摘要
PURPOSE Occlusion time (OT) is an important factor in the treatment of pulmonary arteriovenous malformations (PAVMs) since it can lead to serious complications. The purpose of our study is to calculate the OT of Amplatzer vascular plug (AVP, St Jude Medical), and correlate it to the type of the device used (AVP or AVP 2) and the percent of device oversizing. Technical success rates and complications were also recorded. METHODS We retrospectively studied a total of 19 patients with 47 PAVMs who received percutaneous transcatheter embolization therapy using either AVP or AVP 2. We recorded the location, type, feeding artery diameter, AVP device used, and OT of each PAVM. We correlated the percent of device oversizing and the type of AVP with the OT. We also studied the rate of persistence of PAVM for both devices. RESULTS Forty-six (98%) of the PAVMs were simple. Device diameters ranged from 4.0-16.0 mm with device oversizing ranging between 14% and 120%. There was a statistically significant difference in the OT of AVP and AVP 2 (3 min 54 s vs. 5 min 30 s, P = 0.030). There was a weak positive correlation between OT and device oversizing for AVP (r= 0.246, P = 0.324) and AVP 2 (r= 0.261, P = 0.240). No major complications were identified. Immediate technical success rate was 100%. CONCLUSION The use of AVP 2, and increase in device oversizing were not associated with reduction in the OT of PAVMs. There was no reported difference in safety between the two devices, and no major complications were noted.
引用
收藏
页码:61 / 65
页数:5
相关论文
共 18 条
  • [1] Occlusion Time for Amplatzer Vascular Plug in the Management of Pulmonary Arteriovenous Malformations
    Aal, Ahmed K. Abdel
    Hamed, Maysoon F.
    Biosca, Roderick F.
    Saddekni, Souheil
    Raghuram, Karthikram
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 192 (03) : 793 - 799
  • [2] Transcatheter closure of pulmonary arteriovenous malformations with Amplatzer devices
    Beck, Avi
    Dagan, Tamir
    Matitiau, Avraham
    Bruckheimer, Elchanan
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2006, 67 (06) : 932 - 937
  • [3] Bilateral multiple pulmonary arteriovenous malformations: Endovascular treatment with the amplatzer vascular plug
    Cil, B
    Canyigit, M
    Ozkan, OS
    Pamuk, GA
    Dogan, R
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 17 (01) : 141 - 145
  • [4] Pulmonary arteriovenous malformations - A state of the art review
    Gossage, JR
    Kanj, G
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1998, 158 (02) : 643 - 661
  • [5] Embolization of pulmonary arteriovenous malformations using the Amplatzer vascular plug: successful treatment of 69 consecutive patients
    Hart, Jonathan L.
    Aldin, Zaid
    Braude, Philip
    Shovlin, Claire L.
    Jackson, James
    [J]. EUROPEAN RADIOLOGY, 2010, 20 (11) : 2663 - 2670
  • [6] Embolization of Pulmonary Arteriovenous Malformations with Amplatzer Vascular Plugs: Safety and Midterm Effectiveness
    Letourneau-Guillon, Laurent
    Faughnan, Marie E.
    Soulez, Gilles
    Giroux, Marie-France
    Oliva, Vincent L.
    Boucher, Louis-Martin
    Dubois, Josee
    Prabhudesai, Vikram
    Therasse, Eric
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2010, 21 (05) : 649 - 656
  • [7] Porstmann W., 1977, CURRENT CONCEPTS PED, P23, DOI [10.1007/978-1-4471-1288-4_3, DOI 10.1007/978-1-4471-1288-4_3]
  • [8] Pulmonary arteriovenous malformations treated with embolotherapy: Helical CT evaluation of long-term effectiveness after 2-21-year follow-up
    Remy-Jardin, M
    Dumont, P
    Brillet, PY
    Dupuis, P
    Duhamel, A
    Remy, J
    [J]. RADIOLOGY, 2006, 239 (02) : 576 - 585
  • [9] A new device for vascular embolization: Report on case of two pulmonary arteriovenous fistulas embolization using the Amplatzer Vascular Plug
    Rossi, Michele
    Rebonato, Alberto
    Greco, Laura
    Stefanini, Giulio
    Citone, Michele
    Speranza, Annnarita
    David, Vincenzo
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 29 (05) : 902 - 906
  • [10] Sacks David, 2003, J Vasc Interv Radiol, V14, pS199