Differential contribution of skin impedance and thoracic volume to transthoracic impedance during external defibrillation

被引:14
作者
Deakin, CD
Sado, DM
Petley, GW
Clewlow, F
机构
[1] Southampton Univ Hosp, Shackleton Dept Anaesthet, NHS Trust, Southampton SO16 6YD, Hants, England
[2] Southampton Univ Hosp, Dept Med Phys & Bioengn, NHS Trust, Southampton SO16 6YD, Hants, England
关键词
defibrillation; transthoracic impedance; paddle force;
D O I
10.1016/j.resuscitation.2003.10.001
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Two mechanisms by which firm external paddle force decreases transthoracic impedance (TTI) have been proposed. Decreased impedance at the paddle-skin interface has been assumed to be the primary mechanism, but expulsion of air from the lungs, reducing lung volume is also likely to contribute. The relative contribution of each mechanism is unknown. Methods and results: Thirty five intubated patients undergoing general anaesthesia for cardiac surgery were studied. TTI across external defibrillation paddles was measured as paddle force was increased to 12 kgf. Measurements were performed twice; once allowing the volume of the lungs to change and once with lung volume held at functional residual capacity. TTI with constant lung volume was significantly higher at all paddle forces (P < 0.001), confirming that a reduction in lung volume contributes to the decrease in TTI. At an optimal paddle force of 8 kg, the reduction in lung volume contributed to 16.2% of the overall decrease in TTI. Conclusion: The decrease in TTI seen with increasing external paddle force is due primarily to improved electrical contact at the paddle-skin inter-face, with a decrease in thoracic volume accounting for no more than 16% of the overall decrease at forces used clinically. (C) 2003 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:171 / 174
页数:4
相关论文
共 12 条
[1]  
[Anonymous], 2000, RESUSCITATION, V46, P109
[2]   LUNG COMPLIANCE FOLLOWING CARDIAC-ARREST [J].
DAVIS, K ;
JOHANNIGMAN, JA ;
JOHNSON, RC ;
BRANSON, RD .
ACADEMIC EMERGENCY MEDICINE, 1995, 2 (10) :874-878
[3]   What is the optimal paddle force during paediatric external defibrillation? [J].
Deakin, CD ;
Bennetts, SH ;
Petley, GW ;
Clewlow, F .
RESUSCITATION, 2003, 59 (01) :83-88
[4]   Determining the optimal paddle fore for external defibrillation [J].
Deakin, CD ;
Sado, DM ;
Petley, GW ;
Clewlow, F .
AMERICAN JOURNAL OF CARDIOLOGY, 2002, 90 (07) :812-+
[5]   Effects of positive end-expiratory pressure on transthoracic impedance - implications for defibrillation [J].
Deakin, CD ;
McLaren, RM ;
Petley, GW ;
Clewlow, F ;
Dalrymple-Hay, MJR .
RESUSCITATION, 1998, 37 (01) :9-12
[6]   THE OPTIMAL TECHNIQUE FOR ELECTRICAL CARDIOVERSION OF ATRIAL-FIBRILLATION [J].
EWY, GA .
CLINICAL CARDIOLOGY, 1994, 17 (02) :79-84
[7]   INFLUENCE OF VENTILATION PHASE ON TRANS-THORACIC IMPEDANCE AND DEFIBRILLATION EFFECTIVENESS [J].
EWY, GA ;
HELLMAN, DA ;
MCCLUNG, S ;
TAREN, D .
CRITICAL CARE MEDICINE, 1980, 8 (03) :164-166
[8]   RELATION BETWEEN TRANSCARDIAC AND TRANSTHORACIC CURRENT DURING DEFIBRILLATION IN HUMANS [J].
LERMAN, BB ;
DEALE, OC .
CIRCULATION RESEARCH, 1990, 67 (06) :1420-1426
[9]  
SIMA SJ, 1988, AM J CARDIOL, V62, P1048
[10]   Respiratory system compliance decreases after cardiopulmonary resuscitation and stomach inflation: impact of large and small tidal volumes on calculated peak airway pressure [J].
Wenzel, V ;
Idris, AH ;
Banner, MJ ;
Kubilis, PS ;
Band, R ;
Williams, JL ;
Lindner, KH .
RESUSCITATION, 1998, 38 (02) :113-118