Improvement of electrical blood hematocrit measurements under various plasma conditions using a novel hematocrit estimation parameter

被引:29
|
作者
Kim, Myounggon [1 ]
Kim, Ayoung [2 ]
Kim, Sohee [1 ,2 ]
Yang, Sung [1 ,2 ,3 ]
机构
[1] Gwangju Inst Sci & Technol GIST, Sch Mechatron, Kwangju 500712, South Korea
[2] Gwangju Inst Sci & Technol GIST, Dept Med Syst Engn, Kwangju 500712, South Korea
[3] Gwangju Inst Sci & Technol GIST, Dept Nanobio Mat & Elect, Kwangju 500712, South Korea
来源
BIOSENSORS & BIOELECTRONICS | 2012年 / 35卷 / 01期
基金
新加坡国家研究基金会;
关键词
Electrical impedance spectroscopy (EIS); Hematocrit (HCT); Electrical HCT measurement; Novel HCT estimation parameter; Measurement error; IMPEDANCE; SPECTROSCOPY; SAMPLES;
D O I
10.1016/j.bios.2012.02.010
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This paper presents an electrical method for measurement of Hematocrit (HCT) using a novel HCT estimation parameter. Particularly in the case of electrical HCT measurements, the measurement error generally increases with changes in the electrical conditions of the plasma such as conductivity and osmolality. This is because the electrical properties of blood are a function not only of HCT, but also of the electrical conditions in the plasma. In an attempt to reduce the measurement errors, we herein propose a novel HCT estimation parameter reflecting the characteristics of both the changes in volume of red blood cells (RBCs) and electrical conditions of plasma, simultaneously. In order to characterize the proposed methods under various electrical conditions of plasma, we prepared twelve blood samples such as four kinds of plasma conditions (hypotonic, isotonic, two kinds of hypertonic conditions) at three different HCT levels. Using linear regression analysis, we confirmed that the proposed parameter was highly correlated with reference HCT (HCTref) values measured by microcentrifugation. Thus, the HCT measurement error was less than 4%, despite considerable variations in the conductivity and osmolality of the plasma at conditions of the HCTref. of 20%. Multiple linear regression analysis showed that the proposed HCT estimation parameter also yielded a lower measurement error (1%) than the other parameter previously used for the same purpose. Thus, these preliminary results suggest that proposed method could be used for accurate, fast, easy, and reproducible HCT measurements in medical procedures. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:416 / 420
页数:5
相关论文
共 50 条
  • [1] Improvement of the accuracy of continuous hematocrit measurement under various blood flow conditions
    Kim, Myounggon
    Yang, Sung
    APPLIED PHYSICS LETTERS, 2014, 104 (15)
  • [2] ELECTRICAL-IMPEDANCE AND HEMATOCRIT OF HUMAN BLOOD WITH VARIOUS ANTICOAGULANTS
    ZHAO, TX
    PHYSIOLOGICAL MEASUREMENT, 1993, 14 (03) : 299 - 307
  • [3] THE EFFECT OF VARIOUS ANTICOAGULANTS ON THE SPECIFIC GRAVITY OF BLOOD AND OF PLASMA AND ON THE HEMATOCRIT
    MCLAIN, PL
    RUHE, CHW
    PASTORIUS, GJ
    BLOOD, 1949, 4 (07) : 863 - 868
  • [4] Simultaneous electrical online estimation of changes in blood hematocrit and temperature in cardiopulmonary bypass
    Martin Wekesa Sifuna
    Daisuke Kawashima
    Katsuhiro Matsuura
    Hiromichi Obara
    Yusuke Nakajima
    Masahiro Takei
    Journal of Artificial Organs, 2022, 25 : 305 - 313
  • [5] HEMOGLOBIN ERYTHROCYTES AND HEMATOCRIT IN BLOOD OF SOME MICROTIDAE UNDER LABORATORY CONDITIONS
    KOSTELEC.A
    BULLETIN DE L ACADEMIE POLONAISE DES SCIENCES-SERIE DES SCIENCES BIOLOGIQUES, 1966, 14 (05): : 343 - &
  • [6] Simultaneous electrical online estimation of changes in blood hematocrit and temperature in cardiopulmonary bypass
    Sifuna, Martin Wekesa
    Kawashima, Daisuke
    Matsuura, Katsuhiro
    Obara, Hiromichi
    Nakajima, Yusuke
    Takei, Masahiro
    JOURNAL OF ARTIFICIAL ORGANS, 2022, 25 (04) : 305 - 313
  • [7] Whole blood viscosity, plasma viscosity and serum viscosity in various levels of hematocrit and hemoglobin
    Wajima, T
    Teague, OE
    THROMBOSIS AND HAEMOSTASIS, 1999, : 151 - 151
  • [8] PLASMA AND BLOOD VOLUME IN THE COW USING THE T-1824 HEMATOCRIT METHOD
    REYNOLDS, M
    AMERICAN JOURNAL OF PHYSIOLOGY, 1953, 173 (03): : 421 - 427
  • [9] COMPARATIVE ESTIMATION OF HEMATOCRIT AND TRAPPED PLASMA IN THE PACKED CELL-VOLUME IN MAN, RABBIT AND CHICKEN BLOOD
    FERRANDO, A
    BOBADILLA, IG
    BOBADILLA, MS
    PALOU, A
    ALEMANY, M
    COMPARATIVE BIOCHEMISTRY AND PHYSIOLOGY A-PHYSIOLOGY, 1981, 70 (04): : 611 - 613
  • [10] A correction method using a support vector machine to minimize hematocrit interference in blood glucose measurements
    Shin, Jaeyeon
    Park, Hodong
    Cho, Sungpil
    Nam, Hakhyun
    Lee, Kyoung-Joung
    COMPUTERS IN BIOLOGY AND MEDICINE, 2014, 52 : 111 - 118