The AutoQual ultrasound elastography method for quantitative assessment of lateral strain in post-rupture Achilles tendons

被引:18
作者
Brown, Phillip G. [1 ]
Alsousou, Joseph [3 ]
Cooper, Ashley [2 ]
Thompson, Mark S. [1 ,3 ]
Noble, J. Alison [1 ]
机构
[1] Univ Oxford, Dept Engn Sci, Inst Biomed Engn, Oxford, England
[2] Univ Oxford, Div Med Sci, Med Sci Teaching Ctr, Oxford, England
[3] Univ Oxford, Botnar Res Ctr, Nuffield Dept Orthopaed Rheumatol & Musculoskelet, Oxford, England
关键词
Ultrasound; Elastography; Lateral; Achilles; Tendon; Rupture; IN-VIVO; MOTION; DISPLACEMENT; MANAGEMENT;
D O I
10.1016/j.jbiomech.2013.07.044
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This paper presents the AutoQual elastography method: a novel algorithm that improves the quality of 2D displacement field calculation from ultrasound radio frequency (RE) sequences of acutely ruptured Achilles tendons to determine image-lateral strain fields and has potential use for ligaments and muscles. This method uses 2D bicubic spline interpolation of the RF signal, Quality Determined Search, Automatic Search Range and Adaptive Block Size components as a novel combination that is designed to improve continuity and decrease displacement field noise, especially in areas of low signal strength. We present a simple experiment for quantitatively comparing the AutoQual method to a multiscale (MS) elastography method from ultrasound RF sequences of a 5% agar phantom for rigid body motion and known lateral strain loads with speeds up to 5 mm/s. We finally present examples of four in vivo Achilles tendons in various damage states and with manual or artificially controlled passive flexion of the foot. Results show that the AutoQual method offers a substantial improvement on the MS method, achieving similar performance for rigid body tracking at all speeds, a lower normalized square error at all strains induced and a more continuous strain field at higher compression rates. AutoQual also showed a greater average normalized cross correlation for image blocks in the area of interest, a lower standard deviation of the strain field and a visually more acceptable point tracking for in vivo examples. This work demonstrates lateral ultrasound elastography which is robust to the complex passive motion of the Achilles and to various imaging artifacts associated with imaging tendon rupture. This method potentially has a wide clinical application for assessing in vivo strains in and hence mechanical function of any near skin surface tissues that are longitudinally loaded. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:2695 / 2700
页数:6
相关论文
共 21 条
  • [1] MECHANICAL BEHAVIOUR OF TENDON IN VITRO - A PRELIMINARY REPORT
    ABRAHAMS, M
    [J]. MEDICAL & BIOLOGICAL ENGINEERING, 1967, 5 (05): : 433 - +
  • [2] Alsousou J., 2012, J BONE JOINT SUR S36, V94, P6
  • [3] Low-level mechanical stimulation is sufficient to improve tendon healing in rats
    Andersson, Therese
    Eliasson, Pernilla
    Hammerman, Malin
    Sandberg, Olof
    Aspenberg, Per
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 2012, 113 (09) : 1398 - 1402
  • [4] Non-uniform displacement within the Achilles tendon during passive ankle joint motion
    Arndt, Anton
    Bengtsson, Ann-Sophie
    Peolsson, Michael
    Thorstensson, Alf
    Movin, Tomas
    [J]. KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2012, 20 (09) : 1868 - 1874
  • [5] Measurement of in vivo tendon function
    Bey, Michael J.
    Derwin, Kathleen A.
    [J]. JOURNAL OF SHOULDER AND ELBOW SURGERY, 2012, 21 (02) : 149 - 157
  • [6] Brown P, 2012, 2012 9TH IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI), P1104, DOI 10.1109/ISBI.2012.6235752
  • [7] OPERATIVE VERSUS NONOPERATIVE TREATMENT OF ACHILLES-TENDON RUPTURE - A PROSPECTIVE RANDOMIZED STUDY AND REVIEW OF THE LITERATURE
    CETTI, R
    CHRISTENSEN, SE
    EJSTED, R
    JENSEN, NM
    JORGENSEN, U
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 1993, 21 (06) : 791 - 799
  • [8] A Hybrid Displacement Estimation Method for Ultrasonic Elasticity Imaging
    Chen, Lujie
    Housden, R. James
    Treece, Graham M.
    Gee, Andrew H.
    Prager, Richard W.
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2010, 57 (04) : 866 - 882
  • [9] A quality-guided displacement tracking algorithm for ultrasonic elasticity imaging
    Chen, Lujie
    Treece, Graham M.
    Lindop, Joel E.
    Gee, Andrew H.
    Prager, Richard W.
    [J]. MEDICAL IMAGE ANALYSIS, 2009, 13 (02) : 286 - 296
  • [10] Tendon motion and strain patterns evaluated with two-dimensional ultrasound elastography
    Chernak, Laura A.
    Thelen, Darryl G.