Determination of the Hounsfield value for CT-based design of custom femoral stems

被引:69
作者
Aamodt, A [1 ]
Kvistad, KA [1 ]
Andersen, E [1 ]
Lund-Larsen, J [1 ]
Eine, J [1 ]
Benum, P [1 ]
Husby, OS [1 ]
机构
[1] Univ Trondheim Hosp, Dept Orthopaed Surg, Biomech Lab, N-7006 Trondheim, Norway
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1999年 / 81B卷 / 01期
关键词
D O I
10.1302/0301-620X.81B1.8880
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
CT and advanced computer-aided design techniques offer the means for designing customised femoral stems. Our aim was to determine the HounsfieId (HU) value of the bone at the corticocancellous interface, as part of the criteria for the design algorithm. We obtained transverse CT images from eight human cadaver femora, The proximal femoral canal was rasped until contact with dense cortical bone was achieved. The femora were cut into several sections corresponding to the slice positions of the CT images, After obtaining a computerised image of the anatomical sections using a scanner, the inner cortical contour was outlined and transferred to the corresponding CT image. The pixels beneath this contour represent the CT density of the bone remaining after surgical rasping. Contours were generated automatically at nine HU levels from 300 to 1100 and the mean distance between the transferred contour and each of the HU-generated contours was computed. The contour generated along the 600-HU pixels was closest to the inner cortical contour of the rasped femur and therefore 600 HU seem to be the CT density of the corticocancellous interface in the proximal part of cadaver femora, Generally, femoral bone with a CT density beyond 600 HU is not removable by conventional reamers, Thus, we recommend the 600 HU threshold as one of several criteria for the design of custom femoral implants from CT data.
引用
收藏
页码:143 / 147
页数:5
相关论文
共 10 条
  • [1] BARGAR WL, 1989, CLIN ORTHOP RELAT R, P73
  • [2] Garg A, 1985, Eng Med, V14, P89, DOI 10.1243/EMED_JOUR_1985_014_018_02
  • [3] COMPUTER DESIGN OF CUSTOM FEMORAL STEM PROSTHESES
    GRANHOLM, JW
    ROBERTSON, DD
    WALKER, PS
    NELSON, PC
    [J]. IEEE COMPUTER GRAPHICS AND APPLICATIONS, 1987, 7 (02) : 26 - 35
  • [4] Hua J, 1993, P ISSCP AM ISL FL, P11
  • [5] KVISTAD KA, 1994, P ISSCP ROM IT, P35
  • [6] Rothuizen P., 1987, BIOMECHANICS BASIC A, P109
  • [7] THE MORPHOLOGY OF THE PROXIMAL FEMUR - A 3-DIMENSIONAL RADIOGRAPHIC ANALYSIS
    RUBIN, PJ
    LEYVRAZ, PF
    AUBANIAC, JM
    ARGENSON, JN
    ESTEVE, P
    DEROGUIN, B
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01): : 28 - 32
  • [8] RUEBEN JD, 1992, CLIN ORTHOP RELAT R, V285, P48
  • [9] STULBERG SD, 1989, CLIN ORTHOP RELAT R, P79
  • [10] Tummler H P, 1990, Biomed Tech (Berl), V35 Suppl 3, P38