Impact of varying planning parameters on proton pencil beam scanning dose distributions in four commercial treatment planning systems

被引:15
作者
Alshaikhi, Jailan [1 ,2 ,3 ]
Doolan, Paul J. [4 ]
D'Souza, Derek [2 ]
Holloway, Stacey McGowan [1 ,5 ]
Amos, Richard A. [1 ]
Royle, Gary [1 ]
机构
[1] UCL, Dept Med Phys & Biomed Engn, London, England
[2] Univ Coll London Hosp NHS Fdn Trust, Dept Radiotherapy Phys, London, England
[3] Saudi Particle Therapy Ctr, Riyadh, Saudi Arabia
[4] German Oncol Ctr, Dept Med Phys, Limassol, Cyprus
[5] NIHR Univ Coll London Hosp Biomed Res Ctr, London, England
关键词
particle therapy; proton therapy; treatment planning; MONTE-CARLO; THERAPY; RADIOTHERAPY; INTERPLAY;
D O I
10.1002/mp.13382
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose In pencil beam scanning proton therapy, target coverage is achieved by scanning the pencil beam laterally in the x- and y-directions and delivering spots of dose to positions at a given radiological depth (layer). Dose is delivered to the spots on different layers by pencil beams of different energy until the entire volume has been irradiated. The aim of this study is to investigate the implementation of proton planning parameters (spot spacing, layer spacing and margins) in four commercial proton treatment planning systems (TPSs): Eclipse, Pinnacle(3), RayStation and XiO. Materials and Methods Using identical beam data in each TPS, plans were created on uniform material synthetic phantoms with cubic targets. The following parameters were systematically varied in each TPS to observe their different implementations: spot spacing, layer spacing and margin. Additionally, plans were created in Eclipse to investigate the impact of these parameters on plan delivery and optimal values are suggested. Results It was found that all systems except Eclipse use a variable layer spacing per beam, based on the Bragg peak width of each energy layer. It is recommended that if this cannot be used, then a constant value of 5 mm will ensure good dose homogeneity. Only RayStation varies the spot spacing according to the variable spot size with depth. If a constant spot spacing is to be used, a value of 5 mm is recommended as a good compromise between dose homogeneity, plan robustness and planning time. It was found that both Pinnacle(3) and RayStation position spots outside of the defined volume (target plus margin). Conclusions All four systems are capable of delivering uniform dose distributions to simple targets, but their implementation of the various planning parameters is different. In this paper comparisons are made between the four systems and recommendations are made as to the values that will provide the best compromise in dose homogeneity and planning time. (c) 2019 American Association of Physicists in Medicine
引用
收藏
页码:1150 / 1162
页数:13
相关论文
共 50 条
  • [41] FLASH radiotherapy sparing effect on the circulating lymphocytes in pencil beam scanning proton therapy: impact of hypofractionation and dose rate
    Galts, Antje
    Hammi, Abdelkhalek
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (02)
  • [42] Validation and application of a fast Monte Carlo algorithm for assessing the clinical impact of approximations in analytical dose calculations for pencil beam scanning proton therapy
    Huang, Sheng
    Souris, Kevin
    Li, Siyang
    Kang, Minglei
    Montero, Ana Maria Barragan
    Janssens, Guillaume
    Lin, Alexander
    Garver, Elizabeth
    Ainsley, Christopher
    Taylor, Paige
    Xiao, Ying
    Lin, Liyong
    MEDICAL PHYSICS, 2018, 45 (12) : 5631 - 5642
  • [43] Evaluation and second check of a commercial Monte Carlo dose engine for small-field apertures in pencil beam scanning proton therapy
    Holmes, Jason
    Shen, Jiajian
    Shan, Jie
    Patrick, Christopher L.
    Wong, William W.
    Foote, Robert L.
    Patel, Samir H.
    Bues, Martin
    Liu, Wei
    MEDICAL PHYSICS, 2022, 49 (05) : 3497 - 3506
  • [44] Mitigation of motion effects in pencil-beam scanning-Impact of repainting on 4D robustly optimized proton treatment plans for hepatocellular carcinoma
    Siregar, Hilda
    Baumer, Christian
    Blanck, Oliver
    Chan, Mark
    Engwall, Erik
    Plaude, Sandija
    Spaan, Bernhard
    Timmermann, Beate
    Wulff, Jorg
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2022, 32 (01): : 63 - 73
  • [45] Impact of data transfer between treatment planning systems on dosimetric parameters
    Meffe, Guenda
    Votta, Claudio
    Turco, Gabriele
    Chille, Elena
    Nardini, Matteo
    Romano, Angela
    Chiloiro, Giuditta
    Panza, Giulia
    Galetto, Matteo
    Capotosti, Amedeo
    Moretti, Roberto
    Gambacorta, Maria Antonietta
    Boldrini, Luca
    Indovina, Luca
    Placidi, Lorenzo
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2024, 121
  • [46] Use of single-energy proton pencil beam scanning Bragg peak for intensity-modulated proton therapy FLASH treatment planning in liver-hypofractionated radiation therapy
    Wei, Shouyi
    Lin, Haibo
    Shi, Chengyu
    Xiong, Weijun
    Chen, Chin-Cheng
    Huang, Sheng
    Press, Robert H.
    Hasan, Shaakir
    Chhabra, Arpit M.
    Choi, J. Isabelle
    Simone, Charles B., II
    Kang, Minglei
    MEDICAL PHYSICS, 2022, 49 (10) : 6560 - 6574
  • [47] Technical Note: Investigating interplay effects in pencil beam scanning proton therapy with a 4D XCAT phantom within the RayStation treatment planning system
    den Boer, Erik
    Wulff, Joerg
    Maeder, UIf
    Engwall, Erik
    Baeumer, Christian
    Perko, Zoltan
    Timmermann, Beate
    MEDICAL PHYSICS, 2021, 48 (03) : 1448 - 1455
  • [48] A Monte Carlo evaluation of dose distribution of commercial treatment planning systems in heterogeneous media
    Hasani, Mohsen
    Mohammadi, Kheirollah
    Ghorbani, Mahdi
    Gholami, Soraya
    Knaup, Courtney
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2019, 15 : S127 - S134
  • [49] Proton pencil-beam scanning radiotherapy in the treatment of nasopharyngeal cancer: dosimetric parameters and 2-year results
    Kubes, Jiri
    Vondracek, Vladimir
    Andrlik, Michal
    Navratil, Matej
    Slavikova, Silvia
    Vitek, Pavel
    Dedeckova, Katerina
    Prausova, Jana
    Ondrova, Barbora
    Rotnaglova, Eliska
    Lukes, Petr
    Patzelt, Matej
    Grebenyuk, Alexander
    Rosina, Jozef
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2021, 278 (03) : 763 - 769
  • [50] Proton pencil-beam scanning radiotherapy in the treatment of nasopharyngeal cancer: dosimetric parameters and 2-year results
    Kubeš Jiří
    Vondráček Vladimír
    Andrlik Michal
    Navrátil Matěj
    Sláviková Silvia
    Vítek Pavel
    Dědečková Kateřina
    Prausová Jana
    Ondrová Barbora
    Rotnáglová Eliška
    Lukeš Petr
    Patzelt Matěj
    Grebenyuk Alexander
    Rosina Jozef
    European Archives of Oto-Rhino-Laryngology, 2021, 278 : 763 - 769