Gradient coil design considerations for iron core interventional magnets

被引:3
作者
Ersahin, A [1 ]
Bronskill, MJ [1 ]
Henkelman, RM [1 ]
Collick, B [1 ]
Hinks, RS [1 ]
机构
[1] Univ Toronto, Sunnybrook Hlth Sci Ctr, Dept Med Biophys, Toronto, ON M4N 3M5, Canada
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1998年 / 8卷 / 05期
关键词
interventional MRI; open-configuration magnet; gradient coils; Eddy currents; pole pieces; laminated poles;
D O I
10.1002/jmri.1880080521
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The requirements for access and imaging performance compete in the design of open-concept MR magnets and gradient coils. We conducted a theoretical and experimental investigation of gradient coil design using both solid and laminated pole piece construction to determine whether adequate eddy current control can be obtained without shielded gradient coils while maintaining good patient access and high gradient performance. Eddy currents, gradient characteristics, gradient efficiency, and magnet openness are compared and contrasted for various construction options based on a compact, .27 T iron yoke magnet. The resulting pole pieces and gradient coils have high efficiency for an interventional open-configuration magnet while taking up minimal space between the poles for improved patient access.
引用
收藏
页码:1145 / 1153
页数:9
相关论文
共 17 条
[1]  
[Anonymous], REV MAGN RESON MED
[2]   TEMPORAL AND SPATIAL-ANALYSIS OF FIELDS GENERATED BY EDDY CURRENTS IN SUPERCONDUCTING MAGNETS - OPTIMIZATION OF CORRECTIONS AND QUANTITATIVE CHARACTERIZATION OF MAGNET GRADIENT SYSTEMS [J].
BOESCH, C ;
GRUETTER, R ;
MARTIN, E .
MAGNETIC RESONANCE IN MEDICINE, 1991, 20 (02) :268-284
[3]  
Glover GH, 1987, US Patent Office, Patent No. 4698591
[4]  
HARDY CJ, 1995, P 3 ANN SCI M SOC MA, P489
[5]   USE OF IRON-POWDER MATERIAL FOR MR-IMAGING MAGNET POLE FACES [J].
HILLCOTTINGHAM, RJ ;
EASTHAM, JF ;
YOUNG, IR .
IEEE TRANSACTIONS ON MAGNETICS, 1995, 31 (06) :4074-4076
[6]   MR systems for image-guided therapy [J].
Hinks, RS ;
Bronskill, MJ ;
Kucharczyk, W ;
Bernstein, M ;
Collick, BD ;
Henkelman, RM .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (01) :19-25
[7]   INTENSITY ARTIFACTS IN MRI CAUSED BY GRADIENT SWITCHING IN AN ANIMAL-SIZE NMR MAGNET [J].
HUGHES, DG ;
ROBERTSON, S ;
ALLEN, PS .
MAGNETIC RESONANCE IN MEDICINE, 1992, 25 (01) :167-179
[8]   ANALYTICAL METHOD FOR THE COMPENSATION OF EDDY-CURRENT EFFECTS INDUCED BY PULSED MAGNETIC-FIELD GRADIENTS IN NMR SYSTEMS [J].
JEHENSON, P ;
WESTPHAL, M ;
SCHUFF, N .
JOURNAL OF MAGNETIC RESONANCE, 1990, 90 (02) :264-278
[9]  
KERR A, 1995, P 3 ANN SCI M SOC MA, P364
[10]   A TECHNIQUE FOR MR-GUIDED NEEDLE PLACEMENT [J].
LUFKIN, R ;
TERESI, L ;
CHIU, L ;
HANAFEE, W .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 151 (01) :193-196