Utilization of Anion-exchange Guard Column as an Ion Chromatographic Column of Anions Including Application to Simultaneous Separation of Anions and Cations

被引:4
|
作者
Yoshii, Sakura [1 ]
Mori, Masanobu [2 ]
Kozaki, Daisuke [2 ]
Hosokawa, Takayuki [2 ]
Itabashi, Hideyuki [1 ]
机构
[1] Gunma Univ, Grad Sch Sci & Technol, 1-5-1 Tenjin Cho, Kiryu, Gunma 3768515, Japan
[2] Kochi Univ, Fac Sci & Technol, 2-5-1 Akebono Cho, Kochi 7808520, Japan
关键词
Guard column; ion chromatography; anions; simultaneous separation; ion-exchange; STATIONARY PHASES; RIVER; RETENTION; NITROGEN; AREA;
D O I
10.2116/analsci.19P146
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This study demonstrated that a guard column containing anion-exchange resin has the potential for use as a separation column for acid eluent. Specifically, a 1-cm long anion-exchange guard column with a 4.6-mm internal diameter provided good separation of monovalent inorganic anions, by elution of 8 mM tartaric acid or 4 mM malic acid. Using the guard column with acid eluent could be applied to evaluation of nitrite and nitrate ions in mountain and urban river water samples. When the guard column was connected in front of a cation-exchange separation column (15 cm long x 4.6 mm internal diameter) in a series, the system provided simultaneous separation of anions and cations in eluent of 8 mM tartaric acid and 0.5 mM 18-crown-6 ether by a single injection.
引用
收藏
页码:1117 / 1122
页数:6
相关论文
共 50 条
  • [1] Utilization of Anion-exchange Guard Column as an Ion Chromatographic Column of Anions Including Application to Simultaneous Separation of Anions and Cations
    Sakura Yoshii
    Masanobu Mori
    Daisuke Kozaki
    Takayuki Hosokawa
    Hideyuki Itabashi
    Analytical Sciences, 2019, 35 : 1117 - 1122
  • [2] Tunable separation of anions and cations by column switching in ion chromatography
    Amin, Muhammad
    Lim, Lee Wah
    Takeuchi, Toyohide
    TALANTA, 2007, 71 (04) : 1470 - 1475
  • [3] Improvement of the chromatographic separation performance of an imidazolium ionic liquid functionalized silica column by in situ anion-exchange with dodecyl sulfonate and dodecylbenzene sulfonate anions
    Sun, Min
    Feng, Juanjuan
    Chen, Wenjie
    Li, Leilei
    Duan, Huimin
    Luo, Chuannan
    JOURNAL OF SEPARATION SCIENCE, 2014, 37 (11) : 1283 - 1288
  • [4] Simultaneous determination of anions and cations in rain water by single column ion-exclusion-cation-exchange chromatography
    Xu, Q
    Zhang, W
    Peng, HQ
    Wei, Q
    Jin, LT
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 28 (03) : 278 - 282
  • [5] Simultaneous separation and detection of anions and cations in ion chromatography
    Nesterenko, PN
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2001, 20 (6-7) : 311 - 319
  • [6] Simultaneous Analysis of Inorganic Monovalent Anions/Cations Using Mixed-Bed Single-Column Ion Chromatography
    Kazarian, Artaches A.
    Paull, Brett
    Nesterenko, Pavel N.
    Soisungnoen, Phimpha
    Burakham, Rodjana
    Srijaranai, Supalax
    CHROMATOGRAPHIA, 2015, 78 (3-4) : 179 - 187
  • [7] NEW APPROACH TO THE SIMULTANEOUS SEPARATION OF INORGANIC CATIONS AND ANIONS BY ION CHROMATOGRAPHY BASED ON ION-EXCHANGE AND ELECTROSTATIC INTERACTIONS
    HU, WZ
    MIYAZAKI, A
    HARAGUCHI, H
    ANALYTICAL SCIENCES, 1995, 11 (06) : 999 - 1000
  • [8] Effect of column temperature on the retention of inorganic anions and organic acids in non-suppressed anion-exchange IC
    Yu, Hong
    Li, Ruishu
    CHROMATOGRAPHIA, 2008, 68 (7-8) : 611 - 616
  • [9] Effect of Column Temperature on the Retention of Inorganic Anions and Organic Acids in Non-Suppressed Anion-Exchange IC
    Hong Yu
    Ruishu Li
    Chromatographia, 2008, 68 : 611 - 616
  • [10] SIMULTANEOUS ION CHROMATOGRAPHIC DETERMINATION OF ANIONS AND CATIONS BY SERIES CONDUCTIVITY AND FLAME PHOTOMETRIC DETECTION
    FRENZEL, W
    SCHEPERS, D
    SCHULZE, G
    ANALYTICA CHIMICA ACTA, 1993, 277 (01) : 103 - 111