A simple modified open peritoneal dialysis catheter insertion procedure reduces the need for secondary surgery

被引:4
作者
Li, Yingqin [2 ]
Zhu, Ye [1 ]
Liang, Zibin [3 ]
Zheng, Xiaobin [4 ]
Zhang, Huitao [1 ]
Zhu, Weiping [1 ]
机构
[1] Sun Yat Sen Univ, Affiliated Hosp 5, Nephrol Dept, 52 Meihua Rd, Zhuhai 519000, Peoples R China
[2] Sun Yat Sen Univ, Affiliated Hosp 5, Radiol Dept, 52 Meihua Rd, Zhuhai 519000, Peoples R China
[3] Sun Yat Sen Univ, Affiliated Hosp 5, Radiotherapy Dept, 52 Meihua Rd, Zhuhai 519000, Peoples R China
[4] Sun Yat Sen Univ, Affiliated Hosp 5, Respirat Dept, 52 Meihua Rd, Zhuhai 519000, Peoples R China
关键词
Catheter; Chronic renal failure; Mechanical complication; Peritoneal dialysis; RANDOMIZED CONTROLLED-TRIAL; SINGLE-CENTER EXPERIENCE; PLACEMENT;
D O I
10.1007/s11255-019-02101-9
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Background The aim of this retrospective study was to assess the efficacy of a modified peritoneal dialysis catheter insertion technique for reducing the incidence of mechanical complications. Methods We conducted a retrospective analysis of clinical data of 346 patients undergoing peritoneal dialysis catheter insertion at our peritoneal dialysis center. The traditional procedure was performed in 157 patients (group A) and the modified procedure in 189 patients (group B). The double-polyester-cuff straight Tenckhoff catheter was used in all patients. Results At the end of 1 year, tunnel inflammation was more common in group A (21 patients after 0.011 patient-months follow-up versus 10 patients in group B after 0.007 patient-months of follow-up; p = 0.009). Technical survival rate of the catheter was significantly higher in group B (97.35% in group B vs. 89.81% in group A; p = 0.005). All-cause mortality was not significantly different between the two groups (4.5% in group A vs. 3.2% in group B; p = 0.532). Postoperative mechanical complications were also higher in group A (32 patients [20.4%] in group A vs. 3 patients [1.6%] in group B; p < 0.001). The incidences of complications such as hernia, dialysis fluid leakage, hemorrhage, incision infection, and prolapse of the polyester cuff were similar in the two groups. Conclusion The simple modified peritoneal dialysis catheter insertion procedure decreases the occurrence of catheter migration andomental encapsulation and improves the technical survival rate of the catheter.
引用
收藏
页码:729 / 736
页数:8
相关论文
共 26 条
[11]   A MODIFIED OPEN SURGERY TECHNIQUE FOR PERITONEAL DIALYSIS CATHETER PLACEMENT DECREASES CATHETER MALFUNCTION [J].
Jiang, Chunming ;
Xu, Linfeng ;
Chen, Yun ;
Yan, Xiang ;
Sun, Cheng ;
Zhang, Miao .
PERITONEAL DIALYSIS INTERNATIONAL, 2014, 34 (04) :358-367
[12]   A randomized controlled trial of coiled versus straight swan-neck Tenckhoff catheters in peritoneal dialysis patients [J].
Johnson, David W. ;
Wong, Jennifer ;
Wiggins, Kathryn J. ;
Kirwan, Robyn ;
Griffin, Anthony ;
Preston, John ;
Wall, Daryl ;
Campbell, Scott B. ;
Isbel, Nicole M. ;
Mudge, David W. ;
Hawley, Carmel M. ;
Nicol, David L. .
AMERICAN JOURNAL OF KIDNEY DISEASES, 2006, 48 (05) :812-821
[13]   Prospective Randomized Study for Comparison of Open Surgery with Laparoscopic-Assisted Placement of Tenckhoff Peritoneal Dialysis Catheter-A Single Center Experience and Literature Review [J].
Jwo, Shyh-Chuan ;
Chen, Kuo-Su ;
Lee, Chin-Chan ;
Chen, Huang-Yang .
JOURNAL OF SURGICAL RESEARCH, 2010, 159 (01) :489-496
[14]   Peritoneal Fixation Prevents Dislocation of Tenckhoff Catheter [J].
Kume, Haruki ;
Miyazaki, Hideyo ;
Nagata, Masayoshi ;
Ishikawa, Akira ;
Ishibashi, Yoshitaka ;
Fujita, Toshiro ;
Homma, Yukio .
PERITONEAL DIALYSIS INTERNATIONAL, 2011, 31 (06) :694-697
[15]  
McCartan Donna, 2015, Ulster Med J, V84, P166
[16]  
Minguela I, 2001, PERITON DIALYSIS INT, V21, pS209
[17]   New Peritoneal Catheters: New Catheter Problems? [J].
Moreiras-Plaza, Mercedes ;
Maria Martin-Baez, Isabel ;
Blanco-Garcia, Raquel ;
Vanessa Beato-Coo, Laura ;
Raquel Cossio-Aranibar, Cynthia .
PERITONEAL DIALYSIS INTERNATIONAL, 2014, 34 (05) :556-561
[18]   Peritoneal dialysis catheter placement technique and complications [J].
Peppelenbosch, Arnoud ;
van Kuijk, Willy H. M. ;
Bouvy, Nicole D. ;
van der Sande, Frank M. ;
Tordoir, Jan H. M. .
CLINICAL KIDNEY JOURNAL, 2008, 1 :IV23-IV28
[19]  
SPRINGBACK ED, 2009, INT J MIN MET MATER, V16, P177, DOI DOI 10.1016/S1674-4799(09)60030-3
[20]   Few Outflow Problems With a Self-locating Catheter for Peritoneal Dialysis A Randomized Trial [J].
Stegmayr, Bernd G. ;
Sperker, Wolfgang ;
Nilsson, Christina H. ;
Degerman, Christina ;
Persson, Sven-Erik ;
Stenbaek, Jan ;
Arnerlov, Conny .
MEDICINE, 2015, 94 (48)