Scalar Localization by Computed Tomography of Cochlear Implant Electrode Carriers Designed for Deep Insertion

被引:12
作者
Helbig, Silke [1 ]
Mack, Martin [2 ]
Schell, Boris [2 ]
Bratzke, Hansjuergen [3 ]
Stoever, Timo [1 ]
Helbig, Matthias [1 ]
机构
[1] Goethe Univ Frankfurt, Univ Clin, Dept Otorhinolaryngol, Frankfurt, Germany
[2] Goethe Univ Frankfurt, Univ Clin, Dept Radiol & Diagnost Intervent, Frankfurt, Germany
[3] Goethe Univ Frankfurt, Univ Clin, Dept Forens Med, Frankfurt, Germany
关键词
Cochlear implant; Inner ear; Histology; Computed tomography; Radiology; Temporal bone study; HEARING; ARRAY; PRESERVATION; POSITION; SYSTEM; PANEL;
D O I
10.1097/MAO.0b013e318259520c
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objectives: This study aimed to evaluate the possibility of predicting radiologically the scalar localization of a 31.5-mm-long, free-fitting electrode carrier for cochlear implantation, using conventional planar computed tomography. Study Design: A cross-sectional human temporal bone study was conducted. Setting: Twenty human temporal bones were acquired postmortem and implanted with 31.5- mm-long electrode carriers. Ten of these were implanted into the scala tympani using the round window approach, whereas the other 10 electrodes were inserted into the scala vestibuli by cochleostomy. Computed tomography was then performed, and 2 experienced blinded radiologists evaluated the intracochlear position of the array. Main Outcome Measure: The estimated position of the electrode carrier was described using a 5-point scale. After sectioning and histologic investigation, the results of the radiologic and histologic investigations were compared. Results: In 17 of 20 cases, it was possible to estimate the correct position of the electrode carrier within the basal turn of the cochlea by means of computed tomography. As the insertion angles widened beyond 360 degrees, it became increasing difficult for the radiologists to correctly determine the position of the electrode carrier. Conclusion: The comparison of our temporal bone experiment results with the computed tomography results revealed the difficulty of assessing the correct position of intracochlear electrodes. Scalar localization of deeply inserted electrode carriers cannot be precisely determined by means of computed tomography.
引用
收藏
页码:745 / 750
页数:6
相关论文
共 21 条
[1]   Quality control after insertion of the nucleus contour and contour advance electrode in adults [J].
Aschendorff, Antje ;
Kromeier, Jan ;
Klenzner, Thomas ;
Laszig, Roland .
EAR AND HEARING, 2007, 28 (02) :75S-79S
[2]   Flat-panel volume computed tomography for cochlear implant electrode array examination in isolated temporal bone specimens [J].
Bartling, Soenke H. ;
Gupta, Rajiv ;
Torkos, Attila ;
Dullin, Christian ;
Eckhardt, Grabbe ;
Lenarz, Thomas ;
Becker, Hartmut ;
Stöver, Timo .
OTOLOGY & NEUROTOLOGY, 2006, 27 (04) :491-498
[3]   The Internal Dimensions of the Cochlear Scalae With Special Reference to Cochlear Electrode Insertion Trauma [J].
Biedron, Slavomir ;
Prescher, Andreas ;
Ilgner, Justus ;
Westhofen, Martin .
OTOLOGY & NEUROTOLOGY, 2010, 31 (05) :731-737
[4]   Variational Anatomy of the Human Cochlea: Implications for Cochlear Implantation [J].
Erixon, Elsa ;
Hogstorp, Herman ;
Wadin, Karin ;
Rask-Andersen, Helge .
OTOLOGY & NEUROTOLOGY, 2009, 30 (01) :14-22
[5]   Role of Electrode Placement as a Contributor to Variability in Cochlear Implant Outcomes [J].
Finley, Charles C. ;
Holden, Timothy A. ;
Holden, Laura K. ;
Whiting, Bruce R. ;
Chole, Richard A. ;
Neely, Gail J. ;
Hullar, Timothy E. ;
Skinner, Margaret W. .
OTOLOGY & NEUROTOLOGY, 2008, 29 (07) :920-928
[6]   Hybrid 10 Clinical Trial Preliminary Results [J].
Gantz, Bruce J. ;
Hansen, Marlan R. ;
Turner, Christopher W. ;
Oleson, Jacob J. ;
Reiss, Lina A. ;
Parkinson, Aaron J. .
AUDIOLOGY AND NEURO-OTOLOGY, 2009, 14 :32-38
[7]   Hearing Preservation After Complete Cochlear Coverage in Cochlear Implantation With the Free-Fitting FLEXSOFT Electrode Carrier [J].
Helbig, Silke ;
Baumann, Uwe ;
Hey, Constanze ;
Helbig, Matthias .
OTOLOGY & NEUROTOLOGY, 2011, 32 (06) :973-979
[8]   Combined electric acoustic stimulation with the PULSARCI100 implant system using the FLEXEAS electrode array [J].
Helbig, Silke ;
Van de Heyning, Paul ;
Kiefer, Jan ;
Baumann, Uwe ;
Kleine-Punte, Andrea ;
Brockmeier, Hanna ;
Anderson, Ilona ;
Gstoettner, Wolfgang .
ACTA OTO-LARYNGOLOGICA, 2011, 131 (06) :585-595
[9]   The Value of Digital Volume Tomography in Assessing the Position of Cochlear Implant Arrays in Temporal Bone Specimens [J].
Kurzweg, Thiemo ;
Dalchow, Carsten V. ;
Bremke, Martin ;
Majdani, Omid ;
Kureck, Ingo ;
Knecht, Reinald ;
Werner, Jochen A. ;
Teymoortash, Afshin .
EAR AND HEARING, 2010, 31 (03) :413-419
[10]  