REALIZATION OF A STANDARD RADIOMETER FOR MICROWAVE BRIGHTNESS-TEMPERATURE MEASUREMENTS TRACEABLE TO FUNDAMENTAL NOISE STANDARDS

被引:2
作者
Gu, Dazhen [1 ]
Houtz, Derek [1 ]
Randa, James [1 ]
Walker, David K. [1 ]
机构
[1] NIST, Boulder, CO 80305 USA
来源
2012 IEEE INTERNATIONAL GEOSCIENCE AND REMOTE SENSING SYMPOSIUM (IGARSS) | 2012年
关键词
Blackbody target; brightness temperature; illumination efficiency; microwave radiometer;
D O I
10.1109/IGARSS.2012.6350709
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We describe and demonstrate a standard radiometer for making microwave brightness-temperature measurements that are traceable to fundamental noise standards. The standard radiometer is based on a National Institute of Standards and Technology (NIST) waveguide radiometer for 18-26.5 GHz, fitted with an antenna to measure radiated power. The fraction of the antenna pattern subtended by the radiating target is determined by anechoic-chamber measurements in which we vary the temperature of the target and measure the received power. Sample measurement results with uncertainties are presented. The typical standard uncertainty for a brightness temperature of around 340 K is about 1 K. The approach should be extendable to other waveguide bands where NIST has radiometers and standards.
引用
收藏
页码:2941 / 2944
页数:4
相关论文
共 3 条
[1]  
Gu D., 2012, IEEE T GEOSCI REMOTE, V50
[2]  
Randa J., 2004, 6631 NIST
[3]  
Randa J., 2006, P IGARSS AUG, P3979