Determination of celecoxib in human plasma by liquid chromatography-tandem mass spectrometry

被引:7
作者
Ptacek, Pavel [1 ]
Klima, Josef [1 ]
Macek, Jan [1 ]
机构
[1] Pharmakl Sro, CZ-15800 Prague 13, Czech Republic
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2012年 / 899卷
关键词
Celecoxib; Plasma; LC-MS/MS; Protein precipitation; Pharmacokinetics; CLINICAL PHARMACOKINETICS; ARTHRITIS;
D O I
10.1016/j.jchromb.2012.05.013
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A liquid chromatography-electrospray tandem mass spectrometry method was developed and validated to quantitate celecoxib in human plasma. The assay was based on protein precipitation with methanol and liquid chromatography on a C-18 column (55 mm x 2 mm, 3 mu m). the mobile phase consisted of methanol - 10 mM ammonium acetate (75:25, v/v). Quantification was performed by mass spectrometry in the multiple reaction monitoring mode with negative electrospray ionization at m/z 380 -> 316 and 384 -> 320 for celecoxib and the internal standard celecoxib-D-4, respectively. The lower limit of quantitation was 7.0 ng/ml using 0.1 ml of plasma and linearity was demonstrated up to 1800 ng/ml. Intra-assay and interassay precision expressed by relative standard deviation was less than 4% and inaccuracy did not exceed 6% at all levels. The assay was applied to the analysis of samples from a pharmacokinetic study. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:163 / 166
页数:4
相关论文
共 11 条
[1]   Liquid chromatographic-mass spectrometric determination of celecoxib in plasma using single-ion monitoring and its use in clinical pharmacokinetics [J].
Abdel-Hamid, M ;
Novotny, L ;
Hamza, H .
JOURNAL OF CHROMATOGRAPHY B, 2001, 753 (02) :401-408
[2]   Determination of celecoxib in human plasma and rat microdialysis samples by liquid chromatography tandem mass spectrometry [J].
Bräutigam, L ;
Vetter, G ;
Tegeder, I ;
Heinkele, G ;
Geisslinger, G .
JOURNAL OF CHROMATOGRAPHY B, 2001, 761 (02) :203-212
[3]   Determination of celecoxib in human plasma using solid-phase extraction and high-performance liquid chromatography [J].
Chow, HHS ;
Anavy, N ;
Salazar, D ;
Frank, DH ;
Alberts, DS .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2004, 34 (01) :167-174
[4]   Clinical pharmacokinetics and pharmacodynamics of celecoxib - A selective cyclo-oxygenase-2 inhibitor [J].
Davies, NM ;
McLachlan, AJ ;
Day, RO ;
Williams, KM .
CLINICAL PHARMACOKINETICS, 2000, 38 (03) :225-242
[5]  
Emami J, 2008, DARU, V16, P211
[6]   Celecoxib - A review of its use in the management of arthritis and acute pain [J].
Frampton, James E. ;
Keating, Gillian M. .
DRUGS, 2007, 67 (16) :2433-2472
[7]   Determination of two COX-2 inhibitors in serum and synovial fluid of patients with inflammatory arthritis by ultra performance liquid chromatography-inductively coupled plasma mass spectroscopy and quadrupole time-of-flight mass spectrometry [J].
Gika, Helen G. ;
Theodoridou, Athina ;
Michopoulos, Filippos ;
Theodoridis, Georgios ;
Diza, Eudoxia ;
Settas, Lucas ;
Nikolaidis, Paul ;
Smith, Christopher ;
Wilson, Ian D. .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2009, 49 (03) :579-586
[8]   Determination of celecoxib in human plasma by high-performance liquid chromatography [J].
Jalalizadeh, H ;
Amini, M ;
Ziaee, V ;
Safa, A ;
Farsam, H ;
Shafiee, A .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2004, 35 (03) :665-670
[9]   Investigation of the pharmacokinetics of celecoxib by liquid chromatography-mass spectrometry [J].
Werner, U ;
Werner, D ;
Pahl, A ;
Mundkowski, R ;
Gillich, M ;
Brune, K .
BIOMEDICAL CHROMATOGRAPHY, 2002, 16 (01) :56-60
[10]   Simple and rapid high-performance liquid chromatographic method for determination of celecoxib in plasma using UV detection: Application in pharmacokinetic studies [J].
Zarghi, A ;
Shafaati, A ;
Foroutan, SM ;
Khoddam, A .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2006, 835 (1-2) :100-104