Horizontal and anterior chamber diameter for phakic intraocular lens sizing

被引:5
作者
Oleszko, Adam A. [1 ]
Marek, Jaroslaw J. [1 ]
Muzyka-Wozniak, Maria M. [1 ]
机构
[1] Ophthalmol Clin Ctr SPEKTRUM, Wroclaw, Poland
关键词
angle-to-angle; anterior chamber diameter; biometry; horizontal corneal diameter; phakic intraocular lens; WHITE-TO-WHITE; IMPLANTABLE CONTACT-LENS; CILIARY SULCUS DIAMETER; CORNEAL DIAMETER; HIGH MYOPIA; AGREEMENT; PENTACAM; INTERCHANGEABILITY; REPEATABILITY; METAANALYSIS;
D O I
10.1111/cxo.13101
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Clinical relevance Proper selection of phakic intraocular lens diameter is necessary to avoid complications related with excessive vaulting after its implantation. Background Horizontal corneal diameter, measured as white-to-white distance, is one of the parameters used for phakic intraocular lens diameter calculation. Agreement was assessed between three parameters: white-to-white distance obtained with swept source optical coherence tomography, and white-to-white distance obtained with Scheimpflug camera and angle-to-angle parameter obtained with anterior segment optical coherence tomography. Methods This study included 55 eyes of 43 patients. The white-to-white distance was measured with two biometry techniques: swept source optical coherence tomography and Scheimpflug camera. The angle-to-angle was measured with anterior segment optical coherence tomography. Analysis of agreement was performed by the Bland-Altman method. For every patient, simulation of Visian Implantable Collamer Lens (ICL) sizing was performed using the Online Calculation and Ordering System with different white-to-white distance or angle-to-angle. Results Statistically significant differences were found between swept source optical coherence tomography biometer and anterior segment optical coherence tomography (p = 0.001) and between swept source optical coherence tomography and Scheimpflug camera (p < 0.001). However, there was a good correlation between swept source optical coherence tomography and Scheimpflug camera (intraclass correlation co-efficient = 0.623), with a shift toward higher white-to-white distance values on swept source optical coherence tomography. A relatively high correlation intraclass correlation co-efficient (0.772) and lack of statistically significant differences (p = 0.068) between anterior segment optical coherence tomography and Scheimpflug camera results were observed. Simulation of phakic intraocular lens sizing showed that swept source optical coherence tomography white-to-white distance should not be used interchangeably with Scheimpflug camera white-to-white distance or angle-to-angle. Conclusion White-to-white distance measured with swept source optical coherence tomography was significantly different from values obtained with other methods. Angle-to-angle may support ICL sizing, helping in verification of white-to-white distance values obtained with other devices.
引用
收藏
页码:62 / 68
页数:7
相关论文
共 26 条
[1]  
Benedetti S, 2005, J REFRACT SURG, V21, P116
[2]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[3]   Metaanalysis of cataract development after phakic intraocular lens surgery [J].
Chen, Li-Ju ;
Chang, Yun-Jau ;
Kuo, Jonathan C. ;
Rajagopal, Rama ;
Azar, Dimitri T. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (07) :1181-1200
[4]  
Choi KH, 2007, J REFRACT SURG, V23, P362
[5]   Assessment and Comparison of Anterior Chamber Dimensions Using Various Imaging Techniques [J].
Dinc, Umut Asli ;
Oncel, Banu ;
Gorgun, Ebru ;
Yenerel, Melda Nursal ;
Alimgil, Levent .
OPHTHALMIC SURGERY LASERS & IMAGING, 2010, 41 (01) :115-122
[6]   Device interchangeability on anterior chamber depth and white -to -white measurements: a thorough literature review [J].
Dominguez-Vicent, Alberto ;
Perez-Vives, Cari ;
Ferrer-Blasco, Teresa ;
Garcia-Lazaro, Santiago ;
Montes-Mico, Robert .
INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2016, 9 (07) :1057-1065
[7]   Improving accuracy of phakic intraocular lens sizing using high-frequency ultrasound biomicroscopy [J].
Dougherty, Paul J. ;
Rivera, Robert P. ;
Schneider, David ;
Lane, Stephen S. ;
Brown, David ;
Vukich, John .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (01) :13-18
[8]   Anterior Chamber Changes After Implantable Collamer Lens Implantation in High Myopia Using Pentacam: A Prospective Study [J].
Elmohamady M.N. ;
Abdelghaffar W. .
Ophthalmology and Therapy, 2017, 6 (2) :343-349
[9]   Agreement of white-to-white measurements with the IOLMaster 700, Atlas 9000, and Sirius systems [J].
Ferrer-Blasco, Teresa ;
Esteve-Taboada, Jose J. ;
Martinez-Albert, Noelia ;
Alfonso, Jose F. ;
Montes-Mico, Robert .
EXPERT REVIEW OF MEDICAL DEVICES, 2018, 15 (06) :453-459
[10]   Implantable contact lens for moderate to high myopia - Relationship of vaulting to cataract formation [J].
Gonvers, M ;
Bornet, C ;
Othenin-Girard, P .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (05) :918-924