Intelligent 4D CT sequence scanning (i4DCT): First scanner prototype implementation and phantom measurements of automated breathing signal-guided 4D CT

被引:20
作者
Werner, Rene [1 ,2 ]
Sentker, Thilo [1 ,3 ]
Madesta, Frederic [1 ]
Schwarz, Annette [4 ,5 ]
Vornehm, Marc [4 ,5 ]
Gauer, Tobias [3 ]
Hofmann, Christian [5 ]
机构
[1] Univ Med Ctr Hamburg Eppendorf, Dept Computat Neurosci, D-20246 Hamburg, Germany
[2] Univ Med Ctr Hamburg Eppendorf, Ctr Biomed Artificial Intelligence bAIome, D-20246 Hamburg, Germany
[3] Univ Med Ctr Hamburg Eppendorf, Dept Radiotherapy & Radiooncol, D-20246 Hamburg, Germany
[4] Friedrich Alexander Univ Erlangen, D-91504 Erlangen, Germany
[5] Siemens Healthcare GmbH, D-91301 Forchheim, Germany
关键词
4D CT; motion artifacts; radiotherapy; STEREOTACTIC BODY RADIOTHERAPY; IMAGES; ACQUISITION; ARTIFACTS; 4D-CT;
D O I
10.1002/mp.14106
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Four-dimensional (4D) computed tomography (CT) imaging is an essential part of current 4D radiotherapy treatment planning workflows, but clinical 4D CT images are often affected by artifacts. The artifacts are mainly caused by breathing irregularity during data acquisition, which leads to projection data coverage issues for currently available commercial 4D CT protocols. It was proposed to improve projection data coverage by online respiratory signal analysis and signal-guided CT tube control, but related work was always theoretical and presented as pure in silico studies. The present work demonstrates a first CT prototype implementation along with respective phantom measurements for the recently introduced intelligent 4D CT (i4DCT) sequence scanning concept (). Methods Intelligent 4D CT was implemented on the Siemens SOMATOM go platform. Four-dimensional CT measurements were performed using the CIRS motion phantom. Motion curves were programmed to systematically vary from regular to very irregular, covering typical irregular patterns that are known to result in image artifacts using standard 4D CT imaging protocols. Corresponding measurements were performed using i4DCT and routine spiral 4D CT with similar imaging parameters (e.g., mAs setting and gantry rotation time, retrospective ten-phase reconstruction) to allow for a direct comparison of the image data. Results Following technological implementation of i4DCT on the clinical CT scanner platform, 4D CT motion artifacts were significantly reduced for all investigated levels of breathing irregularity when compared to routine spiral 4D CT scanning. Conclusions The present study confirms feasibility of fully automated respiratory signal-guided 4D CT scanning by means of a first implementation of i4DCT on a CT scanner. The measurements thereby support the conclusions of respective in silico studies and demonstrate that respiratory signal-guided 4D CT (here: i4DCT) is ready for integration into clinical CT scanners.
引用
收藏
页码:2408 / 2412
页数:5
相关论文
共 16 条
[1]  
[Anonymous], 2014, ATMOS CHEM PHYS DISC, V14, pC9360
[2]   Motion management strategies and technical issues associated with stereotactic body radiotherapy of thoracic and upper abdominal tumors: A review from NRG oncology [J].
Brandner, Edward D. ;
Chetty, Indrin J. ;
Giaddui, Tawfik G. ;
Xiao, Ying ;
Huq, M. Saiful .
MEDICAL PHYSICS, 2017, 44 (06) :2595-2612
[3]  
Chan MKH, 2015, STRAHLENTHER ONKOL, V191, P161, DOI 10.1007/s00066-014-0747-5
[4]   Respiratory regularity gated 4D CT acquisition: concepts and proof of principle [J].
Keall, P. J. ;
Vedam, S. S. ;
George, R. ;
Williamson, J. F. .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2007, 30 (03) :211-220
[5]   Respiration-Correlated Image Guidance Is the Most Important Radiotherapy Motion Management Strategy for Most Lung Cancer Patients [J].
Korreman, Stine ;
Persson, Gitte ;
Nygaard, Ditte ;
Brink, Carsten ;
Juhler-Nottrup, Trine .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 83 (04) :1338-1343
[6]   Prospective displacement and velocity-based cine 4D CT [J].
Langner, U. W. ;
Keall, P. J. .
MEDICAL PHYSICS, 2008, 35 (10) :4501-4512
[7]   An in silico performance characterization of respiratory motion guided 4DCT for high-quality low-dose lung cancer imaging [J].
Martin, Spencer ;
Brien, Ricky O' ;
Hofmann, Christian ;
Keall, Paul ;
Kipriditis, John .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (15)
[8]   New prospective 4D-CT for mitigating the effects of irregular respiratory motion [J].
Pan, Tinsu ;
Martin, Rachael M. ;
Luo, Dershan .
PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (15) :N350-N361
[9]   Deviations in delineated GTV caused by artefacts in 4DCT [J].
Persson, Gitte Fredberg ;
Nyaard, Ditte Eklund ;
Brink, Carsten ;
Jahn, Jonas Westberg ;
af Rosenschold, Per Munck ;
Specht, Lena ;
Korreman, Stine Sofia .
RADIOTHERAPY AND ONCOLOGY, 2010, 96 (01) :61-66
[10]   Influence of 4D CT Motion Artifacts on Correspondence Model-based 4D Dose Accumulation [J].
Sothmann, Thilo ;
Gauer, Tobias ;
Werner, Rene .
MEDICAL IMAGING 2018: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2018, 10576