauma Hand & Reconstruct Surg, D-20246 Hamburg, Germany
[2] Hamburg Univ Technol, Biomech Sect, D-21079 Hamburg, Germany
[3] Univ Med Ctr Hamburg Eppendorf, Dept Legal Med, D-22529 Hamburg, Germany
关键词
FEMORAL FRACTURES; ARTHROPLASTY; FIXATION; PLATE;
D O I
10.1007/s00264-012-1697-0
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose Due to increasing life expectancy we see a rising number of joint replacements. Along with the proximal prosthesis in the femur, more and more people have a second implant on the distal ipsilateral side. This might be a retrograde nail or a locking plate to treat distal femur fractures or a constrained knee prosthesis in the case of severe arthrosis. All these constructs can lead to fractures between the implants. The goal of this study was to evaluate the risk of stress risers for interprosthetic fractures of the femur. Methods Thirty human cadaveric femurs were divided into five groups: (1) femurs with a prosthesis on the proximal side only, (2) hip prosthesis on the proximal end and a distal femur nail, (3) femurs with both a hip prosthesis and a constrained knee prosthesis, (4) femurs with a hip prosthesis on the proximal side and a 4.5-mm distal femur locking plate; the locking plate was 230 mm in length, with ten holes in the shaft, and (5) femurs with a proximal hip prosthesis and a 4.5-mm distal femur locking plate; the locking plate was 342 mm in length, with 16 holes in the shaft. Results Femurs with a hip prosthesis and knee prosthesis showed significantly higher required fracture force compared to femurs with a hip prosthesis and a distal retrograde nail. Femurs with a distal locking plate of either length showed a higher required fracture force than those with the retrograde nail. Conclusions The highest risk for a fracture in the femur with an existing hip prosthesis comes with a retrograde nail. A distal locking plate for the treatment of supracondylar fractures leads to a higher required fracture force. The implantation of a constrained knee prosthesis that is not loosened on the ipsilateral side does not increase the risk for a fracture.
引用
收藏
页码:2441 / 2446
页数:6
相关论文
共 14 条
  • [1] Appleton P, 2006, J BONE JOINT SURG BR, V88B, P1065, DOI 10.1302/0301-620X.88B8
  • [2] Periprosthetic fractures of the femur associated with hip arthroplasty
    Fink, B
    Fuerst, M
    Singer, J
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2005, 125 (07) : 433 - 442
  • [3] Treatment of Interprosthetic Fractures of the Femur
    Hou, Zhiyong
    Moore, Blake
    Bowen, Thomas R.
    Irgit, Kaan
    Matzko, Michelle E.
    Strohecker, Kent A.
    Smith, Wade R.
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2011, 71 (06): : 1715 - 1719
  • [4] Stress risers between two ipsilateral intramedullary stems - A finite-element and biomechanical analysis
    Iesaka, K
    Kummer, FJ
    Di Cesare, PE
    [J]. JOURNAL OF ARTHROPLASTY, 2005, 20 (03) : 386 - 391
  • [5] Interprosthetic fracture of the femoral shaft
    Kenny, P
    Rice, J
    Quinlan, W
    [J]. JOURNAL OF ARTHROPLASTY, 1998, 13 (03) : 361 - 364
  • [6] Results of a Second-Generation Constrained Condylar Prosthesis in Primary Total Knee Arthroplasty
    Lachiewicz, Paul F.
    Soileau, Elizabeth S.
    [J]. JOURNAL OF ARTHROPLASTY, 2011, 26 (08) : 1228 - 1231
  • [7] Biomechanical Evaluation of Peri- and Interprosthetic Fractures of the Femur
    Lehmann, Wolfgang
    Rupprecht, Martin
    Hellmers, Nils
    Sellenschloh, Kai
    Briem, Daniel
    Pueschel, Klaus
    Amling, Michael
    Morlock, Michael
    Rueger, Johannes Maria
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2010, 68 (06): : 1459 - 1463
  • [8] The locking attachment plate for proximal fixation of periprosthetic femur fractures-a biomechanical comparison of two techniques
    Lenz, Mark
    Windolf, Markus
    Mueckley, Thomas
    Hofmann, Gunther O.
    Wagner, Michael
    Richards, Robert G.
    Schwieger, Karsten
    Gueorguiev, Boyko
    [J]. INTERNATIONAL ORTHOPAEDICS, 2012, 36 (09) : 1915 - 1921
  • [9] Interprosthetic Femoral Fractures
    Mamczak, Christiaan N.
    Gardner, Michael J.
    Bolhofner, Brett
    Borrelli, Joseph, Jr.
    Streubel, Philipp N.
    Ricci, William M.
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2010, 24 (12) : 740 - 744
  • [10] Management of late periprosthetic femur fractures: a retrospective cohort of 72 patients
    Mukundan, Cibu
    Rayan, Faizal
    Kheir, Ehab
    Macdonald, D.
    [J]. INTERNATIONAL ORTHOPAEDICS, 2010, 34 (04) : 485 - 489