Neutron imaging with Medipix-2 chip and a coated sensor

被引:45
作者
Jakubek, J
Holy, T
Lehmann, E
Pospisil, S
Uher, J
Vacik, J
Vavrik, D
机构
[1] Czech Tech Univ, Inst Expt & Appl Phys, CZ-12800 Prague 2, Czech Republic
[2] Paul Scherrer Inst, CH-5232 Villigen, Switzerland
[3] Acad Sci Czech Republ, Inst Nucl Phys, CZ-25068 Rez, Czech Republic
关键词
neutron radiography; pixel detector; semiconductor detector;
D O I
10.1016/j.nima.2005.11.243
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Neutron transmission radiography can serve as a complementary diagnostic method to X-ray radiography. The hybrid pixel detector device Medipix-2 with a silicon sensor chip (256 x 256 square pixels of 55 mu m size) can be adapted for neutron detection by deposition of a converter layer onto its surface. Thermal neutrons are converted in the layer into secondary radiation which is detected by the pixel detector. Spatial resolution, detection efficiency and other properties were simulated for different converter materials and verified experimentally. The performance of the imager was compared with several types of neutron imaging systems. The results demonstrate superior properties of the Medipix-2 based neutron imager namely in terms of spatial resolution, linearity and dynamic range. The system is very promising in applications such as neutron micro-tomography. (c) 2005 Published by Elsevier B.V.
引用
收藏
页码:143 / 147
页数:5
相关论文
共 2 条
[1]  
JAKUBEK J, IEEE 2003 NUCL SCI S
[2]  
SMITH SW, 1999, DIGIT SIGNAL PROCESS, P425