The Mesoporous Porphyrinic Zirconium Metal-Organic Framework for Pipette-Tip Solid-Phase Extraction of Mercury from Fish Samples Followed by Cold Vapor Atomic Absorption Spectrometric Determination
被引:53
作者:
Rezaei Kahkha, Mohammad Reza
论文数: 0引用数: 0
h-index: 0
机构:
Zabol Univ Med Sci, Dept Environm Hlth Engn, Zabol, IranZabol Univ Med Sci, Dept Environm Hlth Engn, Zabol, Iran
Rezaei Kahkha, Mohammad Reza
[1
]
论文数: 引用数:
h-index:
机构:
Daliran, Saba
[2
]
Oveisi, Ali Reza
论文数: 0引用数: 0
h-index: 0
机构:
Univ Zabol, Dept Chem, Fac Sci, Zabol, IranZabol Univ Med Sci, Dept Environm Hlth Engn, Zabol, Iran
Oveisi, Ali Reza
[3
]
Kaykhaii, Massoud
论文数: 0引用数: 0
h-index: 0
机构:
Univ Sistan & Baluchestan, Dept Chem, Fac Sci, Zahedan, IranZabol Univ Med Sci, Dept Environm Hlth Engn, Zabol, Iran
Solid-phase extraction;
Mesoporous MOF;
Porphyrin;
Zirconium metal organic framework (PCN-222/MOF-545);
Pipette-tip extraction;
Fish analysis;
ENVIRONMENTAL-SAMPLES;
TRACE MERCURY;
HUMAN PLASMA;
CATALYST;
REMOVAL;
PRECONCENTRATION;
SORBENT;
WATER;
ACID;
D O I:
10.1007/s12161-016-0786-x
中图分类号:
TS2 [食品工业];
学科分类号:
0832 ;
摘要:
In this study, the highly stable mesoporous porphyrinic zirconium metal-organic framework, namely PCN-222/MOF-545 (Zr-MOF), was prepared and used for pipette-tip solid-phase extraction of Hg(II). As a high-capacity sorbent, 4 mg of the Zr-MOF was placed into a conventional pipette tip and used, for the first time, for the fast extraction and preconcentration of mercury ions. For desorption, 50 mu L of 10% HCl was used by 15 repeated aspirating/dispensing cycles, and Hg ions in elusion were measured by a cold vapor atomic absorption spectrometer. Affecting parameters on extraction efficiency were studied, and optimum conditions were established as amount of sorbent 2 mg, pH was adjusted to 5.0, the eluting volume was 15 mu L, and extraction was performed on 1.8 mL of the sample. The optimal number of aspirating/dispensing cycles for extraction and desorption of analytes was found to be 10 and 15 cycles, respectively. The limit of detection of the method was found to be 20 ng L-1 with a relative standard deviation of ae<currency>3.1% (for seven replicate analyses of 20 mu g L-1 of mercury). Adsorption capacity and enrichment factor were 35.5 mg g(-1) and 120-fold, respectively. The proposed method was successfully applied for the determination of Hg(II) ions in fish samples.