Biomechanical Comparison of Costotransverse Process Screw Fixation and Pedicle Screw Fixation of the Upper Thoracic Spine

被引:6
|
作者
Little, Andrew S. [1 ]
Brasiliense, Leonardo B. C. [1 ]
Lazaro, Bruno C. R. [1 ]
Reyes, Phillip M. [1 ]
Dickman, Curtis A. [1 ]
Crawford, Neil R. [1 ]
机构
[1] St Josephs Hosp, Div Neurol Surg, Barrow Neurol Inst, Phoenix, AZ 85013 USA
关键词
Biomechanics; Costotransverse screw; Pedicle screw; Thoracic spine;
D O I
10.1227/01.NEU.0000350869.35779.05
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: To compare the biomechanics of costotransverse process screw fixation with those of pedicle screw fixation in a cadaveric model of the upper thoracic spine. METHODS: Ten human thoracic spines were instrumented across the T3-T4 segment with costotransverse and pedicle screws. Nonconstraining pure moments (maximum, 6.0 Nm) were applied to induce flexion, extension, lateral bending, and axial rotation. The range of motion, lax zone, and stiff zone were determined in each specimen in the normal state, after 3-column destabilization, and after instrumentation. After flexibility testing was completed, axial screw pull-out strength was assessed. RESULTS: In all directions of loading, both fixation techniques significantly decreased lax zone and range of motion at T3-T4 compared with the destabilized state (P <.001). During all loading modes except lateral bending, pedicle screw fixation allowed significantly less range of motion than costotransverse screw fixation. Pedicle screws provided 62% greater resistance to axial pull-out than costotransverse screws. CONCLUSION: The costotransverse screw technique seems to provide only moderately stiff fixation of the destabilized thoracic spine. Pedicle screw fixation seems to have more favorable biomechanical properties. These data suggest that the costotransverse process construct is better used as a salvage procedure rather than as a primary fixation strategy.
引用
收藏
页码:178 / 182
页数:5
相关论文
共 50 条
  • [1] A Biomechanical Comparison of a Novel Thoracic Screw Fixation Method: Transarticular Screw Fixation vs Traditional Pedicle Screw Fixation
    Yu, Yang
    Xie, Ning
    Song, Shaohua
    Zhang, Wei
    Guo, Qunfeng
    Ni, Bin
    Luo, Junsheng
    NEUROSURGERY, 2011, 69 : 141 - 145
  • [2] Pedicle screw fixation of the thoracic spine:: An in vitro biomechanical study on different configurations
    Deviren, V
    Acaroglu, E
    Lee, J
    Fujita, M
    Hu, S
    Lenke, LG
    Polly, D
    Kuklo, TR
    O'Brien, M
    Brumfield, D
    Puttlitz, CM
    SPINE, 2005, 30 (22) : 2530 - 2537
  • [3] Pedicle Screw Fixation of Thoracic Spine Fractures
    Bozik, M.
    Magala, M.
    Heger, T.
    Matejicka, D.
    Baka, J.
    Simko, R.
    ACTA CHIRURGIAE ORTHOPAEDICAE ET TRAUMATOLOGIAE CECHOSLOVACA, 2014, 81 (02) : 140 - 151
  • [4] Translaminar screw fixation in the upper thoracic spine
    Kretzer, Ryan M.
    Sciubba, Daniel M.
    Bagley, Carlos A.
    Wolinsky, Jean-Paul
    Gokaslan, Ziya L.
    Garonzik, Ira M.
    JOURNAL OF NEUROSURGERY-SPINE, 2006, 5 (06) : 527 - 533
  • [5] Efficacy of Pedicle Screw Fixation in Unstable Upper and Middle Thoracic Spine Fractures
    Ghasemi, Amir Abbas
    Ashoori, Soudabeh
    TRAUMA MONTHLY, 2016, 21 (01)
  • [6] Accuracy and safety of pedicle screw fixation in thoracic spine trauma
    Fisher, Charles G.
    Sahajpal, Vic
    Keynan, Ory
    Boyd, Michael
    Graeb, Douglas
    Bailey, Christopher
    Panagiotopoulos, Kostas
    Dvorak, Marcel F.
    JOURNAL OF NEUROSURGERY-SPINE, 2006, 5 (06) : 520 - 526
  • [7] A biomechanical comparison between cortical bone trajectory fixation and pedicle screw fixation
    Hiroki Oshino
    Toshihiko Sakakibara
    Tadashi Inaba
    Takamasa Yoshikawa
    Takaya Kato
    Yuichi Kasai
    Journal of Orthopaedic Surgery and Research, 10
  • [8] A biomechanical comparison between cortical bone trajectory fixation and pedicle screw fixation
    Oshino, Hiroki
    Sakakibara, Toshihiko
    Inaba, Tadashi
    Yoshikawa, Takamasa
    Kato, Takaya
    Kasai, Yuichi
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2015, 10
  • [9] Biomechanical evaluation of a dynamic pedicle screw fixation device
    Xu, HZ
    Wang, XY
    Chi, YL
    Zhu, QA
    Lin, Y
    Huang, QS
    Dai, LY
    CLINICAL BIOMECHANICS, 2006, 21 (04) : 330 - 336
  • [10] Biomechanical comparison of anterior lumbar screw-plate fixation versus posterior lumbar pedicle screw fixation
    Lie-hua Liu
    Cong-tao Guo
    Qiang Zhou
    Xiao-bing Pu
    Lei Song
    Hao-ming Wang
    Chen Zhao
    Shi-ming Cheng
    Yang-jun Lan
    Ling Liu
    Journal of Huazhong University of Science and Technology [Medical Sciences], 2014, 34 : 907 - 911