A novel mouthpiece prevents bite injuries caused by intraoperative transcranial electric motor-evoked potential monitoring

被引:3
作者
Oshita, Kyoko [1 ]
Saeki, Noboru [1 ]
Kubo, Takayasu [2 ]
Abekura, Hitoshi [2 ]
Tanaka, Nobuhiro [3 ]
Kawamoto, Masashi [1 ]
机构
[1] Hiroshima Univ, Grad Sch Biomed & Hlth Sci, Dept Anesthesiol & Crit Care, Minami Ku, 1-2-3 Kasumi, Hiroshima, Japan
[2] Hiroshima Univ, Grad Sch Biomed & Hlth Sci, Dept Adv Prosthodont, Hiroshima, Japan
[3] Hiroshima Univ, Grad Sch Biomed & Hlth Sci, Dept Orthoped Surg, Hiroshima, Japan
关键词
Mouthpiece; Bite injuries; Tc-MEP monitoring; SPINE SURGERY; STIMULATION;
D O I
10.1007/s00540-016-2220-3
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Intraoperative transcranial motor-evoked potential monitoring causes contraction of the masseter muscles, which may cause injuries to the oral cavity and damage to the orotracheal tube. We developed a mouthpiece made from vinyl-silicone impression material to prevent these injuries. The purpose of this study was to examine its efficacy and safety. Twenty-two patients undergoing spinal surgery under transcranial motor-evoked potential monitoring were fitted with bespoke vinyl-silicone mouthpieces by dentists before surgery. On induction of general anesthesia and orotracheal intubation, the mouthpiece was attached to the upper and lower dental arches. A lateral cervical X-ray was taken at the end of surgery to examine the condition of the orotracheal tube. The incidence of endotracheal tube deformation was compared with an historic control group of 20 patients in whom a conventional gauze bite block had been previously used before induction of the mouthpiece. The oral cavity was examined by a dentist the day before surgery and 3 days postoperatively, and intraoral injuries were recorded. No endotracheal tube deformation was found in 22 patients fitted with the new mouthpiece. The incidence of tube deformation (none of 22 patients, 0 %) was significantly lower than in those who had been fitted with the gauze bite block (9 of 20 patients, 45.0 %; p < 0.001). Application of the mouthpiece resulted in no tongue or tooth injuries. A novel mouthpiece reduced the incidence of damage to the endotracheal tube caused by intraoperative transcranial motor-evoked potential monitoring.
引用
收藏
页码:850 / 854
页数:5
相关论文
共 50 条
  • [21] Intraoperative monitoring of somatosensory (SSEPs) and transcranial electric motor-evoked potentials (tce-MEPs) during surgical correction of neuromuscular scoliosis in patients with central or peripheral nervous system diseases
    F. Pastorelli
    M. Di Silvestre
    F. Vommaro
    E. Maredi
    A. Morigi
    M. R. Bacchin
    S. Bonarelli
    R. Plasmati
    R. Michelucci
    T. Greggi
    European Spine Journal, 2015, 24 : 931 - 936
  • [22] Low-dose droperidol suppresses transcranial electrical motor-evoked potential amplitude: a retrospective study
    Deguchi, Hiroyuki
    Furutani, Kenta
    Mitsuma, Yusuke
    Kamiya, Yoshinori
    Baba, Hiroshi
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2021, 35 (01) : 175 - 181
  • [23] Somatosensory- and motor-evoked potential monitoring during spine and spinal cord surgery
    P Costa
    A Bruno
    M Bonzanino
    F Massaro
    L Caruso
    I Vincenzo
    P Ciaramitaro
    E Montalenti
    Spinal Cord, 2007, 45 : 86 - 91
  • [24] Somatosensory- and motor-evoked potential monitoring during spine and spinal cord surgery
    Costa, P.
    Bruno, A.
    Bonzanino, M.
    Massaro, F.
    Caruso, L.
    Vincenzo, I.
    Ciaramitaro, P.
    Montalenti, E.
    SPINAL CORD, 2007, 45 (01) : 86 - 91
  • [25] Susceptibility of Transcranial Electric Motor-evoked Potentials to Varying Targeted Blood Levels of Dexmedetomidine during Spine Surgery
    Mahmoud, Mohamed
    Sadhasivam, Senthilkumar
    Salisbury, Shelia
    Nick, Todd G.
    Schnell, Beverly
    Sestokas, Anthony K.
    Wiggins, Cheryl
    Samuels, Paul
    Kabalin, Thomas
    McAuliffe, John
    ANESTHESIOLOGY, 2010, 112 (06) : 1364 - 1373
  • [26] Improving successful rate of transcranial electrical motor-evoked potentials monitoring during spinal surgery in young children
    Yang, Junlin
    Huang, Zifang
    Shu, Haihua
    Chen, Yuguang
    Sun, Xinrui
    Liu, Weifeng
    Dou, Yunling
    Xie, Chaofan
    Lin, Xiang
    Hu, Yong
    EUROPEAN SPINE JOURNAL, 2012, 21 (05) : 980 - 984
  • [27] Influence of Preoperative Motor Score and Patient Comorbidities on Transcranial Motor-Evoked Potential Acquisition in Intracranial Surgery: A Retrospective Cohort Study
    Chen, Adrian C.
    Shah, Harshal A.
    Vilaysom, Sabena
    Ryan, Casey
    Kruse, Aaron
    D'Amico, Randy S.
    Silverstein, Justin W.
    NEUROSURGERY, 2024, 95 (04) : 842 - 848
  • [28] The impact of several craniotomies on transcranial motor evoked potential monitoring during neurosurgery
    Tomio, Ryosuke
    Akiyama, Takenori
    Toda, Masahiro
    Ohira, Takayuki
    Yoshida, Kazunari
    JOURNAL OF NEUROSURGERY, 2017, 127 (03) : 543 - 552
  • [29] Understanding the effect of non-surgical factors in a transcranial motor-evoked potential alert: A retrospective cohort study
    Shigematsu, Hideki
    Yoshida, Go
    Kobayashi, Kazuyoshi
    Imagama, Shiro
    Ando, Muneharu
    Kawabata, Shigenori
    Yamada, Kei
    Kanchiku, Tsukasa
    Fujiwara, Yasushi
    Taniguchi, Shinichirou
    Iwasaki, Hiroshi
    Tadokoro, Nobuaki
    Takahashi, Masahito
    Wada, Kanichiro
    Yamamoto, Naoya
    Funaba, Masahiro
    Yasuda, Akimasa
    Ushirozako, Hiroki
    Tani, Toshikazu
    Matsuyama, Yukihiro
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2021, 26 (05) : 739 - 743
  • [30] Impact of Motor-Evoked Potential Monitoring on Facial Nerve Outcomes after Vestibular Schwannoma Resection
    Tawfik, Kareem O.
    Walters, Zoe A.
    Kohlberg, Gavriel D.
    Lipschitz, Noga
    Breen, Joseph T.
    O'Neal, Kelly
    Zuccarello, Mario
    Samy, Ravi N.
    ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 2019, 128 (01) : 56 - 61