SPECT imaging for radioaerosol deposition and clearance studies

被引:42
作者
Eberl, S
Chan, HK
Daviskas, E
机构
[1] Royal Prince Alfred Hosp, Dept PET & Nucl Med, Sydney, NSW, Australia
[2] Univ Sydney, Dept Pharm, Sydney, NSW 2006, Australia
[3] Royal Prince Alfred Hosp, Dept Resp Med, Sydney, NSW, Australia
来源
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG | 2006年 / 19卷 / 01期
关键词
aerosol deposition; aerosol clearance; lung; SPECT; radionuclide imaging;
D O I
10.1089/jam.2006.19.8
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Planar gamma camera scintigraphy is well established for measuring the deposition and clearance of radioaerosols. Single photon emission computed tomography (SPECT) provides three-dimensional (3D) reconstructions of the radioactivity distribution, thus avoiding the compression of 3D data into two-dimensional (2D) images and potentially offering superior assessment of aerosol deposition patterns. However, SPECT has traditionally been associated with long imaging times, making it unsuitable for measuring deposition and clearance of radioaerosols with fast clearance. Multi-detector SPECT systems can collect complete SPECT studies in <1 min, allowing both initial deposition and clearance over time to be assessed by dynamic SPECT. Simultaneous transmission measurement with an external source provides attenuation correction for absolute activity quantification as well as aiding in the definition of the lung volume of interest. A dynamic SPECT imaging protocol has been developed to allow fast imaging from the oropharynx to the abdomen using gamma cameras with limited axial field of views. This allows activity quantification not only in the lungs, but also in areas outside the thorax. However, fast dynamic SPECT imaging is technically and computationally more demanding and provides less scope for reducing the radioactivity administered to the subjects. It has been shown that dynamic SPECT, compared to planar imaging, is more sensitive in detecting changes in deposition as measured by the Penetration Index (PI). Thus, SPECT can better differentiate between large and small airways, which is important for lung regional analysis.
引用
收藏
页码:8 / 20
页数:13
相关论文
共 46 条
  • [21] Gullberg G. T., 1998, NUCL CARDIOLOGY STAT, P137
  • [22] ACCELERATED IMAGE-RECONSTRUCTION USING ORDERED SUBSETS OF PROJECTION DATA
    HUDSON, HM
    LARKIN, RS
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (04) : 601 - 609
  • [23] ICRP, 1998, ICRP PUBL, V80
  • [24] Quantitative assessment of regional myocardial blood flow with thallium-201 and SPECT
    Iida, H
    Eberl, S
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 1998, 5 (03) : 313 - 331
  • [25] IIDA H, 1994, J NUCL MED, V35, P2019
  • [26] In vivo imaging of nicotinic receptor upregulation following chronic (-)-nicotine treatment in baboon using SPECT
    Kassiou, M
    Eberl, S
    Meikle, SR
    Birrell, A
    Constable, C
    Fulham, MJ
    Wong, DF
    Musachio, JL
    [J]. NUCLEAR MEDICINE AND BIOLOGY, 2001, 28 (02) : 165 - 175
  • [27] The effect of scatter and attenuation on aerosol deposition as determined by gamma scintigraphy
    Lee, ZH
    Berridge, MS
    Nelson, AD
    Heald, DL
    [J]. JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 2001, 14 (02): : 167 - 183
  • [28] LINKS JM, 1991, J NUCL MED, V32, P2253
  • [29] MACEY DJ, 1982, J NUCL MED, V23, P731
  • [30] MEIKLE SR, 1994, J NUCL MED, V35, P360