Dose analysis of InCise 2 multi leaf collimator and IRIS-based stereotactic radiotherapy plans for brain and liver tumors

被引:0
作者
Ermis, Ekin [1 ,2 ]
Blatti-Moreno, Marcela [1 ,2 ]
Leiser, Dominic [1 ,2 ]
Cihoric, Nikola [1 ,2 ]
Schmidhalter, Daniel [2 ,3 ]
Henzen, Dominik [2 ,3 ]
Malthaner, Marco [2 ,3 ]
Hemmatazad, Hossein [1 ,2 ]
Tsikkinis, Alexandros [1 ,2 ]
Sermaxhaj, Burim [1 ,2 ]
Aebersold, Daniel M. [1 ,2 ]
Herrmann, Evelyn [1 ,2 ]
机构
[1] Bern Univ Hosp, Dept Radiat Oncol, Bern, Switzerland
[2] Univ Bern, Bern, Switzerland
[3] Bern Univ Hosp, Dept Radiat Oncol, Div Med Radiat Phys, Inselspital, Bern, Switzerland
关键词
cyberknife; multileaf collimator; brain metastasis; liver tumor; RADIOSURGERY; CYBERKNIFE; CONFORMITY; QUALITY; IMPACT;
D O I
10.1088/2057-1976/ab0beb
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Second-generation multi leaf collimator (MLC) has been introduced for the CyberKnife Robotic Radiosurgery System (Accuray Inc., Sunnyvale, USA) in 2016. The aim of the present study was to compare the plan quality between InCise 2 MLC and IRIS by analyzing dose parameters for small and large targets such as brain metastases and liver tumors, respectively. Ten brain metastases targets and 10 liver targets were selected. The following parameters were analyzed: coverage, dose distributions in planning target volume (PTV) and organs at risk (OAR), conformity index (CI), Nakamura CI (nCI), homogeneity index (HI), gradient index (GI), V12 Gy within brain, treatment time, number of beams, nodes and monitor units (MU). For brain patient group, CI and nCI were lower with IRIS plans (1.18 versus 1.26; p = 0.048 and 1.19 versus 1.28; p = 0.033, respectively), whereas GI values were higher in InCise 2 MLC plans (0.36 versus 0.33; p < 0.01). PTV mean doses were higher with IRIS plans (1.13 versus 1.16; p = 0.03). For liver tumors, InCise 2 MLC plans reached higher GI values (0.32 versus 0.29; p < 0.01). PTV minimum and mean doses were higher with IRIS plans (0.85 versus 0.93; p < 0.01 and 1.17 versus 1.2; p = 0.046, respectively). OAR's mean dose for bowel, left-right kidney and healthy liver tissue were lower with MLC plans (0.02 versus 0.04, 0.02 versus 0.05, 0.07 versus 0.10, and 0.20 versus 0.21; p < 0.01, respectively). Shorter delivery time was reached with InCise 2 MLC for both groups (38 versus 56 min, and 42 versus 56 min; p < 0.01, for brain and liver, respectively). Results of this study showed that second generation MLC plans for intracranial and liver targets were dosimetrically feasible. The main advantages of InCise 2 MLC are the improved dose gradient and delivery time.
引用
收藏
页数:8
相关论文
共 22 条
[1]  
Accuray, 2013, CYBERKNIFE M6TM SER
[2]   The cyberknife: A frameless robotic system for radiosurgery [J].
Adler, JR ;
Chang, SD ;
Murphy, MJ ;
Doty, J ;
Geis, P ;
Hancock, SL .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1997, 69 (1-4) :124-128
[3]   The design and physical characterization of a multileaf collimator for robotic radiosurgery [J].
Asmerom, G. ;
Bourne, D. ;
Chappelow, J. ;
Goggin, L. M. ;
Heitz, R. ;
Jordan, P. ;
Kilby, W. ;
Laing, T. ;
Maurer, C. R., Jr. ;
Noll, J. M. ;
Sayeh, S. ;
Weber, A. .
BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2016, 2 (01)
[4]  
Boda-Heggemann J, 2018, STRAHLENTHER ONKOL, V194, P414, DOI 10.1007/s00066-018-1271-9
[5]   The design, physical properties and clinical utility of an iris collimator for robotic radiosurgery [J].
Echner, G. G. ;
Kilby, W. ;
Lee, M. ;
Earnst, E. ;
Sayeh, S. ;
Schlaefer, A. ;
Rhein, B. ;
Dooley, J. R. ;
Lang, C. ;
Blanck, O. ;
Lessard, E. ;
Maurer, C. R., Jr. ;
Schlegel, W. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (18) :5359-5380
[6]   Characteristics and performance of the first commercial multileaf collimator for a robotic radiosurgery system [J].
Fuerweger, Christoph ;
Prins, Paulette ;
Coskan, Harun ;
Heijmen, Ben J. M. .
MEDICAL PHYSICS, 2016, 43 (05) :2063-2071
[7]  
Grimm Jimm, 2011, J Appl Clin Med Phys, V12, P3368
[8]   CyberKnife® Robotic Stereotactic Radiosurgery [J].
Heron, D. E. ;
Mintz, A. .
TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2010, 9 (05) :431-432
[9]   Dosimetric comparison between cone/Iris-based and InCise MLC-based CyberKnife plans for single and multiple brain metastases [J].
Jang, Si Young ;
Lalonde, Ron ;
Ozhasoglu, Cihat ;
Burton, Steven ;
Heron, Dwight ;
Huq, M. Saiful .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (05) :184-199
[10]   Dosimetric and delivery efficiency investigation for treating hepatic lesions with a MLC-equipped robotic radiosurgery radiotherapy combined system [J].
Jin, Lihui ;
Price, Robert A. ;
Wang, Lu ;
Meyer, Joshua ;
Fan, James ;
Ma, Chang Ming Charlie .
MEDICAL PHYSICS, 2016, 43 (02) :727-733