Auxiliary device for calibration of roller pumps: development and experimental results

被引:1
作者
Vieira, F. U., Jr. [1 ,2 ]
Dantas de Medeiros, J., Jr. [3 ]
Antunes, N. [2 ]
Vieira, R. W. [2 ]
Costa, E. T. [1 ,3 ]
机构
[1] State Univ Campinas UNICAMP, Ctr Biomed Engn, Sao Paulo, Brazil
[2] State Univ Campinas UNICAMP, Sch Med Sci, Sao Paulo, Brazil
[3] State Univ Campinas UNICAMP, Sch Elect & Comp Engn, Dept Biomed Engn, Sao Paulo, Brazil
来源
PERFUSION-UK | 2012年 / 27卷 / 06期
关键词
roller pump; hemolysis; calibration device; CARDIOPULMONARY BYPASS; CENTRIFUGAL PUMPS; RETROGRADE FLOW; DYNAMIC METHOD; HEMOLYSIS; SURGERY; BLOOD;
D O I
10.1177/0267659112453077
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Roller pumps are commonly used in surgeries involving extracorporeal circulation and its regulation may influence the hemolysis rate. Objective: To describe an auxiliary device for adjusting roller pumps using the dynamic calibration method and show preliminary results of its application in surgical procedures. Method:The device was tested in the laboratory environment with three disposable pressure transducers normally used in surgeries. In the operation room, tests were carried out with two groups of patients. Free hemoglobin plasma rates were measured,5 minutes after perfusion started and immediately before the passage of the residual volume of the pump. For Group 1 (n = 22), occlusion measurements were carried out using the drop rate method and for Group 2 (n = 18), measurements were carried out using the auxiliary device and 300 mmHg calibration (mean pressure for the dynamic calibration method). Results:The auxiliary calibration device has shown calibration results statistically equal to those obtained with a reference device in the laboratory environment. We have found less variability of hemolysis rates in the operating room, using the auxiliary device and we did not observe any influence of unbalanced rollers in the hemolysis rates.
引用
收藏
页码:512 / 519
页数:8
相关论文
共 17 条
  • [1] Asante-Siaw Julius, 2006, Interact Cardiovasc Thorac Surg, V5, P128, DOI 10.1510/icvts.2005.126961
  • [2] An investigational study of minimum rotational pump speed to avoid retrograde flow in three centrifugal blood pumps in a pediatric extracorporeal life support model
    Clark, Joseph B.
    Guan, Yulong
    McCoach, Robert
    Kunselman, Allen R.
    Myers, John L.
    Uendar, Akif
    [J]. PERFUSION-UK, 2011, 26 (03): : 185 - 190
  • [3] COOLEY DA, 1987, TEX HEART I J, V14, P113
  • [4] Drabkin DL, 1932, J BIOL CHEM, V98, P719
  • [5] Hydrodynamic performance and heat generation by centrifugal pumps
    Ganushchak, Y.
    Lichtenbelt, W. van Marken
    van der Nagel, T.
    de Jong, D. S.
    [J]. PERFUSION-UK, 2006, 21 (06): : 373 - 379
  • [6] Haemolysis during cardiopulmonary bypass:: an in vivo comparison of standard roller pumps, nonocclusive roller pumps and centrifugal pumps
    Hansbro, SD
    Sharpe, DAC
    Catchpole, R
    Welsh, KR
    Munsch, CM
    McGoldrick, JP
    Kay, PH
    [J]. PERFUSION-UK, 1999, 14 (01): : 3 - 10
  • [7] Clinical evaluation of setting pump occlusion by the dynamic method: effect on flow
    Mongero, L. B.
    Beck, J. R.
    Orr, T. W.
    Kroslowitz, R. M.
    Lee-Sensiba, K.
    Oz, M. C.
    [J]. PERFUSION-UK, 1998, 13 (05): : 360 - 368
  • [8] Superiority of centrifugal pump over roller pump in paediatric cardiac surgery: prospective randomised trial
    Morgan, IS
    Codispoti, M
    Sanger, K
    Mankad, PS
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1998, 13 (05) : 526 - 532
  • [9] Optimal Perfusion During Cardiopulmonary Bypass: An Evidence-Based Approach
    Murphy, Glenn S.
    Hessel, Eugene A., II
    Groom, Robert C.
    [J]. ANESTHESIA AND ANALGESIA, 2009, 108 (05) : 1394 - 1417
  • [10] Centrifugal versus roller head pumps for cardiopulmonary bypass: Effec neurpsychologic outcomes after coronary artery surgery
    Scott, DA
    Silbert, BS
    Doyle, TJ
    Blyth, C
    Borton, MC
    O'Brien, JL
    Horne, DJL
    [J]. JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2002, 16 (06) : 715 - 722