Scanner calibration of a small animal PET camera based on continuous LSO crystals and flat panel PSPMTs

被引:40
作者
Benlloch, J. M.
Carrilero, V.
Gonzalez, A. J.
Catret, J.
Lerche, Ch. W.
Abellan, D.
de Quiros, F. Garcia
Gimenez, M.
Modia, J.
Sanchez, F.
Pavon, N.
Ros, A.
Martinez, J.
Sebastia, A.
机构
[1] IFIC, Inst Fis Corpuscular, E-46071 Valencia, Spain
[2] Univ Politecn Valencia, Dept Ingn Elect, E-46071 Valencia, Spain
[3] GEM IMAGING Co, R&D Div, Valencia, Spain
关键词
PET; continuous scintillator; PSPMT;
D O I
10.1016/j.nima.2006.10.020
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have constructed a small animal PET with four identical detector modules, each consisting of a continuous LYSO crystal attached to a Position Sensitive Photomultiplier Tube (PSPMT). The dimensions of the continuous crystal are 50 x 50 mm 2 and 10 mm thickness. The modules are separated 11 cm between each other in the scanner. In this paper we discuss the method used for the calibration of the camera for this special system with continuous detectors. We also present the preliminary values for the main performance parameters such as spatial and energy resolution, and sensitivity of the system. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:26 / 29
页数:4
相关论文
共 10 条
[1]  
ANGER HO, 1963, NUCLEONICS, V21, P56
[2]   Performance tests of a medical mini gamma-camera (summary) [J].
Benlloch, JM ;
Escat, B ;
Fernández, M ;
Giménez, M ;
Kadi-Hanifi, D ;
Lerche, CW ;
Pavón, N ;
Porras, E ;
Ruiz, JA ;
Sánchez, F ;
Alcañiz, M ;
Cerdá, J ;
Grau, V ;
Sebastià, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 504 (1-3) :232-233
[3]   A flat-panel-based mini gamma camera for lymph nodes studies [J].
Fernández, MM ;
Benlloch, JM ;
Cerdá, J ;
Escat, B ;
Giménez, EN ;
Giménez, M ;
Lerche, CW ;
Martínez, J ;
Pavón, N ;
Sánchez, F ;
Sebastià, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 527 (1-2) :92-96
[4]  
GIMENEZ EN, 2004, IEEE NUCL SCI S C RE
[5]   PERFORMANCE OF A POSITION-SENSITIVE SCINTILLATION DETECTOR [J].
KARP, JS ;
MUEHLLEHNER, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (07) :643-655
[6]   Depth of interaction detection with enhanced position-sensitive proportional resistor network [J].
Lerche, CW ;
Benlloch, JM ;
Sánchez, F ;
Pavón, N ;
Giménez, N ;
Fernández, M ;
Giménez, M ;
Sebastiá, A ;
Martínez, J ;
Mora, FJ .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 537 (1-2) :326-330
[7]   Depth of γ-ray interaction within continuous crystals from the width of its scintillation light-distribution [J].
Lerche, CW ;
Benlloch, JM ;
Sánchez, F ;
Pavón, N ;
Escat, B ;
Gimenez, EN ;
Fernández, M ;
Torres, I ;
Giménez, M ;
Sebastià, A ;
Martínez, J .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (03) :560-572
[8]   Portable mini gamma camera for medical applications [J].
Porras, E ;
Escat, B ;
Benlloch, JM ;
Kadi-Hanifi, D ;
López, S ;
Pavón, N ;
Ruiz, JA ;
Sánchez, F ;
Sebastià, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 486 (1-2) :186-190
[9]   Design and tests of a portable mini gamma camera [J].
Sánchez, F ;
Benlloch, JM ;
Escat, B ;
Pavón, N ;
Porras, E ;
Kadi-Hanifi, D ;
Ruiz, JA ;
Mora, FJ ;
Sebastià, A .
MEDICAL PHYSICS, 2004, 31 (06) :1384-1397
[10]   THE COUNTRATE PERFORMANCE OF THE VOLUME IMAGING PENN-PET SCANNER [J].
SMITH, RJ ;
KARP, JS ;
MUEHLLEHNER, G .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (04) :610-618