Disposable screen-printed electrode modified with bismuth-PSS composites as high sensitive sensor for cadmium and lead determination

被引:56
|
作者
Maria-Hormigos, Roberto [1 ]
Jesus Gismera, Ma [1 ]
Procopio, Jesus R. [1 ]
Teresa Sevilla, Ma [1 ]
机构
[1] Univ Autonoma Madrid, Dept Analyt Chem & Instrumental Anal, Fac Sci, Avda Francisco Tomas Y Valiente 7, E-28049 Madrid, Spain
关键词
Bismuth bulk modified electrodes; Screen-printed electrode; Polystyrene sulfonate composites; Differential pulse anodic stripping voltammetry (DPASV); Cd(II) and Pb(II) determination; ANODIC-STRIPPING VOLTAMMETRY; MERCURY FILM ELECTRODES; TRACE-METALS; POLYSTYRENE SULFONATE; CARBON ELECTRODE; GLASSY-CARBON; CD(II); PB(II); OXIDE; NANOSTRUCTURES;
D O I
10.1016/j.jelechem.2016.02.025
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Bismuth was incorporated in screen-printed carbon electrodes (SPCEs) modified with polystyrene sulfonate (PSS) and carbon nanopowder (CnP) composites. Different strategies, based on bulk approaches using bismuth oxide particles and bismuth (III) solutions, were assayed. The features of the modified electrodes were evaluated by differential pulse anodic stripping voltammetry (DPASV) of Pb(II) and Cd(II) solutions. The best results were obtained when bismuth was incorporated as bismuth oxide particles during the preparation of the PSS-CnP aqueous suspension used to modify the electrodes. Using this optimal modified sensor, the DPASV measurement conditions for Pb(II) and Cd(II) determination were optimized and their figures of merit were evaluated. Measurements were performed using two experimental approaches: the "drop method", by putting the test solution on the surface of the modified SPCE, and the "immersion method" performed by immersing the device in a stirred test solution. The limits of detection (S/N = 3) were 0.27 mu g L-1 (1.3 x 10(-9) M) for Pb(II) and 0.10 mu g L-1 (9.0 x 10(-10) M) for Cd(II), using the "drop method", and 0.029 mu g L-1 (1.4 x 10(-10) M) for Pb(II) and 0.012 mu g L-1 (1.1 x 10(-19) M) for Cd(II) employing the immersion procedure. The optimal bismuth modified SPCE was used for Pb(II) and Cd(II) determination in natural water samples with successful results. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:114 / 122
页数:9
相关论文
共 50 条
  • [41] A disposable electrochemical sensor for simultaneous determination of norepinephrine and serotonin in rat cerebrospinal fluid based on MWNTs-ZnO/chitosan composites modified screen-printed electrode
    Wang, Yuting
    Wang, Shi
    Tao, Li
    Min, Qing
    Xiang, Jin
    Wang, Qiman
    Xie, Jianmei
    Yue, Yuan
    Wu, Shuangchan
    Li, Xiaofei
    Ding, Hong
    BIOSENSORS & BIOELECTRONICS, 2015, 65 : 31 - 38
  • [42] Montmorillonite clay-modified disposable ink-jet-printed graphene electrode as a sensitive voltammetric sensor for the determination of cadmium(II) and lead(II)
    Léopoldine S. Guenang
    Liliane M. Dongmo
    Sherman L. Z. Jiokeng
    Arnaud T. Kamdem
    Giscard Doungmo
    Ignas K. Tonlé
    Victor Costa Bassetto
    Milica Jović
    Andreas Lesch
    Hubert Girault
    SN Applied Sciences, 2020, 2
  • [43] Montmorillonite clay-modified disposable ink-jet-printed graphene electrode as a sensitive voltammetric sensor for the determination of cadmium(II) and lead(II)
    Guenang, Leopoldine S.
    Dongmo, Liliane M.
    Jiokeng, Sherman L. Z.
    Kamdem, Arnaud T.
    Doungmo, Giscard
    Tonle, Ignas K.
    Bassetto, Victor Costa
    Jovic, Milica
    Lesch, Andreas
    Girault, Hubert
    SN APPLIED SCIENCES, 2020, 2 (03):
  • [44] Sensitive and stable monitoring of lead and cadmium in seawater using screen-printed electrode and electrochemical stripping analysis
    Gueell, Raquel
    Aragay, Gemma
    Fontas, Claudia
    Antico, Enriqueta
    Merkoci, Arben
    ANALYTICA CHIMICA ACTA, 2008, 627 (02) : 219 - 224
  • [45] Fabrication of a novel screen-printed carbon electrode based disposable sensor for sensitive determination of an endocrine disruptor BPSIP in environmental and biological matrices
    Singh, Ankit Kumar
    Agrahari, Shreanshi
    Gautam, Ravindra Kumar
    Tiwari, Ida
    MICROCHEMICAL JOURNAL, 2023, 193
  • [46] A screen-printed disposable enzyme electrode system for simultaneous determination of sucrose and glucose
    Hu, T
    Zhang, XE
    Zhang, ZP
    Chen, LQ
    ELECTROANALYSIS, 2000, 12 (11) : 868 - 870
  • [47] ELECTROCHEMICAL DETERMINATION OF METHOTREXATE AT A DISPOSABLE SCREEN-PRINTED ELECTRODE AND ITS APPLICATION STUDIES
    Wang, Shi
    Qi, Zheqiong
    Huang, Hongke
    Ding, Hong
    ANALYTICAL LETTERS, 2012, 45 (12) : 1658 - 1669
  • [48] Disposable and Sensitive Electrochemical Determination of Isoliquiritigenin in Licorice Using Screen-Printed Carbon Electrode Modified With Multi-Walled Carbon Nanotubes
    Tang, Weijia
    Zhu, Linna
    Li, Ran
    Adu-Frimpong, Michael
    Shen, Xin
    Li, Xiaoxiao
    Xu, Ximing
    Tong, Shanshan
    ANALYTICAL LETTERS, 2025, 58 (04) : 565 - 581
  • [49] Integrated three-electrode screen-printed sensor modified with bismuth film for voltammetric determination of thallium(I) at the ultratrace level
    Domanska, K.
    Tyszczuk-Rotko, K.
    ANALYTICA CHIMICA ACTA, 2018, 1036 : 16 - 25
  • [50] Electrochemical determination of tramadol using modified screen-printed electrode
    Mohammadi, Sayed Zia
    Jafari, Maedeh
    Jahani, Peyman Mohammadzadeh
    Ahmadi, Sayed Ali
    Mashayekh, Raana
    JOURNAL OF ELECTROCHEMICAL SCIENCE AND ENGINEERING, 2022, 12 (01): : 127 - 135