Comparison of corneal astigmatism and axis location in cataract patients measured by total corneal power, automated keratometry, and simulated keratometry

被引:22
作者
Srivannaboon, Sabong [2 ]
Soeharnila [2 ,3 ]
Chirapapaisan, Chareenun [1 ]
Chonpimai, Pratuangsri [2 ]
机构
[1] Mahidol Univ, Siriraj Hosp, Dept Ophthalmol, SiLASIK Ctr, Bangkok 10700, Thailand
[2] Mahidol Univ, Siriraj Hosp, Fac Med, Bangkok 10700, Thailand
[3] Klin Mata Nusantara, Jakarta, Indonesia
关键词
SURGICALLY INDUCED ASTIGMATISM; INTRAOCULAR-LENS CALCULATION; DUAL SCHEIMPFLUG ANALYZER; ACCURACY; EYES; SURGERY; REPEATABILITY; TOPOGRAPHER; PENTACAM; VIDEOKERATOGRAPHY;
D O I
10.1016/j.jcrs.2012.07.024
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To compare the corneal astigmatism (magnitude and axis location) derived by total corneal power (TCP), automated keratometry, and simulated keratometry. SETTING: Siriraj Hospital, Mahidol University, Bangkok, Thailand. DESIGN: Prospective comparative study. METHODS: Eyes with previous ocular surgery or abnormalities were excluded. All patients were examined with the ARK 730A autokeratometer and the Galilei analyzer. The steepest and flattest corneal power along with the steepest axis of the TCP, automated keratometry, and simulated keratometry were recorded. Vector analysis (J0 and J45) was calculated. Analysis of variance with Bonferroni correction was performed for multiple comparisons. Outcome measures were the magnitude and axis location of astigmatism. RESULTS: One hundred eyes of 100 cataract patients were randomly selected. There was no statistically significant difference in the mean steepest axis between TCP (93.31 +/- 68.75 [SD]), automated keratometry (94.24 +/- 64.78), and simulated keratometry (92.42 +/- 64.30). However, the mean magnitude of astigmatism measured by TCP (1.23 +/- 0.75) was significantly higher than that measured by automated keratometry (0.93 +/- 0.68) (P = .01) but not than that measured by simulated keratometry (1.08 +/- 0.68) (P = .43); there was no statistically significant difference in J0 or J45. Twenty two (40%) of 54 eyes with more than 1.00 diopter of TCP astigmatism had more than 10 degrees of axis difference from automated keratometry. CONCLUSIONS: The magnitude of TCP astigmatism was higher than that of automated keratometry. The axis location was similar. However, there was more than 10 degrees of axis difference between automated keratometry and TCP in patients with high astigmatism. Financial Disclosure: No author has a financial or proprietary interest in any material or method mentioned. J Cataract Refract Surg 2012; 38:2088-2093 (C) 2012 ASCRS and ESCRS
引用
收藏
页码:2088 / 2093
页数:6
相关论文
共 38 条
  • [31] Comparison of Surgically-induced Astigmatism using Topography and Keratometry in Patients undergoing Phacoemulsification and Small Incision Cataract Surgery: A Prospective Interventional Study
    Shinde, Indrajit
    Joshi, Ajit
    Pandharikar, Aishwarya
    JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH, 2024, 18 (07) : NC1 - NC5
  • [32] Efficacy comparison between toric intraocular lens and aspheric intraocular lens plus steep-axis incision in cataract patients with low corneal astigmatism
    Liu, Ziyuan
    Zhou, Ren
    Xu, Ke
    Ding, Aihua
    Wang, Wei
    Wu, Tao
    Sun, Yanxiu
    ANNALS OF PALLIATIVE MEDICINE, 2021, 10 (03) : 2610 - 2619
  • [33] Comparison of grafted and non-grafted patients with corneal astigmatism undergoing cataract extraction with a toric intraocular lens implant
    Stewart, Christopher M.
    McAlister, James C.
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 38 (08) : 747 - 757
  • [34] Comparison of intraocular lens power calculation formulas with and without total keratometry and ray tracing in patients with previous myopic SMILE
    Elotmani, Anouar Meziane
    Messerschmidt-Roth, Anke
    Nehme, Alexandra
    Mueller, Hans-Helge
    Sekundo, Walter
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2023, 49 (05) : 467 - 473
  • [35] Impact of Total Corneal Astigmatism Estimated With the Abulafia-Koch Formula Versus Measured With a SS-OCT Biometer on the Refractive Outcomes of a Toric Intraocular Lens in Cataract Surgery
    Remignon, Charles-Henry
    Levron, Antoine
    Agard, Emilie
    Douma, Ikrame
    El-Chehab, Hussam
    Brunet, Olivier
    Sejournet, Lucas
    Billant, Jeremy
    Dot, Corinne
    JOURNAL OF REFRACTIVE SURGERY, 2023, 39 (03) : 171 - 179
  • [36] Comparison of Multicolored Spot Reflection Topographer and Scheimpflug-Placido System in Corneal Power and Astigmatism Measurements With Normal and Post-refractive Patients
    Lu, Weicong
    Miao, Yaxin
    Li, Yue
    Hu, Xiuli
    Hu, Qingjie
    Huang, Jinhai
    JOURNAL OF REFRACTIVE SURGERY, 2019, 35 (06) : 370 - +
  • [37] Comparison of two swept-source optical coherence tomography devices, a Scheimpflug camera system and a ray-tracing aberrometer in the measurement of corneal power in patients with cataract
    Ma, Shan
    Gao, Rongyu
    Sun, Jing
    Yang, Jun
    Wen, Kai
    Chen, Xiteng
    Zhao, Fangyu
    Xu, Xinyan
    Tian, Fang
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2024, 262 (05) : 1567 - 1578
  • [38] Comparison of two swept-source optical coherence tomography devices, a Scheimpflug camera system and a ray-tracing aberrometer in the measurement of corneal power in patients with cataract
    Shan Ma
    Rongyu Gao
    Jing Sun
    Jun Yang
    Kai Wen
    Xiteng Chen
    Fangyu Zhao
    Xinyan Xu
    Fang Tian
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2024, 262 : 1567 - 1578