Direct analysis in real time-mass spectrometry (DART-MS) for rapid qualitative screening of toxic glycols in glycerin-containing products

被引:20
|
作者
Self, Randy L. [1 ]
机构
[1] US FDA, Off Regulatory Affairs, Pacific Reg Lab Northwest, Appl Technol Ctr, Bothell, WA 98021 USA
关键词
Direct analysis in real time; Orbitrap; Diethylene glycol; Ethylene glycol; Glycerin; DIETHYLENE GLYCOL; INTOXICATION;
D O I
10.1016/j.jpba.2013.02.037
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In 2007, the United States Food and Drug Administration released guidance recommending testing of glycerin used in regulated consumer products, such as cough syrup preparations, toothpaste, and other pharmaceutical and food products, for the toxic compounds ethylene glycol and diethylene glycol. Regulatory laboratories routinely test glycerin, and products containing glycerin or related compounds for these toxic glycols, using an official gas chromatographic method, to ensure the safety of these products. The current work describes a companion technique to compliment this GC-FID method utilizing Orbitrap mass spectrometry with direct analysis in real time ionization to rapidly screen these samples qualitatively, with results in as little as five seconds, with no sample preparation required. This allows the more time and resource intensive method to be reserved for those rare cases when these compounds are detected, potentially greatly improving laboratory efficiency. The technique was evaluated for qualitative sensitivity and repeatability, and compared against the GC-FID method. The method appears to perform well against these metrics. Published by Elsevier B.V.
引用
收藏
页码:155 / 158
页数:4
相关论文
共 44 条
  • [31] Rapid fingerprinting of source and environmental microplastics using direct analysis in real time-high resolution mass spectrometry
    Zhang, Xianming
    Mell, Alicia
    Li, Frederick
    Thaysen, Clara
    Musselman, Brian
    Tice, Joseph
    Vukovic, Dragan
    Rochman, Chelsea
    Helm, Paul A.
    Jobst, Karl J.
    ANALYTICA CHIMICA ACTA, 2020, 1100 : 107 - 117
  • [32] Direct analysis in real time - high resolution mass spectrometry (DART-HRMS): a high throughput strategy for identification and quantification of anabolic steroid esters
    Doue, Mickael
    Dervilly-Pinel, Gaud
    Pouponneau, Karinne
    Monteau, Fabrice
    Le Bizec, Bruno
    DRUG TESTING AND ANALYSIS, 2015, 7 (07) : 603 - 608
  • [33] Prediction of acrylamide formation in biscuits based on fingerprint data generated by ambient ionization mass spectrometry employing direct analysis in real time (DART) ion source
    Vaclavik, Lukas
    Capuano, Edoardo
    Gokmen, Vural
    Hajslova, Jana
    FOOD CHEMISTRY, 2015, 173 : 290 - 297
  • [34] Rapid analysis of caffeine in various coffee samples employing direct analysis in real-time ionization-high-resolution mass spectrometry
    Danhelova, Hana
    Hradecky, Jaromir
    Prinosilova, Sarka
    Cajka, Tomas
    Riddellova, Katerina
    Vaclavik, Lukas
    Hajslova, Jana
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2012, 403 (10) : 2883 - 2889
  • [35] Rapid analysis of caffeine in various coffee samples employing direct analysis in real-time ionization–high-resolution mass spectrometry
    Hana Danhelova
    Jaromir Hradecky
    Sarka Prinosilova
    Tomas Cajka
    Katerina Riddellova
    Lukas Vaclavik
    Jana Hajslova
    Analytical and Bioanalytical Chemistry, 2012, 403 : 2883 - 2889
  • [36] Quantitative analysis of phosphoric acid esters in aqueous samples by isotope dilution stir-bar sorptive extraction combined with direct analysis in real time (DART)-Orbitrap mass spectrometry
    Bridoux, Maxime C.
    Malandain, Helene
    Leprince, Francoise
    Progent, Frederic
    Machuron-Mandard, Xavier
    ANALYTICA CHIMICA ACTA, 2015, 869 : 1 - 10
  • [37] Application of direct analysis in real time-orbitrap mass spectrometry combined with multivariate data analysis for rapid quality assessment of Yuanhu Zhitong Tablet
    Li, Lele
    Liu, Shang
    Ma, Li
    Guo, Yunlong
    Wang, Yang
    Liu, Shuying
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2017, 421 : 33 - 39
  • [38] Authentication of milk and milk-based foods by direct analysis in real time ionization-high resolution mass spectrometry (DART-HRMS) technique: A critical assessment
    Hrbek, Vojtech
    Vaclavik, Lukas
    Elich, Ondrej
    Hajslova, Jana
    FOOD CONTROL, 2014, 36 (01) : 138 - 145
  • [39] Rapid Analysis of Illegal Cationic Dyes in Foods and Surface Waters Using High Temperature Direct Analysis in Real Time High-Resolution Mass Spectrometry
    Wen, Ruizhi
    Zeng, Dong
    Yang, Zihui
    Jiang, Le
    Ma, Ming
    Chen, Bo
    van Beek, Teris A.
    JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 2018, 66 (28) : 7542 - 7549
  • [40] Direct analysis in real time mass spectrometry and multivariate data analysis: A novel approach to rapid identification of analytical markers for quality control of traditional Chinese medicine preparation
    Zeng, Shanshan
    Wang, Lu
    Chen, Teng
    Wang, Yuefei
    Mo, Huanbiao
    Qu, Haibin
    ANALYTICA CHIMICA ACTA, 2012, 733 : 38 - 47