Comparison of corneal astigmatism and axis location in cataract patients measured by total corneal power, automated keratometry, and simulated keratometry

被引:22
|
作者
Srivannaboon, Sabong [2 ]
Soeharnila [2 ,3 ]
Chirapapaisan, Chareenun [1 ]
Chonpimai, Pratuangsri [2 ]
机构
[1] Mahidol Univ, Siriraj Hosp, Dept Ophthalmol, SiLASIK Ctr, Bangkok 10700, Thailand
[2] Mahidol Univ, Siriraj Hosp, Fac Med, Bangkok 10700, Thailand
[3] Klin Mata Nusantara, Jakarta, Indonesia
来源
JOURNAL OF CATARACT AND REFRACTIVE SURGERY | 2012年 / 38卷 / 12期
关键词
SURGICALLY INDUCED ASTIGMATISM; INTRAOCULAR-LENS CALCULATION; DUAL SCHEIMPFLUG ANALYZER; ACCURACY; EYES; SURGERY; REPEATABILITY; TOPOGRAPHER; PENTACAM; VIDEOKERATOGRAPHY;
D O I
10.1016/j.jcrs.2012.07.024
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To compare the corneal astigmatism (magnitude and axis location) derived by total corneal power (TCP), automated keratometry, and simulated keratometry. SETTING: Siriraj Hospital, Mahidol University, Bangkok, Thailand. DESIGN: Prospective comparative study. METHODS: Eyes with previous ocular surgery or abnormalities were excluded. All patients were examined with the ARK 730A autokeratometer and the Galilei analyzer. The steepest and flattest corneal power along with the steepest axis of the TCP, automated keratometry, and simulated keratometry were recorded. Vector analysis (J0 and J45) was calculated. Analysis of variance with Bonferroni correction was performed for multiple comparisons. Outcome measures were the magnitude and axis location of astigmatism. RESULTS: One hundred eyes of 100 cataract patients were randomly selected. There was no statistically significant difference in the mean steepest axis between TCP (93.31 +/- 68.75 [SD]), automated keratometry (94.24 +/- 64.78), and simulated keratometry (92.42 +/- 64.30). However, the mean magnitude of astigmatism measured by TCP (1.23 +/- 0.75) was significantly higher than that measured by automated keratometry (0.93 +/- 0.68) (P = .01) but not than that measured by simulated keratometry (1.08 +/- 0.68) (P = .43); there was no statistically significant difference in J0 or J45. Twenty two (40%) of 54 eyes with more than 1.00 diopter of TCP astigmatism had more than 10 degrees of axis difference from automated keratometry. CONCLUSIONS: The magnitude of TCP astigmatism was higher than that of automated keratometry. The axis location was similar. However, there was more than 10 degrees of axis difference between automated keratometry and TCP in patients with high astigmatism. Financial Disclosure: No author has a financial or proprietary interest in any material or method mentioned. J Cataract Refract Surg 2012; 38:2088-2093 (C) 2012 ASCRS and ESCRS
引用
收藏
页码:2088 / 2093
页数:6
相关论文
共 50 条
  • [1] Assessing Total Keratometry Astigmatism, Simulated Keratometry, and Total Corneal Topographic Astigmatism Against Manifest Refractive Cylinder
    Alpins, Noel
    Ong, James K. Y.
    Stamatelatos, George
    JOURNAL OF REFRACTIVE SURGERY, 2021, 37 (03) : 198 - 201
  • [2] Repeatability of total Keratometry and standard Keratometry by the IOLMaster 700 and comparison to total corneal astigmatism by Scheimpflug imaging
    Savini, Giacomo
    Taroni, Leonardo
    Schiano-Lomoriello, Domenico
    Hoffer, Kenneth J.
    EYE, 2021, 35 (01) : 307 - 315
  • [3] Repeatability of total Keratometry and standard Keratometry by the IOLMaster 700 and comparison to total corneal astigmatism by Scheimpflug imaging
    Giacomo Savini
    Leonardo Taroni
    Domenico Schiano-Lomoriello
    Kenneth J. Hoffer
    Eye, 2021, 35 : 307 - 315
  • [4] Comparison of Corneal Power and Astigmatism between Simulated Keratometry, True Net Power, and Total Corneal Refractive Power before and after SMILE Surgery
    Qian, Yishan
    Liu, Yan
    Zhou, Xingtao
    Naidu, Rajeev Krishnan
    JOURNAL OF OPHTHALMOLOGY, 2017, 2017
  • [5] Comparing surgically induced astigmatism calculated by means of simulated keratometry versus total corneal refractive power
    Garzon, Nuria
    Rodriguez-Vallejo, Manuel
    Carmona, David
    Calvo-Sanz, Jorge A.
    Poyales, Francisco
    Palomino, Carlos
    Zato-Gomez de Liano, Miguel A.
    Fernandez, Joaquin
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2018, 28 (05) : 573 - 581
  • [6] Optimized keratometry and total corneal astigmatism for toric intraocular lens calculation
    Savini, Giacomo
    Naeser, Kristian
    Schiano-Lomoriello, Domenico
    Ducoli, Pietro
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2017, 43 (09): : 1140 - 1148
  • [7] Keratometry Evaluations With the Pentacam High Resolution in Comparison With the Automated Keratometry and Conventional Corneal Topography
    Modis, Laszlo, Jr.
    Szalai, Eszter
    Kolozsvari, Bence
    Nemeth, Gabor
    Vajas, Attila
    Berta, Andras
    CORNEA, 2012, 31 (01) : 36 - 41
  • [8] Comparison of Corneal Power Calculation by Standard Keratometry and Total Keratometry in Eyes With Previous Myopic FS-LASIK
    Lupardi, Enrico
    Taroni, Leonardo
    Hoffer, Kenneth J.
    Schiano-Lomoriello, Domenico
    Savini, Giacomo
    JOURNAL OF REFRACTIVE SURGERY, 2021, 37 (12) : 848 - 852
  • [9] Simulated Keratometry Versus Total Corneal Power by Ray Tracing: A Comparison in Prediction Accuracy of Intraocular Lens Power
    Savini, Giacomo
    Hoffer, Kenneth J.
    Lomoriello, Domenico Schiano
    Ducoli, Pietro
    CORNEA, 2017, 36 (11) : 1368 - 1372
  • [10] Accuracy of total corneal topographic astigmatism (CORT total) compared to total keratometry (TK)
    Alpins, Noel
    Ong, James
    Stamatelatos, George
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2019, 47 : 45 - 45