Evaluation scale to assess the accuracy of cuff-less blood pressure measuring devices

被引:9
|
作者
Yan, Iris Renfei [1 ]
Poon, Carmen Chung Yan [1 ]
Zhang, Yuan-Ting [1 ,2 ,3 ]
机构
[1] Chinese Univ Hong Kong, Dept Elect Engn, Joint Res Ctr Biomed Engn, Shatin, Hong Kong, Peoples R China
[2] Shenzhen Inst Adv Technol, Inst Biomed & Hlth Engn, Shenzhen, Peoples R China
[3] Chinese Acad Sci, Key Lab Biomed Informat & Hlth Engn, Beijing 100864, Peoples R China
关键词
blood pressure measurement; cuff-less device; error distribution; evaluation scale; standard; BRITISH-HYPERTENSION-SOCIETY; SELF-MEASUREMENT; MONITORING DEVICE; EUROPEAN-SOCIETY; MEDICAL-INSTRUMENTATION; OSCILLOMETRIC DEVICE; CLINICAL-EVALUATION; VALIDATION; ASSOCIATION; ADVANCEMENT;
D O I
10.1097/MBP.0b013e328330aea8
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective The call for early detection of hypertension and cardiac events creates a heavy demand for devices that can be used for blood pressure (BP) monitoring at home and in ambulatory settings. An emerging type of BP monitors without an occluding cuff has drawn great attentions for this application because it is comfortable and capable of providing continuous readings. For the development the cuff-less devices, it is crucial for the clinicians and engineers to joint efforts in establishing an evaluation standard. Methods This study attempts to contribute to its initiation in two ways. First, a new distribution model for measurement differences between the test device and the reference was proposed. We verified the model using evaluation results from 40 devices, of which 80% of the American Association for the Advancement of Medical Instrumentation and British Hypertension Society results were in agreement, as compared with 50%, if the original normal model was used. We further tested a cuff-less device on 85 patients for 999 datasets and that the differences between the proposed distribution that of the device were nonsignificant for systolic BP measurements (Kolmogorov-Smimov = 0.036, P = 0.15). Second, some evaluation scales were studied for their capability to assess the accuracy of cuff-less devices. For mean absolute difference, a map was developed to relate it with the criteria of American Association for the Advancement of Medical Instrumentation, British Hypertension Society, and European Society of Hypertension protocols, on the basis of the proposed distribution model; for mean absolute percentage difference, it is prominent in evaluating devices that have measurement errors often increasing with Bp, which is an issue has not been fully explored in existing standards. Conclusion This study focused on the statistical aspect of establishing standard to assess the accuracy of cuff-less BP measuring devices. The results of our study on the validation reports of various cuff-based devices and an experimental study on a cuff-less device showed that the t4 distribution is better than the normal distribution in portraying the underlying error distribution of both kinds of devices. Moreover, based on both the theoretical and experimental studies, mean absolute difference or mean absolute percentage difference is recommended as continuous scale to assess the accuracy of cuff-less devices for their own distinctive advantages. Blood Press Monit 14:257-267 (C) 2009 Wolters Kluwer Health | Lippincott Williams & Wilkins.
引用
收藏
页码:257 / 267
页数:11
相关论文
共 50 条
  • [41] Wireless body sensor network for continuous cuff-less blood pressure monitoring
    Espina, Javier
    Falck, Thomas
    Muehlsteff, Jens
    Aubert, Xavier
    2006 3RD IEEE/EMBS INTERNATIONAL SUMMER SCHOOL ON MEDICAL DEVICES AND BIOSENSORS, 2006, : 11 - +
  • [42] PPG Sensor Contact Pressure Should Be Taken Into Account for Cuff-Less Blood Pressure Measurement
    Chandrasekhar, Anand
    Yavarimanesh, Mohammad
    Natarajan, Keerthana
    Hahn, Jin-Oh
    Mukkamala, Ramakrishna
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2020, 67 (11) : 3134 - 3140
  • [43] INCREMENTAL VALUE OF A CUFF-LESS BLOOD MEASUREMENT DEVICE COMPARED TO CLINIC BLOOD PRESSURE MEASUREMENT
    Vischer, Annina
    Hotz, Leana
    Derendinger, Felicia
    Krisai, Philipp
    Socrates, Thenral
    Schumacher, Christina
    Mayr, Michael
    Burkard, Thilo
    JOURNAL OF HYPERTENSION, 2024, 42
  • [44] Cuff-Less High-Accuracy Calibration-Free Blood Pressure Estimation Using Pulse Transit Time
    Kachuee, Mohamad
    Kiani, Mohammad Mahdi
    Mohammadzade, Hoda
    Shabany, Mahdi
    2015 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS (ISCAS), 2015, : 1006 - 1009
  • [45] FEASIBILITY OF WATCH-TYPE CUFF-LESS BLOOD PRESSURE MEASUREMENT DEVICE IN DAILY BLOOD PRESSURE MONITORING
    Lee, Youngro
    Seo, Jongmo
    Lee, Haeyoung
    JOURNAL OF HYPERTENSION, 2024, 42
  • [46] INCREMENTAL VALUE OF A CUFF-LESS BLOOD MEASUREMENT DEVICE COMPARED TO CLINIC BLOOD PRESSURE MEASUREMENT
    Vischer, Annina
    Leana, Hotz
    Felicia, Derendinger
    Krisai, Philipp
    Thenral, Socrates
    Christina, Schumacher
    Mayr, Michael
    Burkard, Thilo
    SWISS MEDICAL WEEKLY, 2024, 154 : 82S - 83S
  • [47] Cuff-less blood pressure estimation using hetero-core optical fibers
    Koyama, Yuya
    Kamada, Hiroki
    Watanabe, Kazuhiro
    Nishiyama, Michiko
    ELECTRONICS AND COMMUNICATIONS IN JAPAN, 2024, 107 (01)
  • [48] Cuff-less blood pressure measurement using fingertip photoplethysmogram signals and physiological characteristics
    Yang, S.
    Zhang, Y.
    Cho, Siu-Yeung
    Morgan, Stephen P.
    Correia, Ricardo
    Wen, L.
    OPTICS IN HEALTH CARE AND BIOMEDICAL OPTICS VIII, 2018, 10820
  • [49] Real-time, Cuff-less and Non-invasive Blood Pressure Monitoring
    Abolhasani, Alireza
    Mousazadeh, Morteza
    Khoei, Abdollah
    INFORMACIJE MIDEM-JOURNAL OF MICROELECTRONICS ELECTRONIC COMPONENTS AND MATERIALS, 2020, 50 (02): : 87 - 103
  • [50] Wearable Cuff-less PTT-based System for Overnight Blood Pressure Monitoring
    Zheng, Yali
    Yan, Bryan P.
    Zhang, Yuanting
    Yu, C. M.
    Poon, Carmen C. Y.
    2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2013, : 6103 - 6106