Device Exchange After Primary Left Ventricular Assist Device Implantation: Indications and Outcomes

被引:68
作者
Stulak, John M.
Cowger, Jennifer
Haft, Jonathon W.
Romano, Matthew A.
Aaronson, Keith D.
Pagani, Francis D.
机构
[1] Mayo Clin, Coll Med, Div Cardiovasc Surg, Ann Arbor, MI USA
[2] Univ Michigan Hlth Syst, Dept Cardiac Surg, Ann Arbor, MI USA
关键词
OFF-PUMP INSERTION; HEARTMATE-XVE; EXPERIENCE; MANAGEMENT; STERNOTOMY; SYSTEM;
D O I
10.1016/j.athoracsur.2012.08.031
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Patients are being supported for longer periods with implantable left ventricular assist devices (LVADs) owing to longer transplantation wait times and approval of LVADs for destination therapy. This comes with an increased potential need for device exchange when complications arise. There are few data examining this patient population. Methods. Between August 1998 and January 2012, 45 patients (34 men) underwent 57 device exchanges after primary pulsatile or continuous-flow LVAD implantation. The median age at the initial LVAD implantation was 58 years (range, 28-78 years) and the median time to first device exchange was 15 months (range, immediate-56 months). Indications for primary LVAD included bridge to transplantation in all but 10 patients, and devices included the HeartMate I (Thoratec, Pleasanton, CA) in 16 patients, the HeartMate II (Thoratec) in 21 patients, the HeartWare HVAD (HeartWare, Framingham, MA) in 2 patients, the DuraHeart I (Terumo Heart, Ann Arbor, MI) in 1 patient, and other devices in 5 patients. Indications for reoperation included device/component failure (n = 24), major driveline infection (n = 15), pump thrombus (n = 15), and other indications (n = 2). Results. Pumps implanted in 57 reoperations included the HeartMate I in 15 patients, the HeartMate II in 35 patients, the HeartWare HVAD in 2 patients, the DuraHeart I in 2 patients, and other devices in 3 patients. Early mortality occurred in 2/57 (3.5%) patients. Median follow-up was 18 months (range, 1-113 months); median length of LVAD therapy after the first device exchange was 13 months (range, 1-59 months). Actuarial 1-year survival and freedom from repeated device exchange after the first exchange was 89% and 79%, respectively. Conclusions. Device exchange may be required after LVAD implantation. This can be performed with low early mortality and no adverse effect on late survival. Multiple reoperations may be required in some patients. (Ann Thorac Surg 2013;95:1262-8) (C) 2013 by The Society of Thoracic Surgeons
引用
收藏
页码:1262 / 1268
页数:7
相关论文
共 21 条
[1]   HeartMate Left Ventricular Assist System Exchange: Results and Technical Considerations [J].
Adamson, Robert M. ;
Dembitsky, Walter P. ;
Baradarian, Sam ;
Chammas, Joseph ;
Jaski, Brian ;
Hoagland, Peter ;
McCalmont, Vicki ;
Ortiz, Kristi ;
Stahovich, Marsha ;
Chillcott, Suzanne .
ASAIO JOURNAL, 2009, 55 (06) :598-601
[2]   Incidence and clinical management of life-threatening left ventricular assist device failure [J].
Birks, EJ ;
Tansley, PD ;
Yacoub, MH ;
Bowles, CT ;
Hipkin, M ;
Hardy, J ;
Banner, NR ;
Khaghani, A .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2004, 23 (08) :964-969
[3]  
Cohn WE, 2006, TEX HEART I J, V33, P48
[4]   Left ventricular assist device performance with long-term circulatory support: Lessons from the REMATCH trial [J].
Dembitsky, WP ;
Tector, AJ ;
Park, S ;
Moskowitz, AJ ;
Gelijns, AC ;
Ronan, NS ;
Piccione, W ;
Holman, WL ;
Furukawa, S ;
Frazier, OH ;
Weinberg, AD ;
Heatley, G ;
Poirier, VL ;
Damme, L ;
Long, JW .
ANNALS OF THORACIC SURGERY, 2004, 78 (06) :2123-2129
[5]   HeartMateo® VE LVAS design enhancements and its impact on device reliability [J].
Dowling, RD ;
Park, SJ ;
Pagani, FD ;
Tector, AJ ;
Naka, Y ;
Icenogle, TB ;
Poirier, VL ;
Frazier, OH .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2004, 25 (06) :958-963
[6]   Urgent pump exchange for stroke resulting from a distorted HeartMate XVE inflow conduit [J].
Frazier, O. H. ;
Forrester, Matthew D. ;
Gemmato, Courtney ;
Gregoric, Igor D. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2007, 26 (06) :646-648
[7]   Initial experience with non-thoracic, extraperitoneal, off-pump insertion of the Jarvik 2000 Heart in patients with previous median sternotomy [J].
Frazier, O. H. ;
Gregoric, Igor D. ;
Cohn, William E. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2006, 25 (05) :499-503
[8]   HeartMate vented-electric left ventricular assist system: Technique for intrathoracic or intraperitoneal implantation via a left thoracotomy [J].
Gregoric, I ;
Wadia, Y ;
Radovancevic, B ;
Messner, G ;
Khan, T ;
Frazier, OH .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2004, 23 (06) :759-762
[9]   Clinical experience with sternotomy versus subcostal approach for exchange of the HeartMate XVE to the HeartMate II ventricular assist device [J].
Gregoric, Igor D. ;
Bruckner, Brian A. ;
Jacob, Leon ;
Kar, Biswajit ;
Cohn, William E. ;
La Francesca, Saverio ;
Frazier, O. H. .
ANNALS OF THORACIC SURGERY, 2008, 85 (05) :1646-1650
[10]   Exchange techniques for implantable ventricular assist devices [J].
Gregoric, Igor D. .
ASAIO JOURNAL, 2008, 54 (01) :14-19