The Medtronic Melody® Transcatheter Pulmonary Valve Implanted at 24-mm Diameter-It Works

被引:42
作者
Cheatham, Sharon L. [1 ]
Holzer, Ralf J. [1 ]
Chisolm, Joanne L. [1 ]
Cheatham, John P. [1 ]
机构
[1] Nationwide Childrens Hosp, Ctr Heart, Columbus, OH 43215 USA
关键词
transcatheter valve replacement; conduit dysfunction; congenital heart disease; STENT FRACTURE; REPLACEMENT; CONDUIT;
D O I
10.1002/ccd.24821
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
ObjectivesWe report the Melody valve implanted and/or expanded to 24-mm diameter. BackgroundThe Medtronic Melody valve has been implanted up to 22 mm in the pulmonary position for over a decade. MethodsA retrospective chart review was performed on 82 patients who underwent Melody valve implant. Technical implant method, pre- and postimplant echocardiographic findings, and initial follow-up were reviewed. ResultsBetween 04/2008 and 12/2011, 13 Melody valves were successfully implanted in 11 patients, median age 35 years (range 16-61 years), in the pulmonary (bioprosthetic valve, right ventricle to pulmonary artery conduit, native valve) position (n=9), tricuspid position (bioprosthetic valve n=3), and aortic position (bioprosthetic valve n=1). Ten valves were delivered on a 24-mm balloon in balloon catheter and three were implanted using a 22-mm Ensemble balloon delivery system, followed by postdilation using a 24-mm x 2-cm Atlas balloon catheter. Postimplant, the median peak systolic gradient across the pulmonary valve was 7 mm Hg and median gradient across the tricuspid valve was 3 mm Hg. There was no change in gradient across the Melody valve in the aortic position where valve prosthesis-patient mismatch was present. Postimplant intracardiac echocardiography demonstrated none or mild valve regurgitation. No more than mild regurgitation was noted at a median follow-up of 9.5 months. ConclusionsThe Melody valve can be implanted at 24 mm in the stenotic/regurgitant bioprosthetic pulmonary, tricuspid, and aortic valve, dysfunctional right ventricle to pulmonary artery conduit, and the native right ventricular outflow tract, whereas the valve remains competent with only mild regurgitation. (c) 2013 Wiley Periodicals, Inc.
引用
收藏
页码:816 / 823
页数:8
相关论文
共 23 条
  • [1] A guide to fluoroscopic identification and design of bioprosthetic valves: A reference for valve-in-valve procedure
    Bapat, Vinayak
    Mydin, Izanne
    Chadalavada, Sucharitha
    Tehrani, Hassan
    Attia, Rizwan
    Thomas, Martyn
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 81 (05) : 853 - 861
  • [2] Percutaneous replacement of pulmonary valve in a right-ventricle to pulmonary-artery prosthetic conduit with valve dysfunction
    Bonhoeffer, P
    Boudjemline, Y
    Saliba, Z
    Merckx, J
    Aggoun, Y
    Bonnet, D
    Acar, P
    Le Bidois, J
    Sidi, D
    Kachaner, J
    [J]. LANCET, 2000, 356 (9239) : 1403 - 1405
  • [3] Transcatheter implantation of a bovine valve in pulmonary position - A lamb study
    Bonhoeffer, P
    Boudjemline, Y
    Saliba, Z
    Hausse, AO
    Aggoun, Y
    Bonnet, D
    Sidi, D
    Kachaner, J
    [J]. CIRCULATION, 2000, 102 (07) : 813 - 816
  • [4] Percutaneous insertion of the pulmonary valve
    Bonhoeffer, P
    Boudjemline, Y
    Qureshi, SA
    Le Bidois, J
    Iserin, L
    Acar, P
    Merckx, J
    Kachaner, J
    Sidi, D
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (10) : 1664 - 1669
  • [5] Transcutaneous Melody™ valve implantation in "tricuspid position" after a Fontan Bjork (RA-RV homograft) operation results in biventricular circulation
    Eicken, Andreas
    Fratz, Sohrab
    Hager, Alfred
    Vogt, Manfred
    Balling, Gunter
    Hess, John
    [J]. INTERNATIONAL JOURNAL OF CARDIOLOGY, 2010, 142 (03) : E45 - E47
  • [6] Percutaneous Re-Revalvulation of the Tricuspid Valve
    Gewillig, Marc
    Dubois, Christophe
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2011, 77 (05) : 692 - 695
  • [7] Transcatheter Tricuspid Valve Replacement With the Edwards SAPIEN Valve
    Kenny, Damien
    Hijazi, Ziyad M.
    Walsh, Kevin P.
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2011, 78 (02) : 267 - 270
  • [8] Immediate clinical and haemodynamic benefits of restoration of pulmonary valvar competence in patients with pulmonary hypertension
    Lurz, P.
    Nordmeyer, J.
    Coats, L.
    Taylor, A. M.
    Bonhoeffer, P.
    Schulze-Neick, I.
    [J]. HEART, 2009, 95 (08) : 646 - 650
  • [9] Percutaneous implantation of pulmonary valves for treatment of right ventricular outflow tract dysfunction
    Lurz, Philipp
    Bonhoeffer, Philipp
    [J]. CARDIOLOGY IN THE YOUNG, 2008, 18 (03) : 260 - 267
  • [10] Percutaneous pulmonary valve implantation impact of evolving technology and learning curve on clinical outcome
    Lurz, Philipp
    Coats, Louise
    Khambadkone, Sachin
    Nordmeyer, Johannes
    Boudjemline, Younes
    Schievano, Silvia
    Muthurangu, Vivek
    Lee, Twin Yen
    Parenzan, Giovanni
    Derrick, Graham
    Cullen, Seamus
    Walker, Fiona
    Tsang, Victor
    Deanfield, John
    Taylor, Andrew M.
    Bonhoeffer, Philipp
    [J]. CIRCULATION, 2008, 117 (15) : 1964 - 1972