Dynamic analysis of tapping mode atomic force microscope (AFM) for critical dimension measurement

被引:11
作者
Xiang, Wuweikai [1 ]
Tian, Yanling [2 ]
Liu, Xianping [2 ]
机构
[1] China Acad Engn Phys, Inst Syst Engn, Mianyang 621999, Sichuan, Peoples R China
[2] Univ Warwick, Sch Engn, Coventry CV4 7AL, W Midlands, England
来源
PRECISION ENGINEERING-JOURNAL OF THE INTERNATIONAL SOCIETIES FOR PRECISION ENGINEERING AND NANOTECHNOLOGY | 2020年 / 64卷
基金
欧盟地平线“2020”;
关键词
Tapping mode AFM; Dimension measurement; Transient dynamics; PHASE-CONTRAST; CANTILEVER; DEFORMATION; STIFFNESS; BEHAVIOR; MOTION; SPEED;
D O I
10.1016/j.precisioneng.2020.03.023
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Transient dynamics of tapping mode atomic force microscope (AFM) for critical dimension measurement are analyzed. A simplified nonlinear model of AFM is presented to describe the forced vibration of the micro cantilever-tip system with consideration of both contact and non-contact transient behavior for critical dimen-sion measurement. The governing motion equations of the AFM cantilever system are derived from the developed model. Based on the established dynamic model, motion state of the AFM cantilever system is calculated utilizing the method of averaging with the form of slow flow equations. Further analytical solutions are obtained to reveal the effects of critical parameters on the system dynamic performance. In addition, features of dynamic response of tapping mode AFM in critical dimension measurement are studied, where the effects of equivalent contact stiffness, quality factor and resonance frequency of cantilever on the system dynamic behavior are investigated. Contact behavior between the tip and sample is also analyzed and the frequency drift in contact phase is further explored. Influence of the interaction between the tip and sample on the subsequent non-contact phase is studied with regard to different parameters. The dependence of the minimum amplitude of tip displacement and maximum phase difference on the equivalent contact stiffness, quality factor and resonance frequency are investigated. This study brings further insights into the dynamic characteristics of tapping mode AFM for critical dimension measurement, and thus provides guidelines for the high fidelity tapping mode AFM scanning.
引用
收藏
页码:269 / 279
页数:11
相关论文
共 30 条
[1]  
Adams JD, 2016, NAT NANOTECHNOL, V11, P147, DOI [10.1038/nnano.2015.254, 10.1038/NNANO.2015.254]
[2]   A vibration analysis of a cracked micro-cantilever in an atomic force microscope by using transfer matrix method [J].
Dastjerdi, Shahriar ;
Abbasi, Mohammad .
ULTRAMICROSCOPY, 2019, 196 :33-39
[3]   Analysis of Dynamic Cantilever Behavior in Tapping Mode Atomic Force Microscopy [J].
Deng, Wenqi ;
Zhang, Guang-Ming ;
Murphy, Mark F. ;
Lilley, Francis ;
Harvey, David M. ;
Burton, David R. .
MICROSCOPY RESEARCH AND TECHNIQUE, 2015, 78 (10) :935-946
[4]   Intermittent impact dynamics of a cantilever scanning a surface at high speed [J].
Dey, S. ;
Kartik, V. .
JOURNAL OF APPLIED PHYSICS, 2018, 123 (12)
[5]  
Esmailzadeh E., 2018, ANAL METHODS NONLINE
[6]   Identification of nanoscale dissipation processes by dynamic atomic force microscopy [J].
Garcia, R. ;
Gomez, C. J. ;
Martinez, N. F. ;
Patil, S. ;
Dietz, C. ;
Magerle, R. .
PHYSICAL REVIEW LETTERS, 2006, 97 (01)
[7]   Dynamic atomic force microscopy methods [J].
García, R ;
Pérez, R .
SURFACE SCIENCE REPORTS, 2002, 47 (6-8) :197-301
[8]   Attractive and repulsive tip-sample interaction regimes in tapping-mode atomic force microscopy [J].
García, R ;
San Paulo, A .
PHYSICAL REVIEW B, 1999, 60 (07) :4961-4967
[9]   Robust controlled manipulation of nanoparticles using atomic force microscope [J].
Korayem, M. H. ;
Omidi, E. .
MICRO & NANO LETTERS, 2012, 7 (09) :927-931
[10]   Dynamic analysis of tapping-mode AFM considering capillary force interactions [J].
Korayem, M. H. ;
Kavousi, A. ;
Ebrahimi, N. .
SCIENTIA IRANICA, 2011, 18 (01) :121-129