Changes of intraaortic balloon volume during pumping in a mock circulation system

被引:6
|
作者
Sakamoto, T
Akamatsu, H
Swartz, MT
Shoji, Y
Kazama, S
Fujiyoshi, K
Hashiguchi, Y
Suzuki, A
Miyashita, M
Arai, H
Suzuki, A
机构
[1] ST LOUIS UNIV,ST LOUIS,MO 63103
[2] KOHDEN TOKYO CORP,TOKYO,JAPAN
[3] AISIN HUMAN SYST CO LTD,NAGOYA,AICHI,JAPAN
关键词
intraaortic balloon pumping; mechanical circulatory support; diastolic augmentation; balloon volume;
D O I
10.1111/j.1525-1594.1996.tb04439.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The effects of hemodynamic changes on intraaortic balloon (IAB) volumes were studied experimentally using a helium tachometer in a mock circulatory system. The IAB volume decreased with increases in the heart rate and mean aortic pressure, but the degree of reduction in the IAB volume was different among 4 commercially available IAB drivers. Improvements in IAB consoles are needed to compliment the progress made in percutaneous techniques for IAB insertion resulting in smaller balloons.
引用
收藏
页码:275 / 277
页数:3
相关论文
共 50 条
  • [41] ALTERATIONS OF CORONARY COLLATERAL BLOOD-FLOW VELOCITY DURING INTRAAORTIC BALLOON PUMPING
    FLYNN, MS
    KERN, MJ
    DONOHUE, TJ
    AGUIRRE, FV
    BACH, RG
    CARACCIOLO, EA
    AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (16): : 1451 - 1455
  • [42] Effects of intraaortic balloon pumping on coronary and carotid flow during percutaneous cardiopulmonary support
    Geppert, A
    Frey, B
    Gabriel, H
    Kratochwill, C
    Siostrzonek, P
    ANNALS OF THORACIC SURGERY, 1996, 61 (05): : 1539 - 1541
  • [43] Mock test characteristic of intraaortic balloon pump using linear motor
    Yamada, H
    Wakiwaka, H
    Suzuki, K
    Nakagawa, H
    Fujii, N
    Matsuura, Y
    Fukunaga, S
    ARTIFICIAL ORGANS, 1996, 20 (11) : 1237 - 1241
  • [44] EFFECT OF INTRAAORTIC BALLOON PUMPS ON CEREBRAL-CIRCULATION
    GOMEZ, CR
    MCLAUGHLIN, JR
    NJEMANZE, PC
    STROKE, 1990, 21 (10) : 1512 - 1513
  • [45] Transesophageal atrial pacing as a trigger for intraaortic balloon pumping - Reply
    Joseph, DL
    ANESTHESIOLOGY, 1997, 86 (01) : 265 - 265
  • [46] INITIAL CLINICAL EXPERIENCE WITH INTRAAORTIC BALLOON PUMPING IN CARDIOGENIC SHOCK
    KANTROWITZ, A
    TJONNELAND, S
    FREED, PS
    PHILLIPS, SJ
    BUTNER, AN
    SHERMAN, JL
    JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1968, 203 (02) : 113 - +
  • [47] Intraaortic balloon pumping: why should we hurry up?
    A Macas
    A Mundinaite
    G Baksyte
    Critical Care, 13 (Suppl 1):
  • [48] PERCUTANEOUS INTRAAORTIC BALLOON PUMPING - 4 YEARS FROM TIBLISI
    BREGMAN, D
    ARTIFICIAL ORGANS, 1987, 11 (05) : 405 - 412
  • [49] COUNTERPULSATION AND INTRAAORTIC BALLOON PUMPING IN CARDIOGENIC SHOCK - CIRCULATORY DYNAMICS
    TALPINS, NL
    KRIPKE, DC
    GOETZ, RH
    ARCHIVES OF SURGERY, 1968, 97 (06) : 991 - &
  • [50] RECENT CLINICAL-EXPERIENCE WITH PERCUTANEOUS INTRAAORTIC BALLOON PUMPING
    BERG, GA
    REECE, IJ
    DAVIDSON, KG
    BAIN, WH
    LIFE SUPPORT SYSTEMS, 1986, 4 (03): : 249 - 256