Results of titanium locking plate and stainless steel cerclage wire combination in femoral fractures

被引:13
作者
El-Zayat, Bilal Farouk [1 ]
Ruchholtz, Steffen [2 ]
Efe, Turgay [1 ]
Paletta, Juergen [1 ]
Kreslo, Dimitri [2 ]
Zettl, Ralph [2 ]
机构
[1] Univ Hosp Marburg, Dept Orthopaed, D-35043 Marburg, Germany
[2] Univ Hosp Marburg, Dept Trauma Hand & Reconstruct Surg, D-35043 Marburg, Germany
关键词
Cerclage wire; fracture healing; galvanic corrosion; locking plate; titanium; stainless steel; FEMUR FRACTURES; PERIPROSTHETIC FRACTURES; HIP; FIXATION; LISS;
D O I
10.4103/0019-5413.118200
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Some in vitro studies warn combining different metals in orthopedic surgery. The aim of this study is to determine the impact of combining titanium and stainless steel on bone healing and the clinical course of patients undergoing internal fixation of femoral fractures. Materials and Methods: 69 patients with femoral fractures had polyaxial locking plate osteosynthesis. The locking plate was made of a titanium alloy. Two different cohorts were defined: (a) sole plating and (b) additional stainless steel cerclage wiring. Postoperative radiographs and clinical followup were performed at 6 weeks, 3 months and 12 months. Results: Cohorts A and B had 36 and 33 patients, respectively. Patient demographics and comorbidities were similar in both groups. In two cases in cohort A, surgical revision was necessary. No complication could be attributed to the combination of titanium and stainless steel. Conclusion: The combination of stainless steel cerclage wires and titanium plates does not compromise fracture healing or the postoperative clinical course.
引用
收藏
页码:454 / 458
页数:5
相关论文
共 50 条
[31]   Management of distal femoral periprosthetic fractures by distal femoral locking plate: A retrospective study [J].
Rajiv Thukral ;
S. K. S. Marya ;
Chandeep Singh .
Indian Journal of Orthopaedics, 2015, 49 :199-207
[32]   Combination of low-contact cerclage wiring and osteosynthesis in the treatment of femoral fractures [J].
Angelini A. ;
Battiato C. .
European Journal of Orthopaedic Surgery & Traumatology, 2016, 26 (4) :397-406
[33]   Managing Vancouver B1 fractures by cerclage system compared to locking plate fixation - a biomechanical study [J].
Gordon, Katharina ;
Winkler, Martin ;
Hofstaedter, Thomas ;
Dorn, Ulrich ;
Augat, Peter .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2016, 47 :S51-S57
[34]   Investigation of a fatigue failure in a stainless steel femoral plate [J].
Marcomini, J. B. ;
Baptista, C. A. R. P. ;
Pascon, J. P. ;
Teixeira, R. L. ;
Reis, F. P. .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2014, 38 :52-58
[35]   Addition of a Medial Locking Plate to an In Situ Lateral Locking Plate Results in Healing of Distal Femoral Nonunions [J].
Holzman, Michael A. ;
Hanus, Bryan D. ;
Munz, John W. ;
O'Connor, Daniel P. ;
Brinker, Mark R. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2016, 474 (06) :1498-1505
[36]   Controlled study on Gamma nail and proximal femoral locking plate for unstable intertrochanteric femoral fractures with broken lateral wall [J].
Han, Lei ;
Liu, Jing-Jing ;
Hu, Yun-gen ;
Quan, Ren-fu ;
Fang, Wei-li ;
Jin, Bo ;
Lin, Wei-long .
SCIENTIFIC REPORTS, 2018, 8
[37]   Proximal femoral locking plate versus dynamic hip screw for unstable intertrochanteric femoral fractures [J].
Upadhyay, Sachin ;
Raza, H. K. T. .
JOURNAL OF ORTHOPAEDIC SURGERY, 2014, 22 (01) :130-131
[38]   Reverse distal femoral locking compression plate a salvage option in nonunion of proximal femoral fractures [J].
Patil, Sampat S. Dumbre ;
Karkamkar, Sachin S. ;
Patil, Vaishali S. Dumbre ;
Patil, Shailesh S. ;
Ranaware, Abhijeet S. .
INDIAN JOURNAL OF ORTHOPAEDICS, 2016, 50 (04) :374-378
[39]   Treatment of Delbet II/III Pediatric Femoral Neck Fractures With Proximal Femoral Locking Plate Versus Cannulated Screws [J].
Haider, Shamrez ;
Harris, Timothy J. ;
Turner, Alexander C. ;
Podeszwa, David A. ;
Hartman, Courtney A. ;
Morris, William Z. .
JOURNAL OF PEDIATRIC ORTHOPAEDICS, 2024, 44 (04) :213-220
[40]   Technical difficulties and mechanical failure of distal femoral locking compression plate (DFLCP) in management of unstable distal femoral fractures [J].
Siddiqui, Yasir Salam ;
Mohd, Julfiqar ;
Abbas, Mazhar ;
Gupta, Keshav ;
Khan, Mohammad Jesan ;
Istiyak, Md .
INTERNATIONAL JOURNAL OF BURNS AND TRAUMA, 2021, 11 (01) :9-19