Numerical Simulation of Ion Transport in a Nano-Electrospray Ion Source at Atmospheric Pressure

被引:12
作者
Wang, Wei [1 ,2 ]
Bajic, Steve [2 ]
John, Benzi [1 ]
Emerson, David R. [1 ]
机构
[1] Sci & Technol Facil Council, Sci Comp Dept, Daresbury Lab, Sci Tech Daresbury, Warrington WA4 4AD, Cheshire, England
[2] Waters Corp, Altrincham Rd, Wilmslow SK9 4AX, Cheshire, England
基金
英国工程与自然科学研究理事会; “创新英国”项目;
关键词
Nano-ESI-MS; Ion transport; Space-charge effect; Numerical gas dynamic simulation; IONIZATION-MASS-SPECTROMETRY; TRANSMISSION EFFICIENCY; SPHERICAL-PARTICLES; MS SYSTEM; INTERFACE; TRAJECTORIES; VACUUM; SIMION; FLOW; ELECTROPHORESIS;
D O I
10.1007/s13361-017-1863-0
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Understanding ion transport properties from the ion source to the mass spectrometer (MS) is essential for optimizing device performance. Numerical simulation helps in understanding of ion transport properties and, furthermore, facilitates instrument design. In contrast to previously reported numerical studies, ion transport simulations in a continuous injection mode whilst considering realistic space-charge effects have been carried out. The flow field was solved using Reynolds-averaged Navier-Stokes (RANS) equations, and a particle-in-cell (PIC) method was applied to solve a time-dependent electric field with local charge density. A series of ion transport simulations were carried out at different cone gas flow rates, ion source currents, and capillary voltages. A force evaluation analysis reveals that the electric force, the drag force, and the Brownian force are the three dominant forces acting on the ions. Both the experimental and simulation results indicate that cone gas flow rates of 250 slph (standard liter per hour) are important for high ion transmission efficiency, as higher cone gas flow rates reduce the ion signal significantly. The simulation results also show that the ion transmission efficiency reduces exponentially with an increased ion source current. Additionally, the ion loss due to space-charge effects has been found to be predominant at a higher ion source current, a lower capillary voltage, and a stronger cone gas counterflow. The interaction of the ion driving force, ion opposing force, and ion dispersion is discussed to illustrate ion transport mechanism in the ion source at atmospheric pressure.
引用
收藏
页码:600 / 612
页数:13
相关论文
共 58 条
[1]  
[Anonymous], 2014, THEORY GAS DISCHARGE
[2]  
[Anonymous], INT J ANAL CHEM
[3]  
Ansys Inc., 2017, ANSYS FLUENT US GUID
[4]   SIMION ion optics simulations at atmospheric pressure [J].
Appelhans, AD ;
Dahl, DA .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2005, 244 (01) :1-14
[5]   Measurement of external ion injection and trapping efficiency in the ion trap mass spectrometer and comparison with a predictive model [J].
Appelhans, AD ;
Dahl, DA .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2002, 216 (03) :269-284
[6]   Structural Characterization of Drug-like Compounds by Ion Mobility Mass Spectrometry: Comparison of Theoretical and Experimentally Derived Nitrogen Collision Cross Sections [J].
Campuzano, Iain ;
Bush, Matthew F. ;
Robinson, Carol V. ;
Beaumont, Claire ;
Richardson, Keith ;
Kim, Hyungjun ;
Kim, Hugh I. .
ANALYTICAL CHEMISTRY, 2012, 84 (02) :1026-1033
[7]   Practical implications of some recent studies in electrospray ionization fundamentals [J].
Cech, NB ;
Enke, CG .
MASS SPECTROMETRY REVIEWS, 2001, 20 (06) :362-387
[8]  
Clift R., 2005, Bubbles, drops, and particles
[9]  
Cole R.B., 2010, ELECTROSPRAY MALDI M
[10]   ATMOSPHERIC PRESSURE ION SOURCES [J].
Covey, Thomas R. ;
Thomson, Bruce A. ;
Schneider, Bradley B. .
MASS SPECTROMETRY REVIEWS, 2009, 28 (06) :870-897