Megavoltage CT imaging as a by-product of multileaf collimator leakage

被引:21
|
作者
Ruchala, KJ
Olivera, GH
Kapatoes, JM
Schloesser, EA
Reckwerdt, PJ
Mackie, TR
机构
[1] Univ Wisconsin, Dept Med Phys, Madison, WI 53706 USA
[2] Tomotherapy Inc, Middleton, WI 53562 USA
[3] Univ Wisconsin, Dept Human Oncol, Madison, WI 53706 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2000年 / 45卷 / 07期
关键词
D O I
10.1088/0031-9155/45/7/401
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In addition to their potential for the delivery of highly conformal radiation therapy treatments, tomotherapeutic treatments also Feature increased potential for verification. For example, megavoltage CT allows one to use the megavoltage linac to generate tomographic images of the patient in the treatment position. This is typically done before or after radiation therapy treatments. However, it is also possible to collect MVCT images entirely during the treatment itself. This process utilizes the leakage radiation through the closed leaves of the Nomos MIMiC MLC, along with slight inefficiencies in treatment delivery, to generate MVCT images during treatment that require neither additional time nor dose. The image quality is limited, yet sufficient to see a patient's external boundary, density differences over 8% for 25.0 mm objects and resolutions of 3.0 mm for high-contrast objects. Such images can potentially be viewed during treatment, used to flag additional CT immediately after the treatment and provide a representation of the patient's exact position during treatment for use with dose reconstruction.
引用
收藏
页码:N61 / N70
页数:10
相关论文
共 50 条
  • [1] Multileaf collimator end leaf leakage: implications for wide-field IMRT
    Hardcastle, N.
    Metcalfe, P.
    Ceylan, A.
    Williams, M. J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (21): : N493 - N504
  • [2] Leakage of the Siemens 160 MLC multileaf collimator on a dual energy linear accelerator
    Klueter, Sebastian
    Sroka-Perez, Gabriele
    Schubert, Kai
    Debus, Juergen
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (02): : N29 - N37
  • [3] Using multileaf collimator interleaf leakage to extract absolute spatial information from electronic portal imaging device images
    Gao, Zhanrong
    Szanto, Janos
    Gerig, Lee
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2007, 8 (01): : 1 - 9
  • [4] REAL-TIME TARGET POSITION ESTIMATION USING STEREOSCOPIC KILOVOLTAGE/MEGAVOLTAGE IMAGING AND EXTERNAL RESPIRATORY MONITORING FOR DYNAMIC MULTILEAF COLLIMATOR TRACKING
    Cho, Byungchul
    Poulsen, Per Rugaard
    Sawant, Amit
    Ruan, Dan
    Keall, Paul J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 79 (01): : 269 - 278
  • [5] Evaluation of a dual-level micro multileaf collimator in reducing dose undulation and leakage
    Liu, Y.
    Shi, C.
    Lin, L.
    Papanikolaou, N.
    MEDICAL PHYSICS, 2007, 34 (06) : 2449 - 2449
  • [6] Comparative analysis of several detectors for the measurement of radiation transmission and leakage from a multileaf collimator
    Larraga-Gutierrez, J. M.
    Galvan de la Cruz, O. O.
    Garcia-Garduno, O. A.
    Ballesteros-Zebadua, P.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (03): : 391 - 395
  • [7] Megavoltage image-based dynamic multileaf collimator tracking of a NiTi stent in porcine lungs on a linear accelerator
    Poulsen, Per R.
    Carl, Jesper
    Nielsen, Jane
    Nielsen, Martin S.
    Thomsen, Jakob B.
    Jensen, Henrik K.
    Kjærgaard, Benedict
    Zepernick, Peter R.
    Worm, Esben
    Fledelius, Walther
    Cho, Byungchul
    Sawant, Amit
    Ruan, Dan
    Keall, Paul J.
    International Journal of Radiation Oncology Biology Physics, 2012, 82 (02):
  • [8] Investigation of the impact of leaf design on the radiation leakage through a multileaf collimator for use in proton radiotherapy
    Ainsley, C.
    Avery, S.
    Maughan, R.
    McDonough, J.
    Bloch, P.
    Goulart, D.
    Ingram, M.
    MEDICAL PHYSICS, 2007, 34 (06) : 2590 - 2590
  • [9] Verification of multileaf collimator leaf positions using an electronic portal imaging device
    Samant, SS
    Zheng, W
    Parra, NA
    Chandler, J
    Gopal, A
    Wu, J
    Jain, J
    Zhu, YP
    Sontag, M
    MEDICAL PHYSICS, 2002, 29 (12) : 2900 - 2912
  • [10] In-phantom measurements of the bremsstrahlung dose for megavoltage electron beams incident on high Z materials for use in an electron multileaf collimator
    Reft, C.
    MEDICAL PHYSICS, 2007, 34 (06) : 2461 - 2461