Objective and subjective visual quality after implantation of all optic zone diffractive multifocal intraocular lenses: a prospective, case-control observational study

被引:31
作者
Chen, Tianyu [1 ]
Yu, Fang [1 ]
Lin, Huayou [1 ]
Zhao, Yinying [1 ]
Chang, Pingjun [1 ]
Lin, Lei [1 ]
Chen, Qi [1 ]
Zheng, Qian [1 ]
Zhao, Yun-e [1 ]
Lu, Fan [1 ]
Li, Jin [1 ]
机构
[1] Wenzhou Med Univ, Hosp Eye, Dept Cataract, Wenzhou 325000, Zhejiang, Peoples R China
基金
中国国家自然科学基金;
关键词
Clinical Trial; Optics and Refraction; Lens and zonules; DOUBLE-PASS SYSTEM; LASER VISION CORRECTION; PSEUDOPHAKIC EYES; IMAGE QUALITY; CHROMATIC ABERRATIONS; CLINICAL-EVALUATION; CATARACT-SURGERY; LIGHT SCATTER; PERFORMANCE; INSTRUMENT;
D O I
10.1136/bjophthalmol-2015-307135
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background and aims To evaluate the objective and subjective optical quality of all optic zone diffractive multifocal intraocular lens (IOL). Methods Fifty patients (50 eyes) having phacoemulsification and IOL implantation surgery were enrolled. 25 patients were implanted with all optic zone diffractive multifocal IOL and 25 patients with monofocal IOL. Objective optical quality parameters under 4mm pupil using Optical Quality Analysis System (OQAS), subjective visual acuity (VA) at 85cd/m(2) luminance and questionnaire concerning halo and visual function were assessed. Results The multifocal group gained similar objective optical outcomes, such as OQAS values at contrast 100%, 20%, 9%, modulation transfer function cut-off and Strehl ratio, as the monofocal group. Objective scatter index (OSI) was significantly better in the monofocal group. Distance corrected near VA, uncorrected near VA, distance corrected intermediate VA and uncorrected intermediate VA were significantly better in the multifocal group. The difference in uncorrected distance VA and best corrected distance VA between the monofocal and multifocal groups was not significant. Mild halos and glare were reported in both groups and a significant difference between the two groups was not observed; however, moderate glare and halos were reported only in the multifocal group (2/25). There was no significant difference between the two groups when visual function was compared. Conclusions All optic zone diffractive multifocal IOLs provided almost the same objective optical quality as monofocal IOLs did under 4mm pupil besides good all distance visual performance, but with greater OSI, indicating that implantation provides good optical quality in daytime but with more intraocular scatter light. Trial registration number NCT02234635, post-results.
引用
收藏
页码:1530 / 1535
页数:6
相关论文
共 41 条
  • [1] Visual Performance of Tecnis ZM900 Diffractive Multifocal IOL after 2500 Implants: A 3-Year Followup
    Akaishi, Leonardo
    Vaz, Rodrigo
    Vilella, Graziela
    Garcez, Rodrigo C.
    Tzelikis, Patrick F.
    [J]. JOURNAL OF OPHTHALMOLOGY, 2010, 2010
  • [2] Visual effect of the combined correction of spherical and longitudinal chromatic aberrations
    Artal, Pablo
    Manzanera, Silvestre
    Piers, Patricia
    Weeber, Henk
    [J]. OPTICS EXPRESS, 2010, 18 (02): : 1637 - 1648
  • [3] Evaluation of image quality after implantation of 2 diffractive multifocal intraocular lens models
    Castillo-Gomez, Alfredo
    Carmona-Gonzalez, David
    Maria Martinez-de-la-Casa, Jose
    Palomino-Bautista, Carlos
    Garcia-Feijoo, Julian
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (07) : 1244 - 1250
  • [4] One-year outcomes with new-generation multifocal intraocular lenses
    Cillino, Salvatore
    Casuccio, Alessandra
    Di Pace, Francesco
    Morreale, Raffaella
    Pillitteri, Francesco
    Cillino, Giovanni
    Lodato, Gaetano
    [J]. OPHTHALMOLOGY, 2008, 115 (09) : 1508 - 1516
  • [5] Cochener B, 2011, CLIN OPHTHALMOL, V5, P45, DOI 10.2147/OPTH.S14325
  • [6] Evaluation of the impact of light scatter from glistenings in pseudophakic eyes
    De Hoog, Edward
    Doraiswamy, Anand
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2014, 40 (01) : 95 - 103
  • [7] Comparison of the retinal image quality with a Hartmann-Shack wavefront sensor and a double-pass instrument
    Díaz-Doutón, F
    Benito, A
    Pujol, J
    Arjona, M
    Güell, JL
    Artal, P
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2006, 47 (04) : 1710 - 1716
  • [8] OPTICAL-QUALITY TESTING OF MONOFOCAL INTRAOCULAR-LENS IMPLANTS WITH A 3-MM AND A 4-MM APERTURE
    FAALAND, RW
    GROSSMAN, LW
    IGEL, DA
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1991, 17 (04) : 485 - 490
  • [9] Light Scatter and Disability Glare After Intraocular Lens Implantation
    Ferrer-Blasco, Teresa
    Montes-Mico, Robert
    Cervino, Alejandro
    Alfonso, Jose F.
    [J]. ARCHIVES OF OPHTHALMOLOGY, 2009, 127 (04) : 576 - 577
  • [10] Measurements of vision function and quality of life in patients with cataracts in southern India - Report of instrument development
    Fletcher, AE
    Ellwein, LB
    Selvaraj, S
    Vijaykumar, V
    Rahmathullah, R
    Thulasiraj, RD
    [J]. ARCHIVES OF OPHTHALMOLOGY, 1997, 115 (06) : 767 - 774