Impact of varying planning parameters on proton pencil beam scanning dose distributions in four commercial treatment planning systems

被引:15
|
作者
Alshaikhi, Jailan [1 ,2 ,3 ]
Doolan, Paul J. [4 ]
D'Souza, Derek [2 ]
Holloway, Stacey McGowan [1 ,5 ]
Amos, Richard A. [1 ]
Royle, Gary [1 ]
机构
[1] UCL, Dept Med Phys & Biomed Engn, London, England
[2] Univ Coll London Hosp NHS Fdn Trust, Dept Radiotherapy Phys, London, England
[3] Saudi Particle Therapy Ctr, Riyadh, Saudi Arabia
[4] German Oncol Ctr, Dept Med Phys, Limassol, Cyprus
[5] NIHR Univ Coll London Hosp Biomed Res Ctr, London, England
关键词
particle therapy; proton therapy; treatment planning; MONTE-CARLO; THERAPY; RADIOTHERAPY; INTERPLAY;
D O I
10.1002/mp.13382
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose In pencil beam scanning proton therapy, target coverage is achieved by scanning the pencil beam laterally in the x- and y-directions and delivering spots of dose to positions at a given radiological depth (layer). Dose is delivered to the spots on different layers by pencil beams of different energy until the entire volume has been irradiated. The aim of this study is to investigate the implementation of proton planning parameters (spot spacing, layer spacing and margins) in four commercial proton treatment planning systems (TPSs): Eclipse, Pinnacle(3), RayStation and XiO. Materials and Methods Using identical beam data in each TPS, plans were created on uniform material synthetic phantoms with cubic targets. The following parameters were systematically varied in each TPS to observe their different implementations: spot spacing, layer spacing and margin. Additionally, plans were created in Eclipse to investigate the impact of these parameters on plan delivery and optimal values are suggested. Results It was found that all systems except Eclipse use a variable layer spacing per beam, based on the Bragg peak width of each energy layer. It is recommended that if this cannot be used, then a constant value of 5 mm will ensure good dose homogeneity. Only RayStation varies the spot spacing according to the variable spot size with depth. If a constant spot spacing is to be used, a value of 5 mm is recommended as a good compromise between dose homogeneity, plan robustness and planning time. It was found that both Pinnacle(3) and RayStation position spots outside of the defined volume (target plus margin). Conclusions All four systems are capable of delivering uniform dose distributions to simple targets, but their implementation of the various planning parameters is different. In this paper comparisons are made between the four systems and recommendations are made as to the values that will provide the best compromise in dose homogeneity and planning time. (c) 2019 American Association of Physicists in Medicine
引用
收藏
页码:1150 / 1162
页数:13
相关论文
共 50 条
  • [1] A comparison of the dose distributions from three proton treatment planning systems in the planning of meningioma patients with single-field uniform dose pencil beam scanning
    Doolan, Paul J.
    Alshaikhi, Jailan
    Rosenberg, Ivan
    Ainsley, Christopher G.
    Gibson, Adam
    D'Souza, Derek
    Bentefour, El Hassane
    Royle, Gary
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (01): : 86 - 99
  • [2] Automation of pencil beam scanning proton treatment planning for intracranial tumours
    Placidi, Lorenzo
    Righetto, Roberto
    Vecchi, Claudio
    Zara, Stefania
    Alparone, Alessandro
    Moretti, Roberto
    Amelio, Dante
    Scartoni, Daniele
    Schwarz, Marco
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 105
  • [3] A benchmarking method to evaluate the accuracy of a commercial proton monte carlo pencil beam scanning treatment planning system
    Lin, Liyong
    Huang, Sheng
    Kang, Minglei
    Hiltunen, Petri
    Vanderstraeten, Reynald
    Lindberg, Jari
    Siljamaki, Sami
    Wareing, Todd
    Davis, Ian
    Barnett, Allen
    McGhee, John
    Simone, Charles B., II
    Solberg, Timothy D.
    McDonough, James E.
    Ainsley, Christopher
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (02): : 44 - 49
  • [4] A comparison of two pencil beam scanning treatment planning systems for proton therapy
    Langner, Ulrich W.
    Mundis, Michelle
    Strauss, Dan
    Zhu, Mingyao
    Mossahebi, Sina
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (01): : 156 - 163
  • [5] Effects of spot parameters in pencil beam scanning treatment planning
    Kraan, Aafke Christine
    Depauw, Nicolas
    Clasie, Ben
    Giunta, Marina
    Madden, Tom
    Kooy, Hanne M.
    MEDICAL PHYSICS, 2018, 45 (01) : 60 - 73
  • [6] Proton pencil beam scanning for mediastinal lymphoma: treatment planning and robustness assessment
    Zeng, Chuan
    Plastaras, John P.
    James, Paul
    Tochner, Zelig A.
    Hill-Kayser, Christine E.
    Hahn, Stephen M.
    Both, Stefan
    ACTA ONCOLOGICA, 2016, 55 (9-10) : 1132 - 1138
  • [7] Accurate proton treatment planning for pencil beam crossing titanium fixation implants
    Righetto, Roberto
    Clemens, Luis Pazos
    Lorentini, Stefano
    Fracchiolla, Francesco
    Algranati, Carlo
    Tommasino, Francesco
    Dionisi, Francesco
    Cianchetti, Marco
    Schwarz, Marco
    Farace, Paolo
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 70 : 28 - 38
  • [8] A standardized commissioning framework of Monte Carlo dose calculation algorithms for proton pencil beam scanning treatment planning systems
    Chang, Chih-Wei
    Huang, Sheng
    Harms, Joseph
    Zhou, Jun
    Zhang, Rongxiao
    Dhabaan, Anees
    Slopsema, Roelf
    Kang, Minglei
    Liu, Tian
    McDonald, Mark
    Langen, Katja
    Lin, Liyong
    MEDICAL PHYSICS, 2020, 47 (04) : 1545 - 1557
  • [9] Clinical validation of a GPU-based Monte Carlo dose engine of a commercial treatment planning system for pencil beam scanning proton therapy
    Fracchiolla, Francesco
    Engwall, Erik
    Janson, Martin
    Tamm, Fredrik
    Lorentini, Stefano
    Fellin, Francesco
    Bertolini, Mattia
    Algranati, Carlo
    Righetto, Roberto
    Farace, Paolo
    Amichetti, Maurizio
    Schwarz, Marco
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 88 : 226 - 234
  • [10] Beam-specific planning target volumes incorporating 4D CT for pencil beam scanning proton therapy of thoracic tumors
    Lin, Liyong
    Kang, Minglei
    Huang, Sheng
    Mayer, Rulon
    Thomas, Andrew
    Solberg, Timothy D.
    McDonough, Jame E.
    Simone, Charles B., II
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (06): : 281 - 292