Validation of noninvasive oscillometric blood pressure 2020 up pressure upper arm blood pressure monitoring technology according to the European Society of Hypertension International Protocol revision 2010

被引:3
作者
Reshetnik, Alexander [1 ,2 ,3 ,4 ]
Gohlisch, Christopher [1 ,2 ,3 ,4 ]
Abou-Dakn, Michael [5 ]
Toelle, Markus [1 ,2 ,3 ,4 ]
Zidek, Walter [1 ,2 ,3 ,4 ]
van der Giet, Markus [1 ,2 ,3 ,4 ]
机构
[1] Charite Univ Med Berlin, Hindenburgdamm 30, D-12203 Berlin, Germany
[2] Free Univ Berlin, Hindenburgdamm 30, D-12203 Berlin, Germany
[3] Humboldt Univ, Hindenburgdamm 30, D-12203 Berlin, Germany
[4] Berlin Inst Hlth, Dept Nephrol, Campus Benjamin Franklin,Hindenburgdamm 30, D-12203 Berlin, Germany
[5] St Josephs Hosp Berlin Tempelhof, Dept Gynecol & Obstet, Berlin, Germany
关键词
automatic; blood pressure; European Society of Hypertension International Protocol revision 2010; noninvasive blood pressure 2020; noninvasive; oscillometric; upper arm; validation;
D O I
10.1097/MBP.0000000000000370
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
To validate noninvasive oscillometric blood pressure (NIBP) 2020 up pressure (UP) upper arm blood pressure (BP) monitoring technology according to the European Society of Hypertension International Protocol revision 2010 (ESH-IP revision 2010). According to the protocol, 33 participants were included. NIBP2020 UP technology was integrated into the BP+ host device, and common upper arm cuffs were used. The study included patients from ambulatory Hypertension Units. The validation procedure was performed exactly as required by ESH-IP revision 2010. Overall, 57.6% of the study population was female. Mean age was 45.5 +/- 16.3 years. Mean systolic blood pressure (SBP) was 142.9 +/- 30.6 mmHg and mean diastolic blood pressure (DBP) was 89.0 +/- 24.9 mmHg. The number of measurements within the difference range of up to 5 mmHg was 86 for SBP and 92 for DBP; 95 measurements of SBP and 99 of DBP were within the difference range of up to 10 mmHg, and 97 measurements of SBP and 99 measurement of DBP were in the difference range of up to 15 mmHg. Mean observer-test device differences were -0.2 +/- 4.6 mmHg for SBP and 0.6 +/- 3.1 mmHg for DBP. At least two observer-test differences were up to 5 mmHg for SBP in all participants and for DBP in 31 participants. All participants had at least one measurement with an observer-test difference of up to 5 mmHg for SBP and DBP. The NIBP2020 UP technology integrated into the BP+ host device using common upper arm cuffs passed the validation criteria according to the ESH-IP revision 2010. Copyright (c) 2019 Wolters Kluwer Health, Inc. All rights reserved.
引用
收藏
页码:99 / 101
页数:3
相关论文
共 1 条
[1]   European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults [J].
O'Brien, Eoin ;
Atkins, Neil ;
Stergiou, George ;
Karpettas, Nikos ;
Parati, Gianfranco ;
Asmar, Roland ;
Imai, Yutaka ;
Wang, Jiguang ;
Mengden, Thomas ;
Shennan, Andrew .
BLOOD PRESSURE MONITORING, 2010, 15 (01) :23-38