Characterization of a distributed plasma ionization source (DPIS) for ion mobility spectrometry and mass spectrometry

被引:44
作者
Waltman, Melanie J. [1 ,2 ]
Dwivedi, Prabha [3 ]
Hill, Herbert H., Jr. [3 ]
Blanchard, William C. [4 ]
Ewing, Robert G. [1 ]
机构
[1] Pacific NW Natl Lab, Richland, WA 99354 USA
[2] New Mexico Inst Min & Technol, Dept Chem, Socorro, NM 87801 USA
[3] Washington State Univ, Dept Chem, Pullman, WA 99164 USA
[4] Blanchard & Co Inc, Phoenix, MD 21131 USA
关键词
Distributed plasma ionization source (DPIS); Atmospheric pressure ionization; Ion mobility spectrometry; Mass spectrometry; Corona discharge; Explosives; Drugs;
D O I
10.1016/j.talanta.2008.06.014
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A recently developed atmospheric pressure ionization source, a distributed plasma ionization source (DPIS), was characterized and compared to commonly used atmospheric pressure ionization sources with both mass spectrometry (MS) and ion mobility spectrometry (IMS). The source consisted of two electrodes of different sizes separated by a thin dielectric. Application of a high RF voltage across the electrodes generated plasma in air yielding both positive and negative ions. These reactant ions subsequently ionized the analyte vapors. The reactant ions generated were similar to those created in a conventional point-to-plane corona discharge ion source. The positive reactant ions generated by the source were mass identified as being solvated protons of general formula (H2O)(n)H+ with (H2O)(2) H+ as the most abundant reactant ion. The negative reactant ions produced were mass identified primarily as CO3-, NO3-, NO2-, O-3(-) and O-2(-) of various relative intensities. The predominant ion and relative ion ratios varied depending upon source construction and supporting gas flow rates. A few compounds including drugs, explosives and amines were selected to evaluate the new ionization source. The source was operated continuously for 3 months and although surface deterioration was observed visually, the source continued to produce ions at a rate similar that of the initial conditions. (c) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:249 / 255
页数:7
相关论文
共 23 条
[1]  
Asbury GR, 2000, TALANTA, V50, P1291
[2]  
Asbury GR, 2000, J MICROCOLUMN SEP, V12, P172, DOI 10.1002/(SICI)1520-667X(2000)12:3<172::AID-MCS7>3.0.CO
[3]  
2-W
[4]   A phenomenon connected with the discharge of electricity from pointed conductors [J].
Barnes, HT ;
Shaw, AN .
PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES A-CONTAINING PAPERS OF A MATHEMATICAL AND PHYSICAL CHARACTER, 1909, 82 (555) :336-350
[5]  
BLANCHARD WC, 2007, Patent No. 7157721
[6]  
BLANCHARD WC, 2006, P 15 INT C ION MOB S
[7]  
DINDOSOVA D, 1992, ACTA PHYS U COMENIAN, V33, P77
[8]   COMPARISON OF POSITIVE-IONS FORMED IN NICKEL-63 AND CORONA DISCHARGE ION SOURCES USING NITROGEN, ARGON, ISOBUTANE, AMMONIA AND NITRIC-OXIDE AS REAGENTS IN ATMOSPHERIC-PRESSURE IONIZATION MASS-SPECTROMETRY [J].
DZIDIC, I ;
CARROLL, DI ;
STILLWELL, RN ;
HORNING, EC .
ANALYTICAL CHEMISTRY, 1976, 48 (12) :1763-1768
[9]  
EWING RG, 2004, P 13 INT C MOB SPECT
[10]  
EWING RG, 2005, PITTSB C ORL FL MARC